Installation Instructions ORTHOPHOS XG 5 / Ceph
Testing the system version
All units ordered with a cephalometer are preconfigured for
this version at the factory. You should, however, check the
unit configuration anyway.
To do this, select one of the cephalometer programs on
the Multipad.
If the cephalometer mode cannot be selected, check
the unit configuration (see page 149).
IMPORTANT
if the cephalometer must be retrofitted, then the cephalo-
meter mode must be configured in service routine S017.2
(see page 149).
Other settings
The unit is configured on delivery so that an acoustic signal
sounds at the end of an exposure.
If the customer prefers another configuration, you can
deactivate this acoustic signal with service routine S017.15
(see p. 151).
60 04 902 D 3352
D 3352.031.02.19.02
8 Initial startup
8.3 Switching the units ON