Negative pressure wound therapy
Hide thumbs Also See for RENASYS TOUCH:

Advertisement

Service Manual
REF 66801280/66801281
*smith&nephew
RENASYS™ TOUCH
Negative Pressure Wound Therap
y

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the RENASYS TOUCH and is the answer not in the manual?

Questions and answers

Summary of Contents for smith&nephew RENASYS TOUCH

  • Page 1 *smith&nephew RENASYS™ TOUCH Negative Pressure Wound Therap Service Manual REF 66801280/66801281...
  • Page 3: Table Of Contents

    Device failure RENASYS TOUCH features Repair and returning device Annual service and maintenance Device disposal • Device cleaning RENASYS TOUCH service and maintenance sample checklist • Device physical appearance check LIMITED WARRANTY • Functionality and alarm checks Customer assistance . Functionality check Specifications .
  • Page 4: Purpose

    This Service Manual provides guidance on regular The following equipment is recommended to support service and maintenance checks to be conducted by service and maintenance of RENASYS TOUCH devices. trained personnel on the RENASYS™ TOUCH device Availability and usage is at discretion of each facility.
  • Page 5: Renasys Touch Features

    RENASYS™ TOUCH features Front view of device Left view of device and power Power cord inlet Touchscreen Status indicator AC cord External power indicator Power jack Power supply DC cord Power Start/Pause Lock/Unlock Therapy Home screen Underside of device Therapy set point Inlet port Date and Battery indicator...
  • Page 6: Annual Service And Maintenance

    6. Check that inlet port and air exhaust outlet are free Annual service and maintenance of obstructions, dust or foreign materials. Device cleaning 7. Apply germicidal/antibacterial product to a cotton Recommended personnel: Device cleaning should be swab and clean both inlet port and air exhaust outlet. performed by a trained technician.
  • Page 7: Device Physical Appearance Check

    Device physical appearance check 12. Confirm inlet port is not damaged and device O-ring Recommended personnel: Device physical appearance is present. check should be performed by a trained technician. 13. Confirm bar code label is present and legible on Equipment and materials required: None. bottom of device.
  • Page 8: Functionality And Alarm Checks

    Equipment and materials required: 7. Select device information and locate software RENASYS™ TOUCH canister, RENASYS TOUCH power versions. supply, RENASYS power cord, RENASYS quick click adapter and stopwatch. 8. Record Serial Number, Operating System, Main...
  • Page 9: Operation Check

    Operation check 7. Press and hold Lock/Unlock button for 2 seconds. 1. Connect RENASYS™ TOUCH canister to device by gently pushing canister over device inlet port and engaging both canister clips. 2. Connect canister tubing quick click connector to quick click adapter included with the RENASYS Service Pack (Ref 66021812) and engage tethered cap into open connector of quick click adapter.
  • Page 10: Blockage Alarm Check

    Start Equipment and materials required: Gloves, eye stopwatch. protection, germicidal/antibacterial product, cotton swab, RENASYS TOUCH O-ring, RENASYS TOUCH odor 5. Confirm that within approximately 3 minutes filter, #1 Pozidriv® screwdriver / 5/32" (4mm) slotted touchscreen displays “WARNING BLOCKAGE”, screwdriver, plastic o-ring removal tool.
  • Page 11: Device Vacuum Checks

    13. Press Start/Pause Therapy button to stop device operation. 2. Connect RENASYS TOUCH canister to device by Note: The device pump may run continuously during gently pushing canister over device inlet port and this test (as determined by a quiet “putt-putt” sound).
  • Page 12: Device Failure

    Device failure Device disposal In the event device fails any test, has sustained obvious WEEE Directive [2002/96/EC] on Waste Electrical and damage, has a missing or illegible device label, or gives Electronic Equipment. This statement only applies to any indication product may not be fit for its purpose, the European countries with regard to Waste Electrical and following actions must be taken.
  • Page 13: Renasys Touch Service And Maintenance Sample Checklist

    RENASYS™ TOUCH service and maintenance sample checklist Note: In the event that a device fails any test, has sustained obvious damage, has a missing or illegible label, or has any indication it may not be fit for intended use, please return the unit to Smith & Nephew Service and Repair Center. Service and maintenance sample check list UNIT SERIAL # Steps Yes No...
  • Page 14 Bootloader Version, RENASYS™ TOUCH System Version, Main Application Battery Remaining, Device Lifetime Therapy and Motor Controller Maintenance Due. Bootloader System Version RENASYS TOUCH System Version Battery Remaining Device Lifetime Therapy Maintenance Due Device is set to Clinician mode. Operation check...
  • Page 15 Steps Yes No Comments Restore presets Alarm volume: High Compression rate: High Continuous therapy: 80mmHg Intermittent therapy: 80mmHg for 5 min 0mmHg for 2 min Total Patient Therapy: 0 Component replacement and vacuum check O-ring replaced with new o-ring. Odor filter replaced with new odor filter. The vacuum measured by the pressure gauge is 25 ±...
  • Page 16: Limited Warranty

    CUSTOMER’S SOLE REMEDY, AND SMITH & NEPHEW’S LIMITED WARRANTY SOLE LIABILITY, FOR ANY CLAIM WILL BE THE LIMITATION OF REMEDIES/LIABILITY: REPAIR OR REPLACEMENT BY SMITH & NEPHEW AS The Smith & Nephew negative pressure wound care PROVIDED FOR IN THIS WARRANTY. EXCEPT FOR THIS electro-mechanical device (“Device”) is warranted LIMITED LIABILITY OF SMITH &...
  • Page 17: Specifications

    Specifications (REF: 66801280/66801281) Vacuum Continuous Therapy Levels 25, 40, 50, 60, 70, 80, 90, 100, 120, 140, 160, 180, 200mmHg Intermittent Therapy Levels High: 25, 40, 50, 60, 70, 80, 90, 100, 120, 140, 160, 180, 200mmHg Low: 0, 25, 40, 50, 60, 70, 80, 90, 100, 120, 140, 160, 180mmHg Intermittent Therapy Cycle Times High: 3, 5, 8, 10 minutes Low: 2, 3, 5, 8, 10 minutes Alarms...
  • Page 18 Sound Level Normal Operations no alarms: <43.7dB Battery Operating Time ~ 10-16 hours (therapy) when operating from 25mmHg to 120 mmHg; ~ 8 hours (therapy) at 200mmHg Type Lithium ion Compliance UL 2054 / IEC 62133 Safety Protection Protection Against Electric Device internally powered;...
  • Page 19: Replacement Parts And Accessories

    Continental Europe Power Cord 66800291 South Africa and India Power Cord 66800302 Denmark Power Cord 66800196 RENASYS TOUCH Clinician User Manual (US only) 66020801 RENASYS TOUCH 300ml Canister with solidifier 66801273 RENASYS TOUCH 800ml Canister with solidifier 66801274 RENASYS TOUCH 300ml Canister without solidifier...
  • Page 20: Caution Statements

    Caution statements In order to ensure safe and proper performance, the following conditions must be met: 1. All assembly, operation, adjustment, maintenance and/or repair should be carried out by qualified personnel authorized by Smith & Nephew. 2. No modification of this equipment is allowed. 3.
  • Page 21: Icon And Symbol Glossary

    Icon and symbol glossary Battery indicator • Battery full/charged: green Operation buttons • Battery charging: lightning bolt Power Press and hold for 2 seconds to • Battery low: yellow power on or off device. • Battery critically low: empty with red line Start/Pause Therapy Press to begin •...
  • Page 22 Menu navigation Alarm screens Settings Alarm indicator Press to access the settings menu Press to return to alarm screen you navigated away from Therapy log Return to Alarm Screen Press to access the log screen Press to return to alarm screen you Help navigated away from Press to access the Help menu...
  • Page 23 Direct current Keep dry Single use: do not reuse 40°C/ Storage temperature 104°F ETL listing mark Batch code limit 5°C/ 41°F Manufacturer Date of manufacture Serial number Refer to instruction EU: WEEE symbol not Product catalogue manual/booklet for general waste number Caution refer to Relative humidity...
  • Page 24 Smith & Nephew Medical Ltd. 101 Hessle Rd, Hull HU3 2BN England www.smith-nephew.com www.globalwoundacademy.com only ©2019 Smith & Nephew 66027646 ™Trademark of Smith & Nephew Certain marks Reg. US Pat. & TM Off. This product may be covered by one or more US patents 22392 See www.smith-nephew.com/patents 12/2019...