SLICE6 AIR DAS User’s Manual February 2020 Table of Contents DTS Support ........................4 Introducing the SLICE6 AIR DAS ..................5 Sensor Interface ......................... 6 Supported Sensor Types ....................6 Input Range ........................6 Excitation Sources ......................7 Bridge Completion ......................7 Hardware Filters .......................
Page 3
SLICE6 AIR DAS User’s Manual February 2020 Circular Buffer Mode ....................14 Recorder Mode ......................14 Hybrid Recorder Mode ....................15 Continuous Recorder Mode ..................15 Data Streaming ......................15 Start Record and Event Initiation ................... 15 Appendix A: Hardware Specifications ................17 Mechanical Specifications, Connector Information and Pin Assignments ......
DTS has support engineers worldwide with extensive product knowledge and crash test experience to help via telephone, e-mail or on-site visits. The best way to contact a DTS support engineer is to submit a request through the DTS Help Center portal (support.dtsweb.com).
February 2020 Introducing the SLICE6 AIR DAS SWaP optimized and IP65 rated, the SLICE6 AIR DAS is a complete data acquisition unit for measuring analog signals in extreme environments such as payload ejection/deployment; in- flight/on-board UAVs/drones, rockets, missiles and munitions; and biomechanics. PTPv2 Ethernet communications and real-time data streaming up to 20,000 sps per channel with on-board data storage to flash memory are supported.
SLICE6 AIR DAS Sensor Interface Supported Sensor Types The SLICE6 AIR DAS supports many types of sensors including accelerometers, load cells and pressure sensors. The following general sensor types are supported: • Full- bridge (4-wire) or half-bridge (3-wire) resistive and piezo-resistive sensors •...
Shunt Emulation The SLICE6 AIR DAS contains a shunt emulation circuit, effectively eliminating the need for conventional shunt resistors to perform shunt checks. When “Emulation” is chosen as the shunt calibration method, the software injects a precisely-calculated current into the sensor to create an expected deflection of the sensor's output.
<2.7 W <3.3 W The SLICE6 AIR DAS does not contain an internal battery and must be connected to external power at all times for operation. Without external power applied, the SLICE6 AIR DAS is in a power off state.
February 2020 Power-up and Power-down Procedures When sufficient power is applied, the SLICE6 AIR DAS will power up (on, idle and communi- cation enabled) if an ON signal is present. With power applied but the ON signal absent, the unit is off. Power up (On state) occurs in ~15 s, after which communication is enabled.
SLICE6 AIR DAS User’s Manual February 2020 Power (PWR) LED The power LED is red, green or blue. Action Result Power up Connected to host Power up; not connected to host Power fault (input power out of range) • At power up, the LED cycles from red to green to blue immediately after the status LED has completed its boot-up sequence.
February 2020 Data Memory Size With 15 GB of flash memory available for data storage, the SLICE6 AIR DAS can record ~52 minutes of data at the maximum sampling rate (6 channels at 400,000 sps). Since the recording capacity is very large, it is generally best to limit sampling rates and event durations to the minimum necessary to avoid large and cumbersome data files.
<1 ohm. Thermal Considerations The SLICE6 AIR DAS is a low power device and it is unlikely that self-heating will be an issue in real-world testing if proper mounting methods are observed. Never mount the unit to a thermally non-conductive surface like wood or plastic.
SLICE6 AIR DAS User’s Manual February 2020 Environmental Rating The SLICE6 AIR DAS is IP65 rated: 6 (solid ingress) = totally protected against dust; 5 (liquid ingress) = protected against low pressure water jets from any direction. Care should be taken to prevent prolonged exposure to any potentially harmful environment.
The discussion below provides a general introduction to data collection. Please see the software manual for a detailed discussion and implementation specifics. The SLICE6 AIR DAS is a standalone data logger. Once the system is armed, the PC can be disconnected if desired.
The SLICE6 AIR DAS can be placed in an auto-arm mode that will cause the unit to arm automatically when the power is cycled. This available with any available data collection mode.
Page 16
Level trigger is NOT recommended when SLICE6 AIR DAS is used for destructive testing. Finally, if the SLICE6 AIR DAS remains connected to the PC during data collection, the control software can be used to initiate data collection. The table below summarizes the data collection modes and event/triggering options.
All copyright, patent, and ownership rights are retained. This information shall not be disclosed, Manufacture/fabricate to meet the EU RoHS Directive reproduced in whole or in part, or used for manufacture without prior written consent from DTS. 2011/65/EU and RoHS Annex II phthalates...
SLICE6 AIR DAS User’s Manual February 2020 Ethernet Chaining To share Ethernet communications, chain SLICE6 AIR DAS modules together using the methodology in the table below. SLICE6 AIR DAS #1 SLICE6 AIR DAS #2 SLICE6 AIR DAS #3 Function Function...
SLICE6 AIR DAS User’s Manual February 2020 Appendix B: Hardware Configuration Specifications SLICE6 AIR DAS are typically delivered with a default IP address as follows: 192.168.4.xx where xx is based on the last two digits of the S/N; for example: IP address S/N S6A0047 = 192.168.4.47...
Page 20
SLICE6 AIR DAS User’s Manual February 2020 2. The software will immediately look for all attached devices and list them in the table. (You may also click to refresh the list.) Current IP address for any selected device will cause the unit’s LED to flash.
Page 21
SLICE6 AIR DAS User’s Manual February 2020 4. To enable DHCP, select the check box then select . Proceed to step 7. 5. To disable DHCP and manually enter IP address and other information, unselect the check box. 6. Enter the new parameters and select for each item updated.
Page 31
SLICE6 AIR DAS User’s Manual February 2020 Revision History Date Description 6 Feb 2020 Changed max input voltage from 36 to 30 VDC and updated consumption data. Replaced PPS with 1 PPS. Added 50 kHz spec to fixed hardware filter. Updated STS green LED to include data streaming.