4.1 Point of use (Pre-cleaning)

Pre-cleaning should be done immediately after each
use. The procedure is as follows:
A) C22P probe
1) Remove the protective cover.
2) Clean the probe of all patient's blood or fluid with running tap
water until the surface of the probe looks visually clean.
3) Wipe the whole surface of the probe with gauze pad and remove
superficial visible impurities.
B) Attachment
1) Remove the Attachment and the magnetic sensor from the probe.
2) Immerse the Attachment in sufficient amount of high quality tap
water. Scrub it using soft cloth to remove all visible soil and
from its surface.

4.2 Containment and transportation

Putting the contaminated equipment into exclusive
shock and damage proof container for transportation
is recommended. It is recommended that instruments are reprocessed as
soon as possible and not later than 4 hours after usage.
4.3 Manual Cleaning and disinfection
Prepare
following
cleaning and disinfection:
a) Detergent: Cidezyme® (Johnson & Johnson,
#2258) or another cleaning agent with
approved material compatibility for this
medical device.
b) Disinfectant:
Johnson, # 20391) or another disinfectant
with approved material compatibility for
this medical device.
c) Two tanks, one for cleaning and one for
disinfection - optional:
1
additional
deionized/tap water (sufficient size for
immersion of the submergible part of the probe at full length)
d) Soft, fluff free cloth or single use towel
e) Personal protective equipment (gloves, water repellent protective
skirt, face protection mask or protective glasses, see also
instructions of the manufacturer for the detergent and the
disinfectant)
Manual Cleaning:
items
before
Cidex®
OPA
(Johnson
tank
for
rinsing
manual
Rinsing after manual
&
with
-23-
Point of use
(Pre-cleaning)
Containment and
transportation
Manual cleaning and
disinfection
Manual Cleaning
cleaning
Manual Disinfection
Rinsing after manual
disinfection
Q1E-EP1457

Advertisement

Table of Contents
loading

Table of Contents