Testing Frequency Outputs - GE DigitalFlo XGM868i Service Manual

Panametrics general-purpose gas flow transmitter (1 & 2 channel)
Table of Contents

Advertisement

B.4.8 Testing Frequency Outputs
Frequency outputs may be added to the Model XGM868i by installing an appropriate option card in
card contains two or four frequency outputs, which are designated A, B, C and D.
Prepare for the testing procedure by connecting a frequency counter to
page 8 for frequency output wiring, and Figure 4 on page 12 to identify the
terminal block J2.
1. In the center pane of the Calibration/Test window, double-click on Slot 1.
2. Double-click on the desired output to open the Frequency Output menu. (
this manual.)
The procedure for testing
IMPORTANT:
reconnect the frequency counter on terminal block
numbers.
3. At the Frequency prompt, enter a frequency (between 1 and 10,000 Hz) and click [Next Item]. Verify that the
frequency counter reads the correct value.
4. Repeat steps 2 and 3 to test all of the available frequency outputs. If any of the frequency outputs fails the test,
contact GE for assistance.
Click [Exit Page] to return to the Slot 1 window and calibrate additional inputs/outputs, or click [Exit Page] again to
close the Calibration/Test menu.
DigitalFlow™ XGM868i Service Manual
is identical to that for testing
Outputs B, C or D
Appendix B. Calibration and Testing with PanaView
of the option card. See Figure 2 on
Output A
and
OUT(+)
RTN(-)
will be used as an example in
Output A
Output A
. See Figure 4 on page 12 for the correct pin
J2
Slot 1
. This option
pins for
on
Output A
. However, be sure to
51