Summary of Contents for Stahlwille MANOSKOP 721 Quick

  • Page 1 Instructions for use MANOSKOP 721, 730 Quick ® Status: 07/2016...
  • Page 2: Table Of Contents

    Checking the accuracy of the cut-out value ..................... 13 Adjusting for deviations in cut-out value ......................15 Cleaning ............................... 16 Accessories..............................17 For all models ..............................17 730 Quick ..................17 ® For STAHLWILLE Service MANOSKOP Services ................................17 Disposal ................................ 17...
  • Page 3: Technical Description

    Technical description Technical description All models MANOSKOP 721 and 730 Quick are adjustable ® torque wrenches with a cut-out, tactile and audible cut-out signals.  These torque wrenches have a safety cut-out mechanism.  The wrench is set to cut-out at a certain torque level by setting the required value on the force- free sliding scale.
  • Page 4: Standard Manoskop 721/5 Quick, 721/15 Quick, 721/20 Quick And 721Qr/20 Quick

    Technical description Standard MANOSKOP 721/30 Quick ® Maximum permissible deviation of the set value from the absolute value at cut-out is ± 4 %. MANOSKOP 721 Quick and 730 Quick comply ® ... has a permanent, switchable ratchet with a push- with DIN EN ISO 6789, Type II, Class A.
  • Page 5: Manoskop 730 Quick

    Technical description MANOSKOP 730 Quick ® 730/5 730a/5 730/10 730a/10 730/12 730a/12 range [N‧m] 6–50 6–50 20–100 20–100 25–130 25–130 [ft‧lb] 5–36 – 15–72,5 – 20–95 – [in‧lb] – 50–440 – 180–880 – 225–1150 insertable square drive [mm] 9×12 9×12 9×12 9×12 14×18...
  • Page 6: Safety Instructions

    Safety instructions Safety instructions Intended Purpose MANOSKOP 721 and 730 Quick have been ® designed for controlled tightening of screw joints in a workshop environment. In order to loosen a nut or bolt during the normal tightening process, the  The MANOSKOP may not be used as a ®...
  • Page 7: Correct Torque Settings

    Operation Correct torque settings ... MANOSKOP may be damaged. After such an ® overload, the readings may be inaccurate in such a ... can be lifesaving in some applications. For this way that the user does not notice. reason, please note the following points: CAUTION Impermissible deviation from the triggering accuracy leads to a risk of...
  • Page 8: Selecting The Inserts And Insert Tools

     Make sure that the permissible load capability of the plug-in tool exceeds the capacity of the torque wrench.  Only manufacture special tools in consultation with STAHLWILLE. CAUTION Unsecured plug-in tools lead to a risk of injury.  Slide the insert over the square drive of the ...
  • Page 9: 721/30 Quick

    Operation Attaching insert tools — 730/5 to 721/30 Quick 730/65 Quick  Check that the square drive is fitted to the right side of the torque wrench.  Insert the insert tool into the internal square drive on the face of the head of the wrench. The spring- ...
  • Page 10: Removing Insert Tools

    Operation setting fail-safe mechanism is released, the value  If you are using a ratchet insert tool, set this to the set is automatically locked. desired direction by turning the control knob. Proceed as follows:  Hold down the pressure plate at the grip end to release the setting lock.
  • Page 11: Controlled Counter Clockwise Tightening

    Operation  Release the pressure plate again. For controlled counter clockwise tightening using the MANOSKOP 721/30 Quick in the turned over ® The set value is now safely locked. position, the square drive must first be pushed through to the upper side. For controlled counter clockwise tightening using the MANOSKOP 730/5 ®...
  • Page 12: Using The Torque Wrench

    Operation Using the torque wrench Pull steadily and without any interruption, particularly during the final phase, until you feel a CAUTION jerk and hear a click. The torque level set on the scale has now been reached. An incorrect trigger value leads to a risk of injury.
  • Page 13: Maintenance

    MANOSKOP yourself. Suitable torque ® the torque tester. testers are available from STAHLWILLE. It is also  Use the following formula to make sure that the possible for STAHLWILLE to test the MANOSKOP ® values displayed by the torque tester do not for you.
  • Page 14 Maintenance  Set the torque wrench to 60 % of the maximum  Set the torque wrench to the highest scale scale reading. reading.  Use this setting to perform five measurements on  Use this setting to perform five measurements on the torque tester.
  • Page 15: Adjusting For Deviations In Cut-Out Value

    You may return your torque wrench to range. STAHLWILLE for adjustment. You will then receive the tool back with a new works calibration certificate. You may adjust the torque wrench yourself. In this case, however, STAHLWILLE‘s accuracy guarantee...
  • Page 16: Cleaning

    Consult STAHLWILLE if this occurs.  Replace the plugs as appropriate to protect the  To adjust the lower end of the range, insert the cut-out mechanism against dirt and damp.
  • Page 17: Accessories

     repairs  testing and readjusting (incl. accuracy guarantee and new works calibration certificate)  Training courses  Service portal: service.stahlwille.de Disposal When the tool finally has to be disposed of, please observe your local environmental protection laws. The handle is made of PPC.