Checking the Supplied Options

The instrument can be equipped with both hardware and firmware options. To
check whether the installed options correspond to the options indicated on the
delivery note, proceed as follows.
1. Press the [SETUP] key.
2. Press the "System Config" softkey.
3. Switch to the "Versions + Options" tab in the "System Configuration" dialog
box.
A list with hardware and firmware information is displayed.
4. Check the availability of the hardware options as indicated in the delivery
note.
4.13

Performing a Self-Alignment

When strong temperature changes occur in the environment of the R&S FSW, or
after updating the firmware, you have to perform a self-alignment to align the data
to a reference source.
During self-alignment, do not connect a signal to the RF input connector. Running
a self-alignment with a signal connected to the RF input can lead to false mea-
surement results.

Performing a self-alignment

Before performing this functional test, make sure that the instrument has reached
its operating temperature (for details, refer to the data sheet).
A message in the status bar ( "Instrument warming up..." ) indicates that the oper-
ating temperature has not yet been reached.
Depending on the installation settings, an automatic self-alignment is performed
each time the instrument is switched on. A dialog is displayed indicating how
much warm-up time is still required before self-alignment can be performed.
1. Press the [Setup] key.
2. Press the "Alignment" softkey.
Getting Started 1338.4102.02 ─ 08
Preparing for Use
Performing a Self-Alignment
30