X-ray equipment if not properly used may cause injury. Accordingly, the instructions herein contained should be thoroughly read and understood by everyone who will use the equipment before you attempt to place this equipment in operation. GE Healthcare, will be glad to assist and cooperate in placing this equipment in use.
Chapter 2: Safety and Regulatory Radiation Protection This device emits x-ray radiation for medical purposes. Each Definium AMX 700 has been tested for leakage radiation and the test results are recorded in the final test protocol. As radiation is harmful to human body, the necessary safety precautions must be observed and the radiation protection regulations of individual countries must be complied with.
In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
Safety First This equipment is to be used by authorized medical personnel only. Your Definium AMX 700 is engineered for years of reliable service. To promote optimum safety, be sure to read this manual carefully, before operating the unit. WARNING: This X-Ray unit may be dangerous to patient and operator, unless safe exposure factors, operating instructions and maintenance schedules are observed.
CAUTION: If the GE Portable DR Detector is used in a fixed installation, it is the User’s responsibility to provide the means for audio and visual communication between the Operator and the patient.
To avoid damage do not place a load in excess of 160kg (350lb) on the detector. In accordance with UL 60601-1, the Definium AMX 700, is classified as Type B Equipment, Class I Equipment, Internally Powered Equipment, Ordinary Equipment and continuous operation.
(p. 3-9) Unit Features Since the Definium AMX 700 is battery operated, there’s no need to plug in your mobile x-ray unit before taking exposures. Battery operation makes the Definium AMX 700 easily adapted to operating, intensive care, and emergency room applications.
Page 27
Chapter 3: Hardware Overview Figure 3-1 Definium AMX 700 back and front Back Front Table 3-1 Definium AMX 700 features Item Description Tube head. Refer to Chapter 4: Moving and Positioning-Position the Tube (p. 4-5) for more information. Collimator controls. Refer to Chapter 4: Moving and Positioning-Adjust the Collimator (p.
Page 28
USB ports (between the drive handle and the detector). Allow you to connect USB accessory devices, such as an external CD drive. There are also two USB ports on the front of the unit. Refer to item 18. GE Portable DR Detector and grid. Refer to Appendix A: GE Portable DR Detector Operating Instructions for more information.
This produces X-rays that are recorded. Release the Prep/ Expose button after the exposure is completed. TechSwitch The TechSwitch option may be purchased for use with the Definium AMX 700. The remote TechSwitch can be used instead of the standard wired hand switch. Definium AMX 700 X-Ray System 5161515-1EN Rev.
The CD drive connects to the unit through a USB port. Plug in the CD drive power supply and insert the connection cable into any USB port. NOTE: The Definium AMX 700 does not have a dedicated space or storage area for the CD drive. It is recommended that the CD drive be disconnected and stored before moving the unit.
(XX represents the code) and contact your service representative. Drive the Unit A compact design and dual motor assembly makes driving and maneuvering the Definium AMX 700 an easy process. Definium AMX 700 X-Ray System 5161515-1EN Rev.
Page 37
Then, simply grip the drive handle (Figure 4-2) and steer the unit to your destination. Figure 4-2 Drive handle WARNING: Movement of the Definium AMX 700 x-ray unit could cause the horizontal arm to slide and strike an operator or patient if not latched properly. Latch telescoping arm properly before moving the unit.
Page 38
(As with any mobile equipment, however, when driving the unit be sure to exercise reasonable care.) With the telescoping arm extended for a patient exam, the Definium AMX 700 pivots on the spot, rotating and maneuvering easily for exact positioning. Again, grip the drive handle and turn the unit to the desired position.
Page 39
When climbing inclines or moving over carpeting or other rough surfaces the drive motors will automatically adjust to the floor surface. When descending inclines or pushing the Definium AMX 700 to make it go faster, it may stop and display the message RELEASE HANDLE. If this happens, release the handle to clear the message and then continue with normal operation.
NOTE: The unit may be driven while the system software is booting up. Position the Tube This section describes how to setup and position mechanical components of the Definium AMX 700. Steps for mechanical positioning are organized in the following way.
Figure 4-7 Lock release mechanism Adjust Column and Telescoping Arm In the following steps you can free the Definium AMX 700 components for positioning in one of two ways: by gripping the collimator handle(s) to keep the lock system released or, whenever necessary, overriding the lock system with slight force.
6. 0° Detent Adjust the Collimator The Definium AMX 700 manual collimator limits patient radiation exposure to a desired area at a given distance from the x-ray tube focal spot. 1. Press the field light button (on either the collimator face or hand switch) to activate the high-intensity field lamp.
Page 43
Manual lock method: Pull out the knob, position the collimator to the desired angle (other than 90° detent), and turn the knob clockwise to secure. (Turn the knob counterclockwise to unlock the collimator and reposition.) Figure 4-11 Collimator rotation knob Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Chapter 5: Taking Exposures Chapter 5: Taking Exposures This chapter describes the procedures for working with the Definium AMX 700 unit hardware when acquiring images. Refer to Appendix A: GE Portable DR Detector Operating Instructions-Chapter A4: Image Acquisition for more information about the system software used to acquire, view, and manage images.
11. Park tube and detector. Technique Adjustment The software provided with the GE Portable DR Detector contains the recommended default technique settings for both digital and cassette exams. Technique adjustment should always be made on the Acquisition screen. Refer to...
Page 46
– 0.4, 0.5, 0.64, 0.8, 1, 1.25, 1.6, 2, 2.5, 3.2, 4, 5, 6.40, 8, 10, 12.5, 16, 20, 25, 32, 40, 50, 64, 80, 100, 125, 160, 200, 250 and 320. Refer to Table 5-1 for maximum technique ratings. Definium AMX 700 X-Ray System 5161515-1EN Rev.
Page 47
25% heated tube assembly (75% heat units remaining), while the actual maximum allowable kVp and mAs ratings are based upon a cold tube assembly. The Definium AMX 700 generator contains a proprietary heat loading algorithm which protects the tube assembly for all heat loading conditions.
Main Power Circuit Breaker (p. 6-6) Overview Battery charging is as easy as driving the Definium AMX 700 to a correctly rated wall outlet and plugging the unit in. Ratings are as follows: Table 6-1 Wall socket rating by model number...
When the Definium AMX 700 capacity gauge drops to 0%, inhibiting x-ray exposures, it is possible to access some emergency capacity to finish an exam by turning the Definium AMX 700 off and then back on. After it has completed its power-up tests, the Definium AMX 700 will provide enough capacity for approximately 45 seconds of usage before x-ray exposures are again inhibited.
Recharge Recommended When battery charge dips below approximately 10%, recharge is recommended. As the Definium AMX 700 is used in its various modes of operation, the number of segments lit will slowly decrease. It’s important to note that you do not have to raise your kVp and mAs value to compensate for low battery capacity.
When recharging the batteries, always locate the unit in a well ventilated area. 1. Drive the Definium AMX 700 to a designated battery charging station or to a correctly rated wall outlet. Position the unit to within 10 feet of the outlet.
Page 52
This same protection scheme can cause the main circuit breaker to trip if the Definium AMX 700 is turned on while still plugged into the AC wall outlet after the “Charge Complete” message is displayed (there may be a delay of 30 - 180 seconds after the unit is turned on before the breaker will trip).
(p. 7-2) Cleaning the Unit We recommend that you clean the Definium AMX 700 regularly with a soft cloth and a mild, non-abrasive cleaner. Apply the cleaner to the cloth, not directly to the unit, to safe-guard against damage to the circuitry.
Chapter 7: Maintenance and Service Qualified Service Available GE Healthcare and its associates maintain a worldwide organization of service personnel specially trained on medical x-ray equipment. Consult your GE representative to find out more about all the available service programs. Maintenance Schedule...
Page 56
Auxiliary items (battery Every year (Refer to GE Portable DR Detector PM Schedule.) charger, HV cables, etc.) Detector calibration Every year (Refer to GE Portable DR Detector PM Schedule.) Visual inspection of detector, Every year (Refer to GE Portable DR Detector PM Schedule.)
Two kinds of messages can appear on the message display on the control cover: status messages and error messages. Status messages simply inform you of the operating condition of the Definium AMX 700, and look something like this. Figure 8-1 Ready message Another type of message, called an error message, may appear should a problem occur with the equipment.
Page 59
MIN Displayed on the alphanumeric display any time an attempt is made to decrease the mAs below 0.40. Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Page 60
The Definium AMX 700 was moving faster than the drive motors allow. Release the Drive Handle to clear the message, and then continue with normal operation. This may happen if you push the Definium AMX 700 too hard when driving down a ramp.
Appendix A: GE Portable DR Detector Operating Instructions Appendix A: GE Portable DR Detector Operating Instructions This appendix details all the software that is provided with the GE Portable DR Detector used with the Definium AMX 700. Topics covered in this appendix include: •...
Detector surfaces have been treated with a finish to allow a smooth and easily cleanable surface. Care shall be taken to protect the surface from scratches. Nameplates and Markings A label on the back of the detector contains the GE part number. A1-3 Definium AMX 700 X-Ray System 5161515-1EN Rev.
Chapter A1: Detector Overview Detector Safety Labels The back of the GE Portable DR Detector displays the following symbols and information: Figure A1-3 Labels on back of detector 1. Attention, consult CAUTION: Do Not Drop. accompanying CAUTION: Device weighs 5.89kg documents.
Chapter A1: Detector Overview Detector Grid The GE Portable DR Detector grid is integrated with a holder that fits the detector exactly. The grid fits over the detector handle and has raised edges to fit around the detector. Once together, you may handle the grid and detector as one unit.
• Do not drop the detector at any time. • Do not prop the device on an edge, against wall or bed. Keep detector in cradle, bucky, or other GE- supplied container. • Do not place other objects or patients on the detector if it is not on a flat surface, as shown in Figure A1-5.
Do not leave disposable wipes or cleaning cloths on the detector or grid for more than 60 seconds. • Let the detector dry at least 60 seconds between cleanings. The following chemicals and products have been tested and approved by GE for cleaning the GE Portable DR Detector, grid, and tether. •...
Chapter A2: General Information Virtual Keyboard When used in a mobile system, the GE Portable DR Detector software works with a touch screen instead of a keyboard and mouse. The GE Portable DR Detector provides a “virtual keyboard” on screen that allows you to enter text and numbers.
Chapter A2: General Information Software Start Up and Shutdown This section describes the procedure for starting up and shutting down the GE Portable DR Detector software only. Start Up Refer to Chapter 7: Maintenance and Service-Periodic Maintenance by Service Personnel (p.
This allows you to update the Worklist or print and send images while the unit is moving or if you are in a location that does not have a cable network port. Please contact your GE Service representative to activate this option.
2. On the Worklist or Image Management screen, press [UTILITIES]. The System – System screen (Figure A2-7) appears. Figure A2-7 System – System screen with Software Reset button A2-8 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Every attempt must be made to assure markers are not located in regions of direct radiation, but are located in regions where some patient attenuation of radiation is present without obstructing the anatomical information of interest. A2-9 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
The message tells you to Continue or to Continue/Call Service (continue and call service). NOTE: To clear the system status area, open the Message Log (described below) and close it again. A2-10 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
This screen allows operators and service personnel to display, review, and analyze system status messages. Figure A2-11 Message Log A2-11 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Worklist can automatically update from the central HIS/RIS database. Other units or workstations can be configured to access these procedures. If the GE Portable DR Detector is used in a mobile unit, the Worklist can only update with new information from the HIS or RIS if the unit is connected to the facility network.
Page 90
Chapter A4: Image Acquisition-Conduct a Cassette Exam (p. A4-10) for more information. network connection status Appears when the Definium AMX 700 is connected to the hospital network. wireless network connection If the wireless network connection is enabled, displays the status status of the connection and signal strength.
For example, if you want to see all the procedures that have a status of “Suspended”, press the “Status” column heading. An arrow appears in the column heading to indicate which column is currently being sorted. A3-7 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Figure A3-4 Worklist filter screen The filter screen has several options for accepting or rejecting the information from the Worklist. Table A3-3 describes the filter screen functions. A3-8 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
The message remains until the refresh process is complete. NOTE: You will not be able to make selections or access Worklist functions while the Worklist is refreshing. Figure A3-7 Auto refresh message A3-11 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
To switch the button between “Delete” and “Delete All”: 1. Press on the small arrow on the right side of the button. A list appears with the available actions. A3-12 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
2. Press [OK] to delete the procedures. The selected procedures are deleted and the Worklist appears. [CANCEL] closes the message, retains the procedures, and takes you back to the Worklist. A3-13 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Discontinued and the information is sent to PACS. [CANCEL] closes the message and returns you to the Worklist without deleting procedures. The procedures are removed from the Worklist. A3-14 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
NOTE: The Patient Information button is unavailable when multiple exams are selected at the same time. Figure A3-9 Add Patient or Patient Information A3-15 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Page 102
Worklist. NOTE: This button does not appear if the Patient Information screen is opened from the Image Viewer or Image Management screens. A3-17 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Press [CANCEL] to close the Add Patient screen without saving changes. NOTE: For a new exam on a existing patient, the patient information cannot be edited. A3-18 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
3. Edit the information as necessary. 4. Press [SAVE] to record the changes and return to the Worklist. Pressing [CANCEL] closes the Patient Information screen without saving changes. A3-19 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Chapter A4: Image Acquisition Chapter A4: Image Acquisition This section details the process of acquiring images using the GE Portable DR Detector or free cassette. Topics covered include: • Overview (p. A4-1) • Re-start Completed or Discontinued Exams (p. A4-6) •...
Page 106
NOTE: When acquiring multiple images or if you need to retake an image, you must switch back to the Acquisition screen in order to record the exposure. A4-2 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Page 109
Recommended Position: Displays the recommended patient position relative to the detector. network connection status Appears when the Definium AMX 700 is connected to the hospital network. wireless network connection If the wireless network connection is enabled, displays the status status of the connection and signal strength.
Page 112
The category and exam name appear in the Selected Protocols list. 4. Press the exam again to de-select it. The category and exam name are removed from the Selected Protocols list. A4-8 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
This section describes the adjustments required when conducting a digital exam. Follow this process to conduct an exam using the GE Portable DR Detector. NOTE: If you need to interrupt the exam and resume it at a later time, press the [SUSPEND] button. You will be returned to the Worklist.
6. Make technique adjustments as necessary: kV and mAs. NOTE: Press [RESET TECHNIQUE] at any time to reset the technique to the default protocol settings. 7. Position the GE Portable DR Detector under the patient. 8. Confirm or adjust the Patient Side, if applicable.
Page 115
Follow this process to conduct a Cassette Exam. NOTE: Make sure that the GE Portable DR Detector is in the detector bin before beginning a cassette exam. Exposures will be inhibited if the detector is placed in the bin after the exam has begun. To clear the inhibit, either: –...
Image Management screen. Any acquired images are stored in a temporary database until they are committed to the permanent storage database upon closure of the exam. A4-13 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
If multiple procedures were selected, the discontinued status applies to all procedures that were open when the exam was discontinued. Press [CANCEL] to close the message and return to the Acquisition screen. A4-14 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
You are able to view single or multiple images at once. When viewing multiple images, an aqua border identifies the currently selected image. Table A5-1 describes the functions for the Image Viewer screen. A5-1 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Page 120
NOTE: When acquiring multiple images or if you need to retake an image, you must switch back to the Acquisition screen in order to record the exposure. A5-2 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Refer to Save Changes to Images (p. A5-27) for more information. network connection status Appears when the Definium AMX 700 is connected to the hospital network. wireless network connection If the wireless network connection is enabled, displays the status status of the connection and signal strength. Refer to Chapter A2: General Information-Wireless Network Connection (p.
Page 126
• Normal - image as acquired • Hard - adjusts the image towards black and white • Soft - adjusts the image towards gray The GE Portable DR Detectoruses a “Smart Windowing” algorithm to optimize image quality. Smart Windowing is an automated, image-based, and technique-independent method of selecting brightness and contrast for image display.
Places a Cobb angle (two lines) on the image that you may re-size, move, or angle. Cobb angle specifications are shown in the User Measurements annotation at the bottom right corner of the image. A5-9 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
• To move: Press and drag the ellipse by its edge (not on a handle). • To change shape or re-size: Press and drag a square handle. A5-13 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Page 132
• To move: Select the marker and drag to the desired area. • To switch between R and L: Press the RL button. NOTE: Only one RL marker is inserted per image. A5-14 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
The shutter appears as a yellow box with red handles (Figure A5-11). Anything outside of the box will be blacked out when the shutter is applied. A5-15 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Page 137
The specific limits and retake rules should be ultimately determined by the appropriate staff at your facility. A5-19 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
45° angle, brightness and contrast will change equally; however, if you move the pointer at a 20° angle (more horizontally) the brightness will change more than the contrast. A5-21 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
3. Drag the image thumbnails from the Images - Raw or Images - Processed panel (Figure A5-20) to the Film Composer screen. Figure A5-20 Images - Processed panel A5-23 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Print Images screen and adjust the settings. If the image placement is acceptable, press [PRINT] to print the image. Figure A5-21 Example of Print Preview with incorrect settings A5-24 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Page 143
The accepted range is between 40% and 90%. If True Size or Fit to Film are selected as the Print Mode, the text box remains disabled. A5-25 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
A message appears: “Would you like to save the changes made to the images?” 3. Press [YES]. [NO] closes the Image Viewer screen without saving changes. A5-27 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Image Management screen. In the Image Details section, each row is an image. Selecting a row makes a small preview image, or “thumbnail,” appear. The image may be opened for viewing, adjustment, or deletion. A6-1 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Page 150
Selects where images are to be copied or saved. The first button switches between the local unit and any configured network hosts. NOTE: If the GE Portable DR Detector is used in a mobile unit, network hosts will only be available when the unit is connected to the network.
8. Open the exams from the Local database. NOTE: If the exams are being viewed on a computer that has the DICOM viewer installed, images may be viewed directly from the CD. A6-7 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Refer to Chapter A7: Set Preferences for information about configuring network hosts. NOTE: If the GE Portable DR Detector is used in a mobile unit, the unit must be connected to the network. A6-9 Definium AMX 700 X-Ray System 5161515-1EN Rev.
Completed. NOTE: You can only copy the images from one exam at a time. NOTE: You cannot copy images to multiple Worklist procedures. A6-10 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
CD. [WRITE] begins the copying process. [CANCEL] closes the screen and returns you to the Image Management screen. A6-12 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
NOTE: The original exam is not modified in any way, only a copy is altered. NOTE: Once an anonymous exam is created, there is no way to recover the patient’s identifying information. A6-14 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
If adding a new connection, press [ADD]. 1. Complete requested information for both tabs (see Table A7-1 and Table A7-2) and press [SAVE] to add the network host or save the changes. A7-3 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Page 165
Table A7-2 Add Network Host – Preferences tab description Function Description Allow this host to query the (GE Allows this host to search and filter the GE Portable DR Portable DR Detector system name) Detectorsystem. Allow this host to retrieve from the...
2. Complete requested information for all tabs (see Table A7-3, Table A7-4, and Table A7-5) and press [SAVE] to add the printer or save the changes. A7-8 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
1. Complete or edit the Worklist Default Query screen (Figure A7-10). Table A7-6 describes the fields in detail. 2. Press [SAVE] to change the Default Query. A7-13 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Page 174
__ Days __ Hours NOTE: If the “Days” value is left blank, the worklist query will be based on current time minus the number of hours entered. A7-14 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Figure A7-11 Drop-down lists on the Add Patient screen 1. Follow the procedures below to Add, Edit, or Remove Preset Names. 2. When finished, press [SAVE] to retain the changes you made. A7-15 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
1. On the Detector Exposure Index screen, press an anatomical category to expand it. Press the category again to close it. 2. Change the lower and upper limits in the text boxes as appropriate for the view. A7-29 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Figure A7-21 Technical Mode checkbox Export DEI Log The DEI log records the exposure levels of images. Medical Physicists in your facility or GE Service Personnel use the log to determine if there is a problem with the detector and to determine what adjustments to make to correct exposure errors.
Image Processing Preferences also allow you to view the settings of default factory looks or to create up to five (5) custom looks. NOTE: Custom looks and the default looks mapping can be backed up by having GE Service personnel perform a system backup. It is recommended that a system backup be performed after any change to the default looks mapping or when custom looks are built.
Page 194
• Move the slider right to make images sharper. • Move the slider left to make images smoother. Range: 1 to 10 (discrete setting) with increments of 1 A7-34 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Page 198
The effect of varying AREA and STRENGTH in TE is demonstrated in Figure A7-28 for under-penetrated regions. Increasing STRENGTH while keeping the AREA constant makes the pixel grayer in the shoulder region. Increasing AREA extends the region that becomes gray. A7-38 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Page 205
– If the anatomy is unpaired, the selections are Head Down or Head Up – If the anatomy is paired, the selections are Digits to Back, Digits to Feet, Digits to Front, Digits to Head. A7-45 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Page 206
13. Select all Patient Sizes you want available for the view. The default size will automatically be selected. 14. Select Receptors and Modes for each selected receptor. 15. Select the Default Receptor - cassette or GE Portable DR Detector. A7-46...
Page 207
The default patient size is shown with an asterisk (*) to the left of the size indicator. 17. Select the patient size to edit. 18. Press [EDIT TECHNIQUE]. A7-47 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Page 212
This chapter explains the Quality Assurance and Maintenance process for the GE Portable DR Detector. To assure continued performance of the GE Portable DR Detector, a periodic inspection program must be established. Daily functional checks should be part of this program.
The QAP process begins when the QAP button is clicked. The button is located at the bottom of the Worklist or Acquisition screen. NOTE: QAP cannot be performed if an exam is open or being reviewed. A8-2 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
2. Select Quality Control Tests from the left side of the screen (Figure A8-3). Figure A8-3 QAP selection from left side of screen The Quality Assurance Program screen appears. A8-3 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Page 215
4. Press [START] to begin the Detector Check. 5. Wait for the test to complete. The Detector Check result screen appears automatically after tests have been completed. A8-4 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Detector Check tests. Figure A8-7 Failed Detector Check results Perform QAP QAP involves tests requiring exposures with the flat field phantom. The GE Portable DR Detector QAP kit includes a flat-field phantom (Figure A8-8) sized to fit the collimator. A8-5...
Page 217
2. Select Quality Control Tests from the left side of the screen (Figure A8-10). Figure A8-10 QAP selection from left side of screen The Quality Assurance Program screen appears. A8-6 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Page 219
The Detector QC Tests screen appears automatically when the Detector Check Tests are complete. 7. Follow the instructions on screen to complete this portion of the test: A8-8 Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) Operator Manual...
Periodic maintenance checks are required thirteen months after installation and every year thereafter. A periodic maintenance schedule and a sample mAs/kVp accuracy chart are provided on the following pages. As the manufacturer of the GE Portable DR Detector, GE accepts no responsibility for safety, reliability or performance unless: •...
On the Service User Interface, click on the Configuration Tab and select the Advanced Options. Click on the Radio Buttons available to Enable/ Disable the HIPAA option. By default the HIPAA option is disabled. The Login feature can be turned on or off by your GE Field Engineer. Definium AMX 700 X-Ray System 5161515-1EN Rev.
Understanding Privileges, Groups, and Users Privileges are the rights to access a system or piece of equipment and perform certain functions. Privileges are assigned to groups. The privileges are created by GE and cannot be changed. They are: • GE Service allows access to all functions for service and maintenance personnel.
Understanding Privileges, Groups, and Users (p. B-2) for definitions of privileges. NOTE: The Service privilege is protected and cannot be selected. Only GE Service personnel can assign a group to the Service privilege. 3. Click on a selected checkbox to remove the privilege.
NOTE: Remember the following when creating user names: – You cannot use the names “operator”, “admin”, or “administrator.” These names are reserved for use by the system and GE Service personnel. – User names cannot have spaces or uppercase (capital) letters. Use all lowercase letters in the username and use underscores to separate words.
There is a class of users identified as protected users. The privileges assigned to these users can only be modified by a GE Service Engineer. Select the Show Protected Users checkbox to see these users. Protected users and groups are shown with a violet (light purple) background.
[APPLY CONFIGURATION] button was last clicked. Configure Enterprise Authorization The Configure-Enterprise Tab is used by the site’s IT (Information Technology) or GE Service personnel. It provides connectivity to the site’s user database. If you do not have a network established in your facility, this tab will not be used.