B)
Magnetic sensor attachment
1) Prepare the disinfectant solution as stated in the procedure for the
probe.
2) Immerse the Magnetic sensor attachment into the disinfectant (see Fig.
8). Set a clock to insure the recommended contact time which is 5
minutes.
3) Rinse the Magnetic sensor attachment with deionized water for 1 minute.
(alternatively: immerse it in a tray filled with deionized water (see
Fig. 8) for 5 min.)
4) Visually check the outer surface of the Magnetic sensor attachment
for leavings of the disinfectant. If necessary, repeat the rinsing.
Water
Detergent
Disinfectant
Fig. 8 Immersion of the probe and the Magnetic sensor attachment
-19-