Ecg Check - GE Vivid T8 Basic Service Manual

Hide thumbs Also See for Vivid T8:
Table of Contents

Advertisement

Step
4
Test the probe in each active connector slot.
5
Do a leakage test on the probe,
10-7 'Electrical safety tests' on page 10-19 for
more information.
6
Repeat this procedure for all available probes.

4-3-26 ECG Check

4-3-26-1 Introduction
4-3-26-2 Parts needed
4-3-26-3 Preparations
Vivid T8/Vivid T8 Pro – Basic Service Manual
5718826-100 English Rev. 10
Table 4-1:
Probe and Connectors Checks
Task
See
Not all the types of RS probe connector can match the Vivid T8/
Vivid T8 Pro system.
Figure 4-6. RS type probe connector
1. Not match
2. Match
3. Match
4. Match
The ECG capability on this unit, is intended as use as a trigger
for measurements, but can also be viewed on the screen.
ECG Pads (3 pc)
ECG Cable Europe Kit (5505585) or US Kit (5505577)
None
Functional checks
Expected Results
It will display pictorial data each time.
It passes the test.
4-31