Hitachi MXS2ESLL1 Probe Instruction Manual page 10

Table of Contents

Advertisement

MN1-6403 Rev. 1
Table 2 Name of each part and its explanation
Ultrasonic irradiation part
Probe tip
Deflection portion
Insertion portion
Angle knob for U/D defrection
Angle knob for L/R defrection
Lock knob
Lock lever
Proximal head
Rotation Switch A ,B
Rotation Switch C
Cable
Connector
Ring
Bite block
Caution
Do not pull, bend, twist, or apply excessive force to the cable.
The probe may malfunction due to cable disconnection.
Do not subject the ultrasonic radiation part to hard impact.
The impact may cause damage to the transducer, and that results in noise or no echo in the image.
In most cases, the ultrasonic radiation part itself is not damaged because the part is made of rubber
material.
Do not bend or twist the deflection portion unnecessarily or manually .
This could make the probe unusable.
Keep the angle knob in the free position during an operation of curving.
Operating the probe while locked, it may become unusable.
There is no mark to indicate front direction of probe.
It becomes right axis angle operation, in Continuing pushing rotasion switch and completly turning
diagnostic image
DM# 126288 V1
Name
Ultrasound is radiated from this part. The electronic Sector
transducer is integrated underneath this part.
This contains a temprature sensor controlling surface
temperature and a transducer.
Operating the angle knob allows this part to be bent in any
desired direction.
The probe is inserted into the patient's body up to this section.
This part is used to UP/DOWN operate the deflection portion.
This part is used to LEFT/RIGHT operate the deflection portion.
This part makes the angle knob for L/R defrection fixed any
direction.
This part makes the angle knob for U/D defrection fixed any
direction.
When using the probe, hold this part by hand.
pushing this switches makes rotation of the diagnostic image.
pushing this switches makes rotation of the diagnostic image.
Cable transfers electric input/output signals.
The connector is the part which is connected to the ultrasound
diagnostic instrument.
This part is used for hanging the proximal head
This is put in patient's mouth so that keep the path of the probe
inserted into the body
. See section 4-7.
Explanation
-6-