Hitachi EUP B715 Instruction Manual page 33

Biopsy probe
Table of Contents

Advertisement

6) Alternatively to step 5 suspend the submergible part of the probe
in a tray filled with deionized water/tap water for 5 min.
7) Visually check the cavity and the outer surface of the biopsy probe
for cleanness.
check. If there is still soil visible, repeat all above steps.
B) Magnetic sensor attachment and spacer for EZU-RVS2
1) The temperature of the detergent solution should be between 15-30 °C,
concentration is 1.6%. Please note the minimum contact time of the
detergent in the manufacturer's instruction.
detergent is used, follow the manufacturer's instructions. If a
different detergent is used, please also consider the approved
material compatibility for the medical device.
2) Immerge the Magnetic sensor
attachment and the Spacer for
EZU-RV2S into the diluted detergent
(see figure 26). Wipe the magnetic
sensor attachment under the surface
of the detergent solution with a
single-use fluff free soft cloth to
remove all visible soil.
3) Rinse the Magnetic sensor attachment
and the Spacer for EZU-RV2S
thoroughly with high-quality tap
water to remove all debris and residue
of the detergent solution.
4) Alternatively to step 3 suspend the Magnetic sensor attachment and
the spacer for EZU-RV2S in a tray filled with deionized water/tap
water for 5 min.
5) Visually check the outer surface of the Magnetic sensor attachment
and the Spacer for EZU-RV2S for cleanness.
magnifying glass for visually check. If there is still soil visible,
repeat all above steps.
If necessary, use magnifying glass for visually
-27-
If a different
Water,
Detergent or
Disinfectant
Fig.26
Immersion of the
Magnetic sensor
attachment and the
Spacer for EZU-RV2S
If necessary, use
Q1E-EP1245