2.2.4
Turn plate (MOBILETT Plus/Plus M only)
Fig. 7:
•
Connect the mains cable and switch the system ON.
•
Move the unit to a level surface.
PMF
Button illumination
•
Press one of the turn plate buttons. The buttons must light up when the turn plate is acti-
vated.
PMF
Function
•
Check whether the system can be rotated around the turn plate
•
Press the turn plate button again and check whether the turn plate retracts to the park-
ing position.
•
Repeat the same test with the other turn plate button. If a problem arises, proceed ac-
cording to the section "Adjustment and replacement of the turn plate" in the service in-
structions.
MOBILETT Plus/ -E/ -M
SPR8-215.831.04.03.02
07.05
Inspection and maintenance
Page 28 of 64