General setup
Protecting settings
After saving a password, you can protect the individual
functions of the USM 36. The access to these functions
is then only possible as Expert after entering the pass-
word.
– Switch to the second operating level.
– Select the function group CONFIG4.
– Select the function FUN-SELECT.
– Enter the password and press one of the selector
keys. The list of functions is displayed.
– Use the arrow keys to scroll through the list of func-
tions and use the Right arrow key to set the functions
that you want to protect to OFF.
– Press one of the selector keys to complete the pro-
cessing of the list of functions. The settings are
saved.
USM 36
– Switch to the function MODE.
– Use the navigation to change the user from
EXPERT to INSPECTOR in order to activate the
password protection.
– Switch to the first operating level.
– Try to use a protected function.
It should then not be possible to use protected functions.
In these cases, the message
NOT AVAILABLE FOR OPERATOR.
is displayed at the bottom of the display screen.
To be able to use the protected functions, you have to
be logged in as Expert.
Issue 2 (12/2013)
5 Operation
5-91