8. Error Messages
If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. „ERR 3" is displayed.
Error
Designation
Potential cause and remedy
„ERR 1"
Signal too weak
The pulse signals on the cuff are too weak. Re-
position the cuff and repeat the measurement.*
„ERR 2"
Error signal
During the measurement, error signals were
detected by the cuff, caused for instance by move-
ment or muscle tension. Repeat the measurement,
keeping your arm still.
„ERR 3"
No pressure in
An adequate pressure cannot be generated in the
the cuff
cuff. A leak may have occurred. Check that the
cuff is correctly connected and is not too loose.
Replace the batteries if necessary. Repeat the
measurement.
„ERR 5"
Abnormal result
The measuring signals are inaccurate and no result
can therefore be displayed. Read through the
checklist for performing reliable measurements
and then repeat the measurement.*
Error
Designation
Potential cause and remedy
„HI"
Pulse or cuff
The pressure in the cuff is too high (over 300
pressure too high
mmHg) OR the pulse is too high (over 200 beats
per minute). Relax for 5 minutes and repeat the
measurement.*
„LO"
Pulse too low
The pulse is too low (less than 40 beats per
minute). Repeat the mesurement.*
* Please consult your doctor if this or any other problem occurs
repeatedly.
If you think the results are unusual, please read through the informa-
tion in „section 1." carefully.