GE CF-LP Manual page 4

Programming the calibration factor for clamp-on transducers

Advertisement

Programming the Calibration Factor for CF-LP Clamp-on Transducers
June 2012
PT878
1. Press ESC until you are at the home screen (where pressing ESC no longer changes the screen).
2. Press the Menu key.
3. Using the arrow keys, scroll to Program. Press ENTER.
4. Using the arrow keys, scroll to Correction Factors. Press ENTER.
5. Using the arrow keys, scroll to the Calibration Factor tab. Press ENTER.
6. Using the arrow keys, scroll to the first set of radio buttons to turn the Calibration Factor On.
7. Using the arrow keys, scroll to the next set of radio buttons and choose Single K-Factor.
8. Using the arrow keys, scroll to the Meter K-Factor entry. Press ENTER.
9. In the dialog box, enter the Calibration Factor engraved on the mounting rail. If there is no engraving, enter
1.000. Press ENTER.
10.Press the F3 (OK) button to accept the changes.
Note: For a visual representation of the steps above, please refer to the menu map (Figure 152) in the PT878 User
Manual.