Confirming The External Control Input; Confirming The Standard Serial Output - A&D AD4329A-DLC Instruction Manual

Table of Contents

Advertisement

12.3. Confirming the external control input

By short-circuiting between each input terminal and COM,
the corresponding "o" to switched to "1".

12.4. Confirming the standard serial output

Press [PRINT] to transmit ST,G,+00000.0kg<CR><LF>.
Communication specifications depend on settings for F-30 to F-34.
AD4329A-DLC
EXT1
EXT2
EXT3
EXT4
EXT5
EXT6
Page 46
CHc1ext
oooooo
ooooo1
oooo1o
ooo1oo
oo1ooo
o1oooo
1ooooo
CHc1Cl
Cl
Cl 5end
2 seconds later