YOKOGAWA vigilantplant EXA ISC202G User Manual

YOKOGAWA vigilantplant EXA ISC202G User Manual

2-wire inductive conductivity transmitter
Table of Contents

Advertisement

User's
Manual
Model ISC202G [Style: S2], ISC202S [Style: S3]
2-wire Inductive Conductivity
Transmitter
IM 12D06A03-01E
IM 12D06A03-01E
7th Edition

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the vigilantplant EXA ISC202G and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for YOKOGAWA vigilantplant EXA ISC202G

  • Page 1 User’s Manual Model ISC202G [Style: S2], ISC202S [Style: S3] 2-wire Inductive Conductivity Transmitter IM 12D06A03-01E IM 12D06A03-01E 7th Edition...
  • Page 2: Table Of Contents

    4-2. Explanation of operating keys ................. 4-2 4-3. Setting passcodes ..................4-3 4-3-1. Passcode protection ......................4-3 4-4. Display examples .................... 4-3 4-5. Display functions ..................... 4-4 IM 12D06A03-01E 7th Edition: Oct. 2009(YK) All Rights Reserved, Copyright © 2001, Yokogawa Electric Corporation IM 12D06A03-01E...
  • Page 3 5. Parameter setting ....................5-1 5-1. Maintenance mode ..................5-1 5-1-1. Introduction ........................5-1 5-1-2. Manual calibration to determine the cell constant (C.C.) ..........5-2 5-1-3. Second Line display. Referring to the first compensated conductivity......5-3 5-1-4. Second Line display. Referring to the second compensated conductivity......5-4 5-1-5.
  • Page 4 12. APPENDIX 2 ...................... 2-1 12-1. Preface ....................... 2-1 12-2. Wiring diagrams .................... 2-1 12-3. Peripheral products ..................2-2 12-3-1. PH201G*B Dedicated Distributor .................2-2 12-3-2. BA20 Junction Terminal Box ..................2-3 12-3-3. WF10J Extension Cable ....................2-4 12-4. Quick reference for parameter setting ............2-5 12-4-1.
  • Page 5: Preface

    Although the instrument has a weatherproof construction, the transmitter can be harmed if it Yokogawa is not responsible for damage to the becomes submerged in water or becomes exces- instrument, poor performance of the instrument sively wet.
  • Page 6 The following safety symbols are used on the Warranty and service Yokogawa products and parts are guaranteed product as well as in this manual. free from defects in workmanship and material DANGER under normal use and service for a period of...
  • Page 7 Yokogawa le plus proche. tedesco e francese. Se si desidera ricevere i man- uali operativi di prodotti Ex in lingua locale, met- tersi in contatto con l’ufficio Yokogawa più vicino o con un rappresentante. Alle Betriebsanleitungen für ATEX Ex bezogene Produkte stehen in den Sprachen Englisch, Deutsch und Französisch zur Verfügung.
  • Page 8 IM 12D06A03-01E...
  • Page 9: Introduction And General Description

    This manual also includes a basic troubleshooting guide to answer typi- cal user questions. Yokogawa can not be responsible for the performance of the EXA transmitter if these instructions are not followed. 1-1. Instrument check Upon delivery, unpack the instrument carefully and inspect it to ensure that it was not damaged during shipment.
  • Page 10 1-2 Introduction FISCO field device ISC TRANSMITTER ISC TRANSMITTER IECEx KEM 07.0028X Zone 0 Ex ia IIC T4 MODEL ISC202S-P MODEL ISC202S-F IP65 SUFFIX SUFFIX SEE CONTROL DRAWING KEMA 07ATEX0052 X Ex ia IIC T4 II 1G SEE CONTROL DRAWING SUPPLY FISCO FISCO...
  • Page 11: Application

    For best results, read this manu- al in conjunction with the corresponding sensor user’s manual. Yokogawa designed and built the EXA to meet the CE regulatory standards. To assure the user of continued accurate performance in even the most demanding industrial installations.
  • Page 12 1-4 Introduction IM 12D06A03-01E...
  • Page 13: General Specifications

    - Temperature : ± 0.3ºC (0.6ºF) Note: The following tolerance is added to the devices. For more information contact your above performance. local Yokogawa sales offices. mA output tolerance: ±0.02 mA of “4-20 mA” Maximum load resistance : - Step response ≤...
  • Page 14 2-2 Specifications P) Housing: - Material : Cast aluminium case with chemically resistant coat- ing, cover with flexible polycarbonate window. - Color : Case : Off-white (Equivalent to Possible Munsell 2.5Y8.4/1.2) Cover : Deepsea Moss 31.5 V green(Equivalent to Mun- sell 0.6GY3.1/ 2.0) Voltage (V) - Cable gland :...
  • Page 15 Specifications 2-3 V) Explosionproof type Refer to Control Drawings. Item Description Code CSA Intrinsically safe Approval Item Description Code Applicable standard: C22.2, No. 0-M1991, C22.2, No. 04-M2004, C22.2, No. 157-M1992, FM Intrinsically safe Approval C22.2, No. 61010-1 Applicable standard: FM3600, FM3610, FM3810 Intrinsically Safe for Class I, Division 1, Groups ABCD Ex ia Class I, Division...
  • Page 16 G. Configuration : Local with 6 keys • The external earth connection facility shall be H. Software : Firmware based on Yokogawa stack. connected reliably. I. Hardware : Yokogawa HART® Modem F9197UB • The instrument modification or parts replacement J.
  • Page 17 B. Supply voltage: 9 to 32 V DC FBUS) C. Operating current: 26.0 mA M. Other control systems: D. Operating values: According to IEC 1158-2 YOKOGAWA PRM, DTM E. Bus connection : Fieldbus interface base on IEC1158-2 according to FISCO- Model F.
  • Page 18: Model And Suffix Codes

    2-6 Specifications 2-2. Model and suffix codes 1. 2-wire Inductive conductivity transmitter (General purpose) [Style: S2] Model Suffix Code Option Code Description ISC202G 2-wire Inductive conductivity transmitter Type mA with HART Profibus Language Japanese English Option Mounting Hardware Pipe, wall mounting bracket (Stainless steel) Panel Mounting bracket(Stainless steel) Hood Hood for sun protection (Carbon steel)
  • Page 19: Control Drawing Isc202S Ma Hart ® Specification (Iecex)

    Specifications 2-7 2-3. Control Drawing ISC202S mA HART ® Specification (IECEx). Intrinsically safe design Ex ia or ib IEC Ex standard Ex ia IIC : T4 for ambient temp. < 55°C C ertified safety barrier or pow er T6 for ambient temp. < 40°C w ith R int=300 C ertificate nr.
  • Page 20: Control Drawing Isc202S Ma Hart ® Specification (Atex)

    2-8 Specifications 2-4. Control Drawing ISC202S mA HART ® Specification (ATEX) Intrinsically safe design CEN ELEC standard EEx ia IIC: T4 for ambient temp. < 55°C EEx ia or ib Certified safety barrier or pow er T6 for ambient temp. < 40°C with Rint=300 C ertificate nr.
  • Page 21: Control Drawing Isc202S Ma Hart ® Specification

    Specifications 2-9 2-5. Control Drawing ISC202S mA HART ® Specification (FM Intrinsically safe design) Intrinsically safe design FM Approved safety barrier or FM Class I, Div.1, Group ABCD, T4 for ambient temp. < 55°C power supply T6 for ambient temp. < 40°C with Rint = 300 ISC202S transmitter (HART compatible)
  • Page 22 2-10 Specifications 2-6. Control Drawing ISC202S mA HART ® Specification (FM Non-incendive design) Nonincendive design Class I, Div.2, Group ABCD, T4 for ambient temp. < 55°C T6 for ambient temp. < 40°C ISC202S transmitter FM Approved power supply Voc ≤ 31.5 V DC ISC40S Sensor For electrical data: terminals 11...
  • Page 23: Control Drawing Of Isc202S Ma Hart ® Specification (Csa)

    Specifications 2-11 2-7. Control Drawing of ISC202S mA HART ® Specification (CSA) Intrinsically safe design C SA certified safety barrier or C SA Ex ia C lass I, D iv.1, G roup A BC D , T4 for ambient temp. < 55°C pow er supply w ith R int=300 T6 for ambient temp.
  • Page 24: Control Drawing Of Isc202S Ff/Pb Specification (Iecex)

    2-12 Specifications 2-8. Control Drawing of ISC202S FF/PB Specification (IECEx) Ex ia IIC T4 for ambient temp. 55 C Ui = 24 V Ui = 17,5 V Ii = 250 mA Ii = 380 mA Pi = 1,2 W Pi = 5,32 W ISC202S-F or ISC202S-P Sensor...
  • Page 25: Control Drawing Of Isc202S Ff/Pb Specification (Atex)

    Specifications 2-13 2-9. Control Drawing of ISC202S FF/PB Specification (ATEX) Ex ia IIC T4 for ambient temp. 55 C Ui = 24 V Ui = 17,5 V Ii = 250 mA Ii = 380 mA Pi = 1,2 W Pi = 5,32 W ISC202S-F or ISC202S-P Sensor...
  • Page 26: Control Drawing Of Isc202S Ff/Pb Specification (Fm Intrinsically Safe Entity)

    2-14 Specifications 2-10. Control Drawing of ISC202S FF/PB Specification (FM Intrinsically safe Entity) FM Class I, DIV. 1, Group ABCD Sensor T4 for ambient temp. 55 C Connections Max. cablelength: 60 mtr. ISC202S-F Cable dia. : 3…12 mm. or ISC202S-P Sensor Connections FM Approved...
  • Page 27 Specifications 2-15 The cable used to interconnect the devices needs to comply with the following parameters: Loop resistance R’: 15 … 150 Ω/km; Inductance per unit length L’: 0,4 … 1 mH/km Capacitance per unit length C’: 80 … 200 nF/km (C’...
  • Page 28: Control Drawing Of Isc202S Ff/Pb Specification (Fm Intrinsically Safe Fisco)

    2-16 Specifications 2-11. Control Drawing of ISC202S FF/PB Specification (FM Intrinsically safe FISCO) FM Class I, DIV. 1, Group ABCD Sensor T4 for ambient temp. 55 C Connections ISC202S-F Max. cablelength: 60 mtr. or ISC202S-P Cable dia. : 3…12 mm. Sensor Connections FM Approved...
  • Page 29 Specifications 2-17 In each I.S. Fieldbus segment only one active source, normally the FM Approved FISCO barrier, is allowed to provide the necessary power for the Fieldbus system. All other equipment connected to the bus cable has to be passive (not providing energy to the system), except to a leakage current of 50μA for each connected device.
  • Page 30: Control Drawing Of Isc202S Ff/Pb Specification (Fm Non-Incendive Entity)

    2-18 Specifications 2-12. Control Drawing of ISC202S FF/PB Specification (FM Non-incendive Entity) FM Class I, DIV. 2, Group ABCD T4 for ambient temp. 55 C Sensor T6 for ambient temp. 40 C Connections ISC202S-B Max. cablelength: 60 mtr. or ISC202S-D Cable dia.: 3…12 mm.
  • Page 31: Control Drawing Of Isc202S Ff/Pb Specification (Fm Non-Incendive Fnico)

    Specifications 2-19 2-13. Control Drawing of ISC202S FF/PB Specification (FM Non-incendive FNICO) FM Class I, DIV. 2, Group ABCD T4 for ambient temp. 55 C Sensor T6 for ambient temp. 40 C Connections ISC202S-B Max. cablelength: 60 mtr. or ISC202S-D Cable dia.: 3…12 mm.
  • Page 32: Control Drawing Of Isc202S Ff/Pb Specification (Csa)

    2-20 Specifications 2-14. Control Drawing of ISC202S FF/PB Specification (CSA) CSA Ex ia Class I, DIV. 1, Group ABCD T4 for ambient temp. 55 C Ui = 24 V Ui = 17,5 V Ii = 250 mA Ii = 380 mA Pi = 1,2 W Pi = 5,32 W ISC202S-F...
  • Page 33: Installation And Wiring

    Installation and wiring 3-1 3. INSTALLATION AND WIRING 3-1. Installation and dimensions 3-1-1. Installation site The EXA transmitter is weatherproof and can be installed inside or outside. It should, however, be installed as close as possible to the sensor to avoid long cable runs between sensor and transmitter. In any case, the cable length should not exceed 60 mtr (200 feet).
  • Page 34: Preparation

    3-2 Installation and wiring Unit: mm (inch) (0.14) SPACING PANEL CUTOUT PANEL CUTOUT Fig. 3-2b. Panel mounting using two (2) self-tapping screws Unit: mm (inch) Pipe mounting Pipe mounting (Vertical) (Horizontal) (2.20) 2-Ø6.5 (0.26) (7.87) 4-Ø10 (0.4) (4.5) (2.75) Nominal 50 A (O.D. Ø60.5 mm) (2 inch pipe) Figure 3-3.
  • Page 35: Wiring Of Sensors

    Installation and wiring 3-3 to 2.5 mm (26 to 14 AWG). The glands will form a tight seal on cables with an outside diameter in the range of 6 to 12 mm (0.24 to 0.47 inches). HAND HELD COMPUTER COMMUNICATOR HOLD FAIL MODE...
  • Page 36: Wiring Of The Power Supply

    3-4 Installation and wiring 3-4 Wiring of the power supply 3-4-1 General precautions WARNING First make sure that the DC-power supply is according the specifications given. DO NOT USE ALTERNATING CURRENT OR MAINS POWER SUPPLY! ! The cable leading to the distributor (power supply) or safety barrier transports power to an output signal from the transmitter.
  • Page 37: Other Sensor Systems

    Where a convenient installation is not possible using the standard cables between sensors and transmit- ter, a junction box and extension cable may be used. The Yokogawa BA10 junction box and the WF10 extension cable should be used. These items are manufactured to a very high standard and are neces- sary to ensure that the specifications of the system are not compromised.
  • Page 38 3-6 Installation and wiring Extension cable may be purchased in bulk quantities, cut to length. Then it is necessary to terminate the cable as shown below. Termination procedure for WF10 cable. 1. Slide 3 cm of heat shrink tube (9 x 1.5) over the cable end to be terminated. 2.
  • Page 39: Operation; Display Functions And Setting

    Operation 4-1 4. OPERATION; DISPLAY FUNCTIONS AND SETTING 4-1. Operator interface This section provides an overview of the operation of the EXA operator interface. The basic procedures for obtaining access to the three levels of operation are described briefly. For a step-by-step guide to data entry, refer to the relevant section of this user’s manual.
  • Page 40: Explanation Of Operating Keys

    4-2 Operation Output hold flag Fail flag Menu pointer flags MODE HOLD FAIL Units Commissioning MEASURE Main display function menu OUTPUT DISPLAY SET HOLD YES NO HOLD TEMP SERVICE Message display Commissioning MODE mode access key Key prompt flags Selection keys YES : Accept setting NO : Change setting Measure/Maintenance...
  • Page 41: Setting Passcodes

    Operation 4-3 4-3. Setting passcodes 4-3-1. Passcode protection In Service Code 52, EXA users can set the passcode protection for each one of the three operating lev- els. This procedure should be completed after the initial commissioning (setup) of the instrument. The passcodes should then be recorded safely for future reference.
  • Page 42: Display Functions

    4-4 Operation 4-5. Display functions Display functions daigram is shown below. Display Functions μ S / c m Actual cell constant YES NO μ S / c m μ S / c m Reference temperature MODE DISP.1 DISP.2 μ S / c m μ...
  • Page 43: Parameter Setting

    Parameter setting 5-1 5. PARAMETER SETTING 5-1. Maintenance mode 5-1-1. Introduction Standard operation of the EXA transmitter involves use of the Maintenance (or operating) mode to set up some of the parameters. Access to the maintenance mode is available via the six keys that can be pressed through the flexible window in the instrument front cover.
  • Page 44: Manual Calibration To Determine The Cell Constant (C.c.)

    5-2 Parameter setting 5-1-2. Manual calibration to determine the cell constant (C.C.) DISP.1 Enter the Conductivity value of the sample at reference temperature HOLD HOLD ENABLED TO ELSE TO MEASURE IM 12D06A03-01E...
  • Page 45: Second Line Display. Referring To The First Compensated Conductivity

    Parameter setting 5-3 5-1-3. Second Line display. Referring to the first compensated conductivity. DISP.2 IF TEMP.2 = TEMP.1 (Section 5-2-4, 5-2-5) HOLD if enabled CALIB HOLD HOLD IM 12D06A03-01E...
  • Page 46: Second Line Display. Referring To The Second Compensated Conductivity

    5-4 Parameter setting 5-1-4. Second Line display. Referring to the second compensated conductivity. Conductivity Value Compensation method for second conductivity value. IM 12D06A03-01E...
  • Page 47 Parameter setting 5-5 5-1-5. Manual activation of HOLD This option is only available if HOLD is enabled in Section 5-2. IM 12D06A03-01E...
  • Page 48: Commissioning Mode

    5-6 Parameter setting 5-2. Commissioning mode In order to obtain peak performance from the EXA ISC202, you must set it up for each custom application. *OUTP : mA output is set as default to 0-1000 mS/cm. For enhanced resolution in more stable measuring processes, it may be desirable to select for example 0-100 μS/cm range.
  • Page 49 Parameter setting 5-7 5-2-1. Linear output (Range) YES NO 0% corresponds to 4 mA Note: 100% corresponds to 20 mA IM 12D06A03-01E...
  • Page 50 5-8 Parameter setting 5-2-2. HOLD Here HOLD is enabled/disabled. When enabled: • The analyser’s output will be set to HOLD when entering Maintenance-, Commissioning- or Service menu. • Up on exiting one of the menus, the user is asked if HOLD should remain activated.
  • Page 51: Temperature Compensation

    Parameter setting 5-9 5-2-3. Temperature compensation Why temperature compensation? The conductivity of a solution is very dependent on temperature. Typically for every 1 °C change in temperature the solution conductivity will change by approximately 2 %. The effect of temperature varies from one solution to another and is determined by several factors like solution composition, concentra- tion and temperature range.
  • Page 52 5-10 Parameter setting Calculation example Calculate the temperature co-efficient of a liquid from the following data. Conductivity 124.5 μS/cm at a liquid temperature of 18.0 °C and a conductivity 147.6 μS/cm at a liquid temperature of 31.0 °C. Substituting the data in the above formula gives the following result. 147.6 - 124.5 α...
  • Page 53: Temperature Compensation For First Conductivity Value

    Parameter setting 5-11 5-2-4. Temperature compensation for first conductivity value (To *TEMP.2) IM 12D06A03-01E...
  • Page 54: Temperature Compensation For Second Conductivity Value

    5-12 Parameter setting 5-2-5. Temperature Compensation for second conductivity value (To *SERV ) IM 12D06A03-01E...
  • Page 55: 5-3.Service Codes

    Parameter setting 5-13 5-3. Service Codes Don't set or input service code numbers other than the code numbers defined in this manual. Setting an undefined service code may make the transmitter malfunction. When an undefined service code is input by some accident, push the MODE key and escape from the service level.
  • Page 56: Temperature Compensation Functions

    5-14 Parameter setting 5-3-3. Temperature compensation functions Code 20 *T.R.°C Choose a temperature to which the measured conductivity (or resistivity) value must be compensated. Normally 25°C is used, therefore this temperature is chosen as default value. Limitations for this setting are: -20 to 140 °C. If *T.UNIT in code 11 is set to °F, default value is 77°F and the limitations are 0 - 280°F.
  • Page 57: Keys

    Parameter setting 5-15 Code Display Function Function detail Default values Temperature compensation functions *T.R.°C Set reference temp. keys to set value 25 °C *T.C.1 Set temp. coef. 1 Adjust compensation factor 2.10 % if set to *T.C. in section 5-2-5. per °C Set value with keys...
  • Page 58 5-16 Parameter setting 5-3-4. mA output functions Code 31 *OUTP.F For the ISC202 the output may be chosen as linear to input, or configured in a 21 point table to a particular linearization. Enable the table setup in code 31, and configure the table in code 35.
  • Page 59: Ma Output Functions

    Parameter setting 5-17 Concentration Output function is done in de following order: • Set *OUTP.F. (Service Code 31) to table • Set the Concentration range in % (Service Code 55) • Set table values (%output and Conductivity values) in *TABLE (Service Code 35) Code Display Function...
  • Page 60: User Interface

    5-18 Parameter setting 5-3-5. User interface Code 50 *RET. When Auto return is enabled, the transmitter reverts to the measuring mode from anywhere in the configuration menus, when no button is pressed during the set time interval of 10 minutes. Code 52 *PASS Passcodes can be set on any or all of the access levels, to restrict access to...
  • Page 61: Communication Setup

    Parameter setting 5-19 5-3-6. Communication setup Code 60 *COMM. The settings should be adjusted to suit the communicating device connected to the output. The communication can be set to HART or to PH 201*B ® distribution (see Appendix 2). *ADDR. Select address 00 for point to point communication with 4-20mA transmission.
  • Page 62 5-20 Parameter setting Code Display Function Function detail Default values Communication *COMM. Communication Set HART communication ® Set HART communication ® Set communication PH 201*B on Communication write enable write Communication write protect enable *ADDR. Network address Set address 00 to 15 *HOUR Clock setup Adjust to current date and time using...
  • Page 63: Calibration

    6. CALIBRATION 6-1 When is calibration necessary? Calibration of conductivity instruments is normally not required, since Yokogawa delivers a wide range of sensors, which are factory calibrated traceable to NIST standards. The cell constant values are nor- mally indicated on the top of the sensor or on the integral cable. These values directly can be entered in service code 03 (section 5-3-1).
  • Page 64: Calibration Procedure

    6-2 Calibration 6-2. Calibration procedure MODE MEASURE DISPLAY HOLD MODE Press the MODE key The legend CALIB appears, and the YES/NO key prompt flags flash Put the sensor in standard solution Press YES Set the value using the Select the flashing digit with the Increase it’s value by pressing the When the correct value is displayed, press ENT to enter the change...
  • Page 65: Maintenance

    The EXA instrument contains a lithium cell to support the clock function when the power is switched off. This cell needs to be replaced at 5 yearly intervals (or when discharged). Contact your nearest Yokogawa service centre for spare parts and instructions. 7-2. Periodic maintenance of the sensor Maintenance advice listed here is intentionally general in nature.
  • Page 66: Troubleshooting

    Trouble shooting 8-1 8. TROUBLESHOOTING 8-1. Introduction The EXA ISC202 microprocessor based conductivity analyser continuously monitors the condition of all key components of the measuring system to ensure that measurement is dependable. If a fault is detected this is immediatly signalled. Errors are shown on the display with a code. Table shows the errors which can be detected and gives information to help locate the fault or identify the error.
  • Page 67: Error Messages And Explanation

    Impossible AIR SET Too high ZERO Replace sensor E10* EEPROM write failure Software problem Unplug the unit, try again call Yokogawa Service Impossible adjustment Temperature Abnormal cable resistance Check reference see section 5-3-2 Outspan span too small Max. zero suppression is 90%...
  • Page 68: Spare Parts

    Spare parts 10-1 10. SPARE PARTS See Customer Maintenance Parts List. IM 12D06A03-01E...
  • Page 69: User Setting For Non-Linear Output Table (Code 31, 35)

    Appendix 1-1 11. APPENDIX 1 11-1. User setting for non-linear output table (code 31, 35) Output signal value S/cm S/cm S/cm Output 4-20 10.4 11.2 12.0 12.8 13.6 14.4 15.2 16.0 16.8 17.6 18.4 19.2 20.0 11-2. User entered matrix data (code 23 to 28) Medium: T1 data T2 data...
  • Page 70: Matrix Data Table (User Selectable In Code 22)

    1-2 Appendix 11-3. Matrix data table (user selectable in code 22) Matrix, Solution Temp ( ºC) Data 1 Data 2 Data 3 Data 4 Data 5 1. Sulfuric acid H 2 SO 4 33.8 mS 63.5 mS 95.0 mS 124.5 mS 154.0 mS 0.5 –...
  • Page 71: Configuration Checklist For Isc202

    Appendix 1-3 11-4. Configuration Checklist For ISC202 Primary choices Defaults Alternatives Reference Menu or on page Service code (SC) Measurement Conductivity 1st Temp. compensation NaCl in water Fixed T.C., Matrix 5-2-4 ‘Commissioning’ 2nd Temp. compensation NaCl in water Fixed T.C., Matrix (inactive) 5-2-5 ‘Commissioning’...
  • Page 72: Device Description (Dd) Menu Structure

    1-4 Appendix 11-6. Device Description (DD) menu structure The Device Description (DD) is available from Yokogawa or the HART foundation. An example is shown ® below of the ON LINE menu structure. This manual makes no attempt to explain the operation of the Hand Held Communicator (HHC).
  • Page 73: Preface

    Appendix 2-1 12. APPENDIX 2 12-1. Preface This appendix contains these items. 1. Method of wiring and parameter setting the following:. PH201G*B Dedicated Distributor BA2O Junction Terminal Box WF1OJ Extension Cable 2. Quick reference for parameter setting 3. Installation factor adjustment. To ensure that this measurement system can be operated safely and also exhibit its full performance, be sure to read this appendix before use.
  • Page 74 2-2 Appendix 12-3. Peripheral products 12-3-1. PH201G*B Dedicated Distributor Model and suffix codes Model Suffix Code Option Code Description PH201G Distributor Power Supply 100V AC 220V AC Style B Option Terminal for Power connection T2.2.2E.EPS Communication setup with PH201G (style B) distributor (service code 60) This communication is a one-way to PH201G (Style B) distributor, a power supplier for the EXA 202 transmitters.
  • Page 75: Ba20 Junction Terminal Box

    Appendix 2-3 12-3-2. BA20 Junction Terminal Box Model and suffix codes Suffix Option Model Description code code Terminal box BA20 T1.3.5.eps External Dimensions and Wiring 4-Φ6.6 Mounting hole Inlet for extension cable (Φ21.5 hole) Equivalent to DIN Pg13.5 cable gland Ground terminal M4 Screw (50)
  • Page 76: Wf10J Extension Cable

    2-4 Appendix 12-3-3. WF10J Extension Cable Model and suffix codes Suffix Option Model Description code code WF10J Extension cable Cable end Finished ends Cable length 10 m 20 m 30 m 40 m T1.3.6.eps (Note) The maximum extension cable length is 50m including sensor cable length.
  • Page 77: Quick Reference For Parameter Setting

    Appendix 2-5 12-4. Quick reference for parameter setting Preparation for operation entails deciding maintenance-mode settings and measurement range set- ting-related issues. Service level (mode) settings are described in Sec. 5-3. Shipping-time measure- ment parameter setting defaults are shown in Table 12-1. In general there is no need to change these defaults.
  • Page 78: Settings Performed In Maintenance Mode

    (we recommend Yokogawa's Personal Conductivity Meter). The reference tempera- ture of both meters should be the same, and as far as possible you should perform the calibration at the reference temperature, as there will otherwise be temperature compensation errors if the temperature compensation functions of the two meters are not exactly the same.
  • Page 79: Commissioning Mode Settings

    Appendix 2-7 12-4-2. Commissioning Mode Settings 12-4-2-1 Output Range Setting Refer to Sec. 5-2-1 1. What does output range setting involve? When linear output (linear relationship between conductivity and analog output) is selected, the meas- urement range corresponding to the 4 to 20 mA DC analog output range must be set. The default measurement range is 0 to 1,000 mS/cm.
  • Page 80: Temperature Compensation

    2-8 Appendix 3. Examples While washing a reactor, HOLD is turned ON to prevent interference with control. When washing finish- es, and a new batch starts, HOLD is turned OFF again and the sensor reverts to conductivity measure- ment. 4. Auto Return If AutoReturn functions are enabled, the instrument reverts to measurement mode after a preset time (10 min.
  • Page 81: Actual Setting Examples

    Appendix 2-9 12-4-3. Actual Setting Examples WARNING WARNING Do not enter any Service Codes other than those specified in this document. If you enter a non-specified Service Code it may affect the program firmware or data, and adversely affect the operation of the instrument. This section provides representative key operation flowcharts.
  • Page 82: Setting Output In Terms Of Concentration

    2-10 Appendix 12-4-3-1 Setting Output in terms of Concentration START Set Reference Temperature *OUTP Method 1: Linear relationship between Set Measurement Range Enter conductivities corresponding to 0% conductivity and concentration and 100% output concentration Method 2: Line-segment modeling of non-linear relationship *SERV 31 select 1 between conductivity and concentration...
  • Page 83 Appendix 2-11 Sample 1 NaCl concentration Meas. range 250-320 g/l Reference temp. 60°C Temp. compensation Use NaCl compensation (default) Concentration ― conductivity relationship is linear START Service Code 20 60°C 0% out -> 404 mS/cm OUTP 100% out -> 450 mS/cm (These conductivity values apply to reference temperature) Select 1 Service Code 55...
  • Page 84 2-12 Appendix Sample 2 NaOH concentration Meas. range 0-10% Reference temp. 25°C Temp. compensation Use existing (temperature, concentration) matrix START 25°C (default) Service Code 20 Select 1 Service Code 31 Service Code 35 Enter conductivity values corresponding to output current values. Existing table (below) relates output current to conductivity at 25°C NaOH conc (%) Output Current (%)
  • Page 85 Appendix 2-13 Sample 3 Concentration of sulfuric acid mixture Meas. range 93-97% Reference temp. 50°C Temp. compensation Enter temperature compensation data into a 5 x 5 matrix START Service code 20 50°C(Default) Service code 31 Select 1 Enter conductivity values corresponding to output current values. Service code 35 Existing table (below) relates output current to conductivity at 50°C conc (%)
  • Page 86 2-14 Appendix Sample 4 Concentration of hydrochloric acid Meas. range 30-40% Reference temp. 35°C Temp. compensation Enter temperature compensation coefficient, Relate concentration to conductivity by table START Service code 20 35 °C (Default) Service code 31 Select 1 Service code 35 Enter conductivity values corresponding to output current values.
  • Page 87: Key Operation Procedure Examples

    Appendix 2-15 12-4-3-2 Key Operation Procedure Examples This uses Sample 1 above as the example 1. Reference Temperature Setting Temperature compensation converts the measured conductivity to the equivalent at the reference tem- perature. The nearer the liquid temperature to the reference temperature, the lower the conversion error.
  • Page 88 2-16 Appendix 2. Output Table Setting Example *SERV *CODE 35 845 mS/cm Enter repeatedly until you reach 20% 810 mS/cm Enter repeatedly until you reach 40% 772 mS/cm Enter repeatedly until you reach 60% 731 mS/cm Enter repeatedly until you reach 80% 688 mS/cm Enter repeatedly until you reach 100%...
  • Page 89 Appendix 2-17 4. Temperature Compensation Coefficient Setting Example *OUTP *HOLD *TEMP *NaCl Escape MODE *SERV CODE 21 TC 1.24 example4E.eps The temperature compensation coefficient is calculated as follows: T = Measured temperature ( °C ) α Kt = Conductivity coefficient at temperature T ( °C ) T - T Tref = Reference temperature ( 25 °C ) Kref = Conductivity coefficient at temperature Tref...
  • Page 90 2-18 Appendix 6. One-point Calibration example Normally you would do a laboratory analysis when the sensor is installed and use the resulting value to calibrate the transmitter. If the concentration at start-up time is known, then you can use concentration to conductivity tables to determine the conductivity, and use that.
  • Page 91: Installation Factor Adjustment

    Appendix 2-19 12-5. Installation factor adjustment ACCESS CODE : 03 DISPLAY : *C.C. Adjustment: Adjust the ratio between the measured conductance of the sensor and the specific conductivity of the solution. Explanation: The installation factor for the ISC40 detector is the ratio of the measured conductivity of the sensor and the specific conductivity of the solution.
  • Page 92 2-20 Appendix IM 12D06A03-01E...
  • Page 93: Appendix 3 Quality Inspection

    Appendix 3-1 13. APPENDIX 3 QUALITY INSPECTION 13-1. ISC202G 2-Wire Inductive Conductivity Transmitter Quality ISC202G 2-Wire Inductive Conductivity Inspection Transmitter Standards Scope This inspection standard applies to the ISC202G 2-Wire Inductive Conductivity Transmitter. Inspection Items 2.1 Insulation resistance test 2.2 Current output test 2.3 Temperature indication check * 2.4 Resistance (conductivity) indication check Note: Items marked with an asterisk (*) may only be confirmed by a test certificate.
  • Page 94 3-2 Appendix Table 1 Current Output (Indication) Output Range 4.0 mA 4 ±0.02 mA DC 12.0 mA 12 ±0.02 mA DC 20.0 mA 20 ±0.02 mA DC Temperature Indication Check <Test> From the last step in Item 3.2, Current Output Test, press the [ENT] key repeatedly until the message display shows “PT1000.”...
  • Page 95 Appendix 3-3 Table 3 SEL.10 Indication Setpoint of Resistance Box 2 Indication Range RES. 1 0.300±0.002 30.00 RES. 3 800.00 8.00±0.04 RES. 5 30.00k 300±2 Change the number of turns of wire onto the ISC40GJ sensor from ten to one. Press the [ENT] key.
  • Page 96 3-4 Appendix ISC202G Resistance Box 1 ISC40 Resistance Box 2 10 turns or 1 turn DC Power Source Load Resistance DC Ammeter Figure 1 Testing Circuit and Test Equipment QIS 12D06A03-01E IM 12D06A03-01E...
  • Page 97 Appendix 3-5 IM 12D06A03-01E...
  • Page 98: Isc202S 2-Wire Inductive Conductivity Transmitter

    3-6 Appendix 13-2. ISC202S 2-Wire Inductive Conductivity Transmitter Quality ISC202S Inspection 2-Wire Inductive Conductivity Transmitter Standards Scope This inspection standard applies to the ISC202S 2-Wire Inductive Conductivity Transmitter. Inspection Items 2.1 Insulation resistance test * 2.2 Dielectric strength test 2.3 Current output test 2.4 Temperature indication check * 2.5 Resistance (conductivity) indication check Note: Items marked with an asterisk (*) may only be confirmed by a test certificate.
  • Page 99 Appendix 3-7 When the message display shows “4 mA,” the output current must be within in the range shown in Table 1. Press the [ENT] key repeatedly until the message displays shows “12 mA.” When it shows “12 mA,” the output current must be within the range shown in Table 1. To skip the current output not needed to be checked, just press the [ENT] key.
  • Page 100 3-8 Appendix Press the [ENT] key. The message display shows “ WAIT ” momentarily, followed by “RES.2.” There is no need to check here, so just press the [ENT] key. The message display shows “ WAIT ” momentarily, followed by “RES.3.” Set resistance box 2 to the value shown in Table 3 and check the indication.
  • Page 101 Appendix 3-9 ISC202S Resistance Box 1 ISC40 Resistance Box 2 10 turns or 1 turn DC Power Source Load Resistance DC Ammeter Figure 1 Testing Circuit and Test Equipment QIS 12D06A03-21E IM 12D06A03-01E...
  • Page 102 3-10 Appendix IM 12D06A03-01E...
  • Page 103: Isc202G, Isc202S 2-Wire Inductive Conductivity Transmitter (Fieldbus Communication)

    Appendix 3-11 13-3. ISC202G, ISC202S 2-Wire Inductive Conductivity Transmitter (Fieldbus Communication) Quality ISC202G, ISC202S Inspection 2-Wire Inductive Conductivity Transmitter (Fieldbus Communication) Standards Scope This inspection standard applies to the ISC202G and ISC202S 2-Wire Inductive Conductivity Transmitters (Fieldbus specification). Inspection Items 2.1 Insulation resistance test * 2.2 Dielectric strength test 2.3 Temperature indication check...
  • Page 104 Resistance Box 2 6.27 mS/cm 300 Ω 6.27 ±0.04 mS/cm 1.567 mS/cm 1.567 ±0.009 mS/cm 1.2 k 376 μS/cm 376 ±3.7 μS/cm Fieldbus Communication Functional Check Check for normal function using Fieldbus equipment specified by Yokogawa. QIS 12D06A03-61E IM 12D06A03-01E...
  • Page 105 Appendix 3-13 Figure 1 Testing Circuit and Test Equipment QIS 12D06A03-61E IM 12D06A03-01E...
  • Page 106 3-14 Appendix IM 12D06A03-01E...
  • Page 107: Isc202G, Isc202S 2-Wire Inductive Conductivity Transmitter (Profibus Communication)

    Appendix 3-15 13-4. ISC202G, ISC202S 2-Wire Inductive Conductivity Transmitter (Profibus Communication) Quality ISC202G, ISC202S Inspection 2-Wire Inductive Conductivity Transmitter (Profibus Communication) Standards Scope This inspection standard applies to the ISC202G and ISC202S 2-Wire Inductive Conductivity Transmitters (Profibus specification). Inspection Items 2.1 Insulation resistance test * 2.2 Dielectric strength test 2.3 Temperature indication check...
  • Page 108 3-16 Appendix In this state, change the resistance value of the decade resistance box 1 as shown in Table 1. The corresponding temperature indication must be within the range. Table 1 Temperature Indication Check (PT1000) Reference Resistance of Indication Range Temperature Resistance Box 1 25 °C...
  • Page 109 Appendix 3-17 Profibus Communication Functional Check Check for normal function using Profibus equipment specified by Yokogawa. In the tests of Item 3.3 to 3.4, check the communication function using Profibus communication. Figure 1 Testing Circuit and Test Equipment QIS 12D06A03-71E...
  • Page 110 3-18 Appendix IM 12D06A03-01E...
  • Page 111 Customer Model ISC202G [Style: S2] Maintenance Inductive Conductivity Transmitter Parts List Item Part No. Description Cover Assembly K9315CA Polyurethane Coating K9315CN Epoxy-polyester Coating Internal Works Assembly with amplifier assembly K9661ED For mA + HART K9661EE For FF K9661EF For Profibus Housing Assembly K9661HA Polyurethane Coating...
  • Page 112 Pipe/Wall Mounting Hardware (Option Code : /U) Panel Mounting Hardware (Option Code : /SCT) Option Code : /H Hood to sun protection (Option Code : /PM) Item Parts No. Description K9171SS Universal Mount Set (/U) K9311BT Tag Plate (/SCT) K9311KA Fitting Assembly (/PM) K9311KG Hood Assembly (/H)
  • Page 113 Customer Model ISC202S [Style: S3] Maintenance Inductive Conductivity Transmitter Parts List Item Part No. Description Cover Assembly K9315CA Polyurethane Coating K9315CN Epoxy-polyester Coating Internal Works Assembly with amplifier assembly For mA + HART For FF For Profibus For mA + HART (Non-incendive) Housing Assembly Polyurethane Coating Epoxy-polyester Coating...
  • Page 114 Pipe/Wall Mounting Hardware (Option Code : /U) Panel Mounting Hardware (Option Code : /SCT) Option Code : /H Hood to sun protection (Option Code : /PM) Item Parts No. Description K9171SS Universal Mount Set (/U) K9311BT Tag Plate (/SCT) K9311KA Fitting Assembly (/PM) K9311KG Hood Assembly (/H)
  • Page 115: Revision Record

    Revision Record Manual Title : Model ISC202G [Style: S2], ISC202S [Style: S3] 2-wire Inductive Conductivity Transmitter Manual Number : IM 12D06A03-01E Edition Date Remark (s) Sep. 2004 Newly published Sep. 2005 Revised pages are shown below. PREFACE some corrected (i page); After-sales Warranty modified (iii page). Sec.
  • Page 116 Edition Date Remark (s) Apr. 2008 Style of ISC202S changed to S3 and related description changed as follows. p2-5, Style of ISC202S changed to S3 for FM approval; p3-1, Some of dimensions in Figure 3-1 corrected; p1-4 (Appendix), Note of HART protrocol DD files URL added;...
  • Page 117 Model ISC202G [Style: S2], ISC202S [Style: S3] User’s 2-Wire Inductive Conductivity Manual Transmitter Supplement Thank you for selecting our Model ISC202G [Style: S2] and/or ISC202S [Style: S3] 2-Wire Inductive Conductivity Transmitter. User's Manual, IM 12D06A03-01E, 7th Edition, supplied with the product, some revisions/additions have been made.
  • Page 118 Handle impaired. with care. Yokogawa is not responsible for damage to the instrument, poor performance of the instrument Although the instrument has a weatherproof or losses resulting from such, if the problems...
  • Page 119: J) Temperature Compensation: - Sensor Types: Pt1000Ω Or 30Kω Ntc - Automatic

    2-2 Specifications P) Housing: - Material : Cast aluminium case with chemically resistant coat- ing, cover with flexible polycarbonate window. - Color : Case : Off-white (Equivalent to Possible Munsell 2.5Y8.4/1.2) Cover : Deepsea Moss 31.5 V green(Equivalent to Mun- sell 0.6GY3.1/ 2.0) Voltage (V) - Cable gland :...
  • Page 120 �ppendix 2�1� Sample 3 Concentration of sulfuric acid mixture Meas. range 9��97% Reference temp. 50°C Temp. compensation �nter temperature compensation data into a 5 x 5 matrix START Service code 20 50°C (Default) Service code 31 Select 1 Enter conductivity values corresponding to output current values. Service code 35 Existing table (below) relates output current to conductivity at 50°C Output Current (%)
  • Page 121 Certification No.: KCC-REM-YHQ-EEN242-3 EQUIPMENT NAME : 2-Wire Inductive Conductivity ISC202G, ISC202S DATE OF MANUFACTURE : See the nameplate of the product. APPLICANT : Yokogawa Electirc Corporation MANUFACTURER : Yokogawa Electirc Corporation COUNTRY OF ORIGIN : JAPAN KCC-REM- YHQ-EEN242-3 All Rights Reserved, Copyright © 2008, 1st Edition: Sep.2011 (YK) IM 12D06A03-01E-S01 Subject to change without notice.

This manual is also suitable for:

Vigilantplant exa isc202s

Table of Contents