RENASYS TOUCH
(refer to the User Manual for full details on RENASYS TOUCH)

1. Switching the device between Continuous and Intermittent therapy

- To switch between Continuous and Intermittent therapy, ensure therapy is paused and press the therapy mode toggle on the home screen
Therapy mode toggle
Continuous therapy home screen
2. Starting therapy (Continuous)
Step 1:
Ensure that the device is powered on and
the Continuous therapy home screen is displayed
Step 2:
Use the + or – buttons to select prescribed
pressure setting
Step 3:
Press the Start/Pause therapy button to
start delivering therapy
When therapy is active the status indicator on top
of the device illuminates green and the therapy
indicator at the top of the screen will rotate orange
3. Starting therapy (Intermittent)
Step 1:
Ensure that the device is powered on and
the Intermittent therapy home screen is displayed
Step 2:
Change the High and Low pressure
settings and cycle times by selecting the setting
and use the + or – to the prescribed pressure/
time
Step 3:
Press the Start/Pause therapy button to
start delivering therapy
When therapy is active the status indicator on top
of the device illuminates green and the therapy
indicator at the top of the screen will rotate orange

Quick Reference

Intermittent therapy home screen
Therapy set point
Decrease value
Therapy mode toggle
All 4 settings can be adjusted within the following pressure ranges:
High Therapy
25, 40, 50, 60, 70, 80, 90, 100, 120, 140, 160, 180, 200mmHg
High Cycle times:
3, 5, 8, 10 minutes
Low Therapy
0, 25, 40, 50, 60, 70, 80, 90, 100, 120, 140, 160, 180mmHg
Low Cycle times:
2, 3, 5, 8, 10 minutes
Increase value
Status indicator
Start/Pause therapy
Status indicator

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the RENASYS TOUCH and is the answer not in the manual?

Questions and answers

Alan
February 17, 2025

Smith and nephew renasys touch machine Green Light stays on, but the screen is blank and power button doesn't seem to work or power off

Summary of Contents for smith&nephew RENASYS TOUCH

  • Page 1 – Quick Reference (refer to the User Manual for full details on RENASYS TOUCH) 1. Switching the device between Continuous and Intermittent therapy - To switch between Continuous and Intermittent therapy, ensure therapy is paused and press the therapy mode toggle on the home screen...
  • Page 2 4. Checking for a seal Follow on screen seal checking instructions. Once seal is achieved the screen will display “seal achieved” and then display “delivering therapy” Look for “raisin-like” appearance Listen for a “hissing” sound, indicating a leak Feel the dressing, which should be hard to the touch Possible site of leak • Wrinkle or crease • S kin fold or crevice...

Table of Contents