Product Specifications
Models: SC-2004-OC, SC-2004-OC-230, SC-2004FC-OC, SC-2008-OC, SC-2008-OC-230
Electrical Rating: 120 VAC, 60 Hz, 0.5 A or 230 VAC, 50 Hz, 0.5 A
Electrical Classification: Class II
Type Applied Part: Type BF
Ingress Protection: IP20
Mains Isolation: Unplug
Mode of Operation: Continuous
Essential Performance: The pump's cyclical inflation and deflation of the garment(s)
Cycle Time (SC-2004-OC): 90 ± 10 seconds (inflation 72 seconds - 18
seconds/garment chamber, deflation 18 seconds)
Cycle Time (SC-2004FC-OC): 60 ± 10 seconds (inflation 50 seconds - 12.5
seconds/garment chamber, deflation 10 seconds)
Cycle Time (SC-2008-OC): 60 ± 10 seconds (inflation 54 seconds - 6.5
seconds/garment chamber, deflation 6 seconds)
Pressure Range: 20-80 mmHg
Precision: 1 mmHg
Accuracy: ± 20%
Warranty: Pump 3 years, garment 1 year
Expected Service Life: 5 years
Regulatory Classification: AU IIa, CA 2, BR II, EU IIa, US 2
Weight (SC-2004-OC, SC-2004FC-OC): 4.65 lbs. (2.1 kg)
Weight (SC-2008-OC): 5.25 lbs. (2.38 kg)
Dimensions: 4.5" x 11.75" x 7.38" (114 mm x 298 mm x 187 mm)