Appendix B. Calibration and Testing with PanaView
B.5.3m Initializing the XGM868i to Default Values (cont.)
Note: When you first initialize the XGM868i, the number of LCD parameters is set to OFF. You must program the
LCD to display any measured parameters.
1. In the Edit Functions menu, double-click on the Site Edit Menu.
2. In the Site Edit Menu, double-click on the Global entry, and then on the Input/Output option.
3. From the Input/Output option, highlight and double-click on the Display option in the center pane.
4. The first prompt asks for the # of LCD Parameters. Double-click on the desired number (from OFF through 1-4 and
KEY).
The OFF setting switches the measurement display off, while the KEY setting enables users to change the measurement
display via the arrow keys, without accessing the Keypad Program. If you select KEY:
To view a parameter other than the one currently displayed, press the [] or [] keys to scroll through the various
parameters.
To scroll through the channel options on a two-channel XGM868i, press the [] or [] keys until you have
reached the desired option.
Note: For a 1-channel XGM868i, skip to step 6.
5. Scroll to the desired Channel option, as listed in Table 9.
6. For each channel, select the desired Measurement Parameter, as shown in Table 10 on page 61.
7. Repeat steps 3 and 4 for each parameter. PanaView returns to the Global I/O menu.
60
Table 9: Channel Options
Option
Description
CH1
Channel 1
CH2
Channel 2
SUM
CH1+CH2
DIF
CH1-CH2
AVE
(CH1+CH2)/2