FD1/FD2/FD3
HIGH SENSITIVITY
POCKET DOPPLERS

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the FD1 and is the answer not in the manual?

Questions and answers

Summary of Contents for Huntleigh FD1

  • Page 1 FD1/FD2/FD3 HIGH SENSITIVITY POCKET DOPPLERS...
  • Page 2: Table Of Contents

    Standards Compliance ................16 FHR Performance ................... 16 General ....................17 Environmental ..................17 6. End of Life Disposal ............18 7. Warranty & Service ............19 Dopplex, and Huntleigh are registered trademarks of Huntleigh Technology Ltd. 2004. © Huntleigh Healthcare Ltd. 2004-2014...
  • Page 3: Doppler Measurement Sites And Recommended Probes

    Doppler Measurement Sites and Recommended Probes 1 Jugular Vein 11 Subclavian Artery VP4HS, VP5HS VP4HS, VP5HS 2 Subclavian Vein 12 Brachial Artery VP4HS, VP5HS VP8HS, EZ8 3 Fetus 13 Ulnar Artery OP2HS, OP3HS VP8HS, EZ8 4 Femoral Vein 14 Radial Artery VP4HS, VP5HS VP8HS, EZ8 5 Great Saphenous Vein...
  • Page 4: Safety

    Do not use in the sterile fi eld unless additional barrier precautions are taken. Do Not : • immerse in any liquid, (except FD1/FD3 probe) • use solvent cleaner, • use high temperature sterilising processes (such as autoclaving), •...
  • Page 5: Patient Applied Parts

    Do not dispose of batteries in fi re as this can cause them to explode. Do not attempt to recharge normal dry-cell batteries. They may leak, cause a fi re or even explode. This product contains sensitive electronics, therefore, strong radio frequency fi...
  • Page 6: Introduction

    Unpacking / Preliminary Checks Contents Delivery Inspection Huntleigh takes every precaution to ensure that goods reach you in perfect condition. However, accidental damage can occur in transit and storage. For this reason we recommend that a thorough visual inspection is made immediately the unit is received.
  • Page 7: Product Controls

    Product Controls ● ● Headphone Socket ● RS232 Port ● ● LCD Panel ● ● Loud-speaker ● ● On/Off Button ● ● Probe Holder ● ● Volume Control ● Start/Stop Button ● Mode Button Battery ● ● Compartment ● ● Pocket Clip Product Labelling Patient applied parts (ultrasound probes) are type BF according to the...
  • Page 8 Huntleigh Healthcare Ltd. Manufactured 35 Portmanmoor Road, Cardiff, CF24 5HN, United Kingdom T: +44 (0)29 20485885 sales@huntleigh-diagnostics.co.uk www.huntleigh-diagnostics.com Legal Manufacturer in association with the CE mark in Europe ArjoHuntleigh AB Hans Michelsensgatan 10 211 20 Malmö, Sweden...
  • Page 9: Operation

    Operation Refer to diagram on inside front cover for Doppler Measuring sites and Recommended Probes. To connect the probe, align the arrow on the connector with the slot on the probe and push fi rmly. To disconnect the probe, pull the connector sharply. DO NOT pull the cable. Note: During use, an automatic noise reduction feature operates on low level signals to improve sound quality.
  • Page 10: Obstetric Mode

    A wind-like sound is heard from the placenta. *Note: For FD1/FD3 : Gel is not required when probe is used underwater. Standard Mode - FD1/FD2/FD3 In this mode the FHR, averaged over 4-heart beats, is displayed on the 3-digit...
  • Page 11 Smoothed Mode - FD2 Only This mode is used to obtain more stable heart rate readings. In this mode, FHR is averaged over 8 beats. The LCD displays a solid heart symbol. Manual Mode - FD2 Only This mode is used when a fetal heart beat is audible in the loudspeaker or headphones but, due to noise or a low signal level, the FD2 cannot reliably calculate the heart rate.
  • Page 12: After Use

    After Use Press and release the On/Off button. If you forget to switch the unit off, it will automatically shut-off after 3 minutes. Refer to the cleaning section before storing or using the unit on another patient. Store unit together with probe and accessories in the soft carry case provided.
  • Page 13: Care And Cleaning

    Periodically, and whenever the integrity of the system is in doubt, carry out a check of all functions as described in the relevant section of the IFU. If there are any defects to the housing contact Huntleigh or your distributor for repair or to order a replacement.
  • Page 14: General Cleaning And Disinfecting

    General Cleaning and Disinfecting Always keep the external surfaces clean and free of dirt and fl uids using a clean dry cloth. Wipe any fl uids from the surface of the product using a clean dry cloth. Wipe with a cloth dampened in 70% Isopropyl Alcohol. Completely dry with a clean, dry lint free cloth.
  • Page 15: Maintenance And Repair

    Warning: Sodium Hypochlorite @ 10,000 ppm for disinfecting should only be used in situations described in the High Risk defi nition. Unnecessary use of this concentrated solution for Low and Medium risk situations may result in damage to the product. Do not allow Sodium Hypochlorite solutions to come into contact with metal parts.
  • Page 16: Specifi Cations

    Mode of operation. Continuous Degree of protection Main Unit: IP20 against harmful ingress FD1 / FD3 probes : IPX7 of particles and/or water. All other probes (Tip only): IPX1 Degree of safety of Equipment not suitable for use in the presence of...
  • Page 17: General

    General Max. Audio Output 500mW rms typical (Loudspeaker) Auto shut-off 3 minute no signal 10 minute unconditional Max. output Power: 25 mW rms (32Ω) Headphones Connector: 3.5mm stereo jack socket Max. applied voltage: +9Vdc RS232 Interface Data format: RS232C (FD2 only) Connector: 8pin subminiature DIN socket Max.
  • Page 18: End Of Life Disposal

    End of Life Disposal This symbol signifi es that this product, including its accessories and consumables is subject to the WEEE (Waste Electrical and Electronic Equipment) regulations and should be disposed of responsibly in accordance with local procedures.
  • Page 19: Warranty & Service

    Warranty & Service Huntleigh Healthcare Ltd. standard terms and conditions apply to all sales. A copy is available on request. These contain full details of warranty terms and do not limit the statutory rights of the consumer. Service Returns If for any reason Dopplex unit has to be returned, please: •...
  • Page 20 In the European Union, the user should also report the serious incident to the Competent Authority in the member state where they are located. Manufactured in the UK by Huntleigh Healthcare Ltd on behalf of; ArjoHuntleigh AB Hans Michelsensgatan 10 211 20 Malmö, Sweden...