Hitachi C42 Instruction Manual page 24

Probe
Table of Contents

Advertisement

3) The Magnetic sensor attachment should be left in the detergent
solution according to the specified contact time of the detergent
manufacturer.
4) Rinse the Magnetic sensor attachment with running tap water for 1
minute. (alternatively: immerse it in a tray filled with deionized
water/tap water (see Fig. 8) for 5 min.)
5) Visually check the outer surface of the Magnetic sensor attachment
for cleanness. If necessary, use magnifying glass for visually check.
If there is still soil visible, repeat all above steps.
Manual disinfection:
A) C42
1) Prepare the disinfectant solution in a tank with cold water (please
follow the instructions of the disinfectant manufacturer regarding
application, concentration, microbiological efficiency, service
life and contact time).
2) Confirm the concentration of the disinfectant before immersing the
probe. Although Cidex® OPA does not need to be diluted, it is
recommended to use test strips to verify the concentration. The test
strips can indicate whether or not the concentration is above the
Minimum
Effective
expiration date of the test stripes. Temperature of disinfectant
solution should be minimum 20 °C. The minimum contact time is 5 minutes.
If a different disinfectant is used, follow the manufacturer's
instructions. Please also consider the material compatibility for the
medical device.
3) Immerse the immersible part of the probe into the disinfectant (see
Fig. 8). Set a clock to insure the recommended contact time which is
5 minutes.
4) Rinse the immersible part of the probe with deionized water for 1 minute.
(alternatively: immerse the immersible part of the probe in a tray
filled with deionized water (see Fig. 8) for 5 min.)
5) Visually check the outer surface of the probe for leavings of the
disinfectant. If necessary, repeat the rinsing.