TROUBLESHOOTING/MAINTENANCE
7.0 Troubleshooting/Maintenance
PROBLEM
"OVER TEMP" light
1-
is on.
.
2-
.
3-
.
T/Pump will not heat.
T/Pump will not
pump.
T/Pump will not turn
on.
Water leaks from
The T/Pump must always be connected to a T/Pad or the hose ends must be joined
Clik-Tite connector
together to prevent water leakage. If the connector leaks, check for:
on pump.
1-
.
2-
.
If problem persists, remove pump for servicing.
8.0 Preventive Maintenance/Service
Only qualified medical equipment service
personnel should repair the T/Pump. Improper
repair may result in death or serious injury,
equipment damage, or malfunction.
Perform the following Preventive Maintenance and
Service procedures as described in the T/PUMP SERVICE
MANUAL*:
Oil the pump motor every 6 months.
Perform the FUNCTIONAL CHECK AND SAFETY
INSPECTION every 6 months and after making
repairs.
* Contact your local dealer for a copy of the
T/PUMP SERVICE MANUAL.
8
POSSIBLE CAUSE
Pad or hose is kinked. Hose clamps are
closed.
Water level is low, or reservoir is empty.
T/Pump is filled with water that is too hot
(it must be below 24°C).
T/Pump is tipped.
Water level is low, or reservoir is empty.
Fuse blown.
Damaged O ring.
Locking ring on Clik-Tite connector is not
snapped into place (see fig. 2).
TP472 T/PUMP
REMEDY
Unkink pad or hose. Turn switch off, unplug
pump, and empty water. Open hose clamps.
Refill with room temperature distilled water to
proper level. Plug T/Pump in, turn switch on, and
wait
40 minutes.
Place T/Pump on level surface.
Fill with room temperature distilled water to
proper level.
Remove T/Pump from use. Refer to qualified
medical equipment service personnel.
If defective, replace Clik-Tite connector.
Snap Clik-Tite locking ring shut.