• Failure to observe the instructions and warnings in this manual invalidates
all performance statements and could result in damage to the unit.
• Water temperature should not exceed 122°F (50°C).
The Decontamination Kit (EVADSK) includes one (1) breathing tube, outlet plug
(PA3PG), one (1) filter inlet plug (EVAICP67), white pad and two (2) gaskets.
PA3PG
Breathing Tube
Outlet Plug
and Gasket
Cleaning
The EVA Powered Air-Purifying Respirator (PAPR) system should be cleaned
and inspected after each use and prior to storage. Detach the battery pack,
breathing tube, and loose-fitting hood from the blower unit. Inspect all parts for
damage or other signs of excessive wear. Replace all damaged parts prior to
storage or next use. Refer to the EVA and EVAHL User Manuals for additional
guidance.
Inspect Prior to Use
Inspect the sealing groove on the upper EVA / EVAHL housing, breathing
tube outlet plug / gasket, and filter inlet plug / gasket. Wipe away debris as
necessary to ensure a secure seal.
Bullard Decontamination is compatible with
EVA (serial no. > 48500) and EVAHL (serial no. > 9000).
Use with serial numbers < 48500 / 9000 will result in water ingress which
will damage the internal components and void the warranty.
www.bullard.com
Decontamination Kit
(for EVA / EVAHL)
�
LIMITATIONS OF USE AND WARNINGS
White Pad
�
WARNING
EVAICP67
Filter Inlet Plug
and gasket