Test Conditions; Calibration Procedure - National Instruments FD-11603 Manual

8-channel, ±10 v voltage input device for fielddaq
Table of Contents

Advertisement

Equipment
Calibrator
SHM125M-Pigtail I/O
Cable (0.3 m, 1 m, 3 m
lengths) (x4)
Banana Plug (optional)
(x2)

Test Conditions

The following setup and environmental conditions are required to ensure the FD-11603 meets
published specifications:
Keep connections to the device as short as possible. Long cables and wires act as
antennas, picking up extra noise that can affect measurements.
Verify that all connections to the device are secure.
Use shielded copper wire for all cable connections to the device. Use twisted-pairs wire
to eliminate noise and thermal offsets.
Maintain an ambient temperature of 23 ±5 °C. The device temperature will be greater
than the ambient temperature.
Keep relative humidity below 80%.
Allow a warm-up time of at least 10 minutes to ensure that the FieldDAQ device
measurement circuitry is at a stable operating temperature.

Calibration Procedure

The calibration process includes the following steps:
1.
Initial Setup
on page 4
2.
Verification
on page 4
3.
Adjustment
on page 7
4.
EEPROM Update
5.
Reverification
Table 2. Recommended Test Equipment
Recommended Model
Fluke 5502A
NI part number
785950-0R3/01/03
on page 8
on page 8
High-precision voltage source with
gain uncertainty of ≤50 ppm and
offset uncertainty of ≤50 μV
Use banana plugs with an opening
large enough to accommodate four
22 AWG signal wires
FD-11603 Calibration Procedure | © National Instruments | 3
Minimum Requirements