Configuration
Calibrate Flow Sensor
Press the »Calibrate Flow Sensor« key on the
Standby screen. The Standby soft key window
is replaced by instructions for performing the
calibration.
Follow the instructions on the screen:
– Close all fresh gas control valves.
– Remove expiratory hose from breathing sys-
tem.
– To start Flow Sensor Calibration, press rotary
knob.
When the calibration begins, the instructions
are removed, and a »Flow Calibration in
progress« message is displayed above the
Standby soft keys.
When the calibration is completed, one of two
messages is posted above the Standby soft
keys: »Flow Calibration completed - recon-
nect expiratory hose« or »Flow Calibration
Failed«.
Troubleshooting Flow Calibration Failure
If the flow sensor cannot be calibrated, retry the cal-
ibration.
If the flow sensor still cannot be calibrated, call
DrägerService or your local authorized service
organization.
134
Instructions for Use Fabius MRI SW 3.n
Need help?
Do you have a question about the Fabius and is the answer not in the manual?
Questions and answers