CO
filter check before ventilation
2
NOTE
Before the CO
filter check, you need to have finished
2
a successful CO
zero calibration. Otherwise the CO
2
filter check may be outside of the tolerance range.
1
Remove the CO
sensor from the cuvette.
2
A
B
2
Attach the CO
sensor (A) to the test filter (B).
2
3
Select Filter check.
4
Confirm. The filter check starts and the screen dis-
plays Filter check in progress. After a successful
filter check, the line briefly displays Filter check
OK.
5
Press Alarm Reset to exit.
6
Attach the CO
sensor back to the cuvette.
2
Selecting the ventilation mode
C
Press the appropriate ventilation mode key (A) for
approximately 3 seconds.
Or
Pocket Guide Oxylog 3000 plus
1
2
The selected ventilation mode will be activated.
The active ventilation mode is displayed in the upper left
2
corner of the display (C).
Setting ventilation parameters
If the changed settings are not confirmed after
5 seconds, the alarm ! Confirm settings appears. If the
settings are still not confirmed after 10 seconds, the
alarm ! Settings not confirmed appears. After that the
former settings are restored.
When the PEEP setting is increased above 10 mbar, a
message Confirm PEEP above 10 mbar? will appear
to request confirmation of the change. The PEEP set-
ting can be increased to the desired setting after the
message is confirmed with the rotary knob.
The device can be configured to show Ti or I:E as a pri-
mary parameter that can be set. If Ti is configured as the
primary parameter, I:E will be shown in the information
window when Ti is selected, and vice versa.
Display operating controls
Oxylog 3000 plus
A
B
Press the appropriate ventilation mode key (A).
Press the rotary knob (B) to confirm.
Set the required control knob below the display.
Or
Select, set and confirm a parameter on the display
with the rotary knob.
B
C
Getting started
Oxylog 3000 plus
D
E
A
19