AVERAGE PIECE WEIGHT (APW)
When the MODE button
is released, CLr.PW Pcs is
displayed.
NOTE: If no APW has been previously stored,
the CLr.PW display is skipped and the display
shows PUt 10 Pcs.
CLEARING A STORED APW
Press the YES button to clear the stored APW.
RECALLING A STORED APW
1.
Press the NO button to
recall the existing APW.
2. Press the FUNCTION MODE
button to temporarily
display the APW value.
ESTABLISHING THE AVERAGE PIECE WEIGHT (APW)
The display shows PUt 10 Pcs.
ESTABLISHING A NEW APW
1.
Press the NO button to
increment the sample size.
Choices are 5, 10, 20, 50,
100 and 200.
2. To establish the APW,
place the specified
quantity of samples on
the scale and press the
YES button to capture
the weight.
BEGIN COUNTING
1.
Place the parts on the
scale and read the count.
If a container is used, be
sure to tare the empty
container first.
PAGE 4 OF 6
OPERATION CONTINUED
LEGAL FOR TRADE
When the indicator is used in trade or a legally
controlled application, it must be set up, verified and
sealed in accordance with local weights and measures
regulations. It is the responsibility of the purchaser to
ensure that all pertinent legal requirements are met.
SETTINGS
Before verification and sealing, perform the following steps:
1. Verify that the menu settings meet the local weights
2. Perform a calibration.
3. Set Legal for Trade to ON in the Setup menu.
4. Exit the menu.
5. Disconnect power from the indicator and open the
6. Set the security switch to the ON position.
7.
8. Reconnect power and turn the indicator on.
VERIFICATION
The local weights and measures official or authorized
service agent must perform the verification procedure.
Please contact your local weights and measures office
for further details.
and measures regulations.
housing.
Close the housing.
NOTE: For installations that employ the audit
trail sealing method, steps 5 to 8 are not
required. However, the security switch may be
set to ON to safeguard against unintentional
changes to configuration and calibration
settings.
NOTE: When Legal for Trade is set to ON and the
security switch is set to ON, the following menu
settings cannot be changed: Span Calibration,
Linearity Calibration, GEO, LFT, Calibration
Unit, Capacity, Graduation, Power On Unit, Zero
Range, Auto Zero Tracking, Expanded Mode,
Count Mode, Kilogram Unit, Pound Unit, Gram
Unit, Ounce Unit, Pound Ounce Unit, Stable
Only. To enable editing of these menu settings,
return the security switch to the OFF position
and set LFT menu item to OFF.