Summary of Contents for OSANG Healthcare GluNEO Lite Smart

  • Page 3 We understand that self-testing of blood glucose level provides a way to control your diabetes and may give you peace of mind by testing regularly. GluNEO Lite Smart has been developed to provide you with a fast and accurate reading in a convenient and simple process.
  • Page 4 4. Please refer to page 33 for cleaning and disinfecting GluNEO Lite Smart Blood Glucose Monitoring System. 5. Do not change your medication based on the GluNEO Lite Smart meter test results without contacting your physician or healthcare professional first. 6. Do not assemble or disassemble the blood glucose test meter.
  • Page 5: Table Of Contents

    Lancing device clear tip is used for alternate site testing-------------------------- 31-32 6. Caring for Your GluNEO Lite Smart System: Cleaning your meter and maintenance ---------------------------------------------------33 Storage of your GluNEO Lite Smart system ------------------------------------------ 34-35 Bluetooth function ---------------------------------------------------------------------------------- 36 7. Transferring Your Result Data ----------------------------------------------------------------37 8.
  • Page 6: Learning The System

    Learning the System Intended use The GluNEO Lite Smart test meter is used with GluNEO Lite Smart test strip for the measurement of glucose in fresh whole blood samples(Capillary and/or venous). It should not be used for the diagnosis or screening of diabetes or for the testing of new born babies.
  • Page 7: Gluneo Lite Smart Kit Contents

    9. 3V Coin (1EA) Battery (CR2032) 4. Lancets 5. Instructions for use Your GluNEO Lite Smart Blood Glucose Monitoring System has been sealed to protect the contents. If you find your seal has been broken, please return it to the place of purchase.
  • Page 8: Gluneo Lite Smart Meter

    Serial Number Battery Cover CAUTION 1. Do not use GluNEO Lite Smart glucose test meter in a dry environment, especially If synthetic materials are present. Synthetic clothes, carpets, etc., may cause damaging static discharge in a dry environment. 2. Do not use GluNEO Lite Smart glucose test meter near cellular or cordless...
  • Page 9: Gluneo Lite Smart Meter Display

    GluNEO Lite Smart Meter Display Low Battery Warning Symbol Alarm Symbol Troubleshooting Symbol Delete Results DATE Average Test Result System Before Having a Meal After Having a Meal After Taking Medication After Sport Activity Control Solution Symbol Blood Drop Symbol for Test...
  • Page 10: Gluneo Lite Smart Test Strip

    Contact Bar Caution 1. Store the GluNEO Lite Smart test strip vials in a cool, dry place. Keep out of direct sunlight. Do not freeze. 2. Store test strips in their original vial only. Do not mix the test strips in new vials or in any other container.
  • Page 11: Before Testing

    Before Testing WARNING To reduce the chance of infection: 1. Lancing device and lancets should NOT be shared with others. 2. Always use a new lancet and a new blood glucose test strip. 3. Practice using the lancing device and become accustomed with its use. 4.
  • Page 12: Setting Your Meter

    Setting Your Meter The GluNEO Lite Smart Meter has a wide variety of functions to choose. From the setup mode you can turn activity/meal flags, set the date/time, designate three unique averages, and set up to five daily alarms. User Activity Option...
  • Page 13 CAUTION 1. VERY IMPORTANT: Set the right test unit. Your GluNEO Lite Smart Meter is pre-set to mg/dL as test unit. A wrong test unit will lead to wrong test result display. Please consult your healthcare professional if you’re not sure which test unit to use. Without setting the date properly, the average glucose level and the results in memory will not show proper values.
  • Page 14: Auto-Coding Function

    Auto-coding Function 1. GluNEO Lite Smart has automatic code recognition function (e.g. CODE 125) 2. It recognizes the code number automatically. This is a very convenient function preventing the inconve- nience of setting the code number on the glucose test meter every time.
  • Page 15: Performing A Glucose Control Solution Test

    GluNEO Lite Smart control solution contains a known amount of glucose that reacts with GluNEO Lite Smart test strip. By comparing your control solution test results with the expected range printed on the test strip vial, you are able to check that the meter and test strips are working properly, and that you are performing the test correctly.
  • Page 16 1. Control solution is for in-vitro diagnostic use only. This means that it is only used for testing outside the body. 2. Control solution is exclusively made for the GluNEO Lite Smart Blood Glucose Monito- ring System. Do not use other brand control solutions made by other manufacturers.
  • Page 17 - Check the expiration dates on all the products you are using. Be sure you are using GluNEO Lite Smart Control Solution. - Try another control solution test. If this test falls out of range, try another control solution test with a new unopened bottle of test strips.
  • Page 18 Follow-up Action 1. Check your meter and test strips with the glucose control solution. 2. Perform your test again. CAUTION 1. If your glucose control solution falls out of range, do not perform a glucose test and please contact your local representative. 2.
  • Page 19: Performing Your Test Testing Blood - Preparation

    Performing Your Test Testing Blood - Preparation Lancing device is a convenient medical tool for collecting capillary blood samples for glucose monitoring or other tests which require one or two drops of blood. Being a more advanced lancing device, it provides ultimate safety and comfort for obtaining blood samples.
  • Page 20 COMPONENTS Clear Tip Ejection System Sliding Barrel Lancet Holder Depth Indicator/Adjustable Tip Release Button USING THE CLEAR TIP (AST) 1. Place the lancing device and lancet on the skin in the chosen area, press and hold it continuously for a few seconds, then push the release button to take a blood sample. 2.
  • Page 21: Performing A Blood Test With Your Fingertip

    Make sure the code number matches on the test strip bottle (e.g. CODE 125). GluNEO Lite Smart automatically recognizes the test strip code number and adjusts the meter accordingly. A blinking test strip will appear at the top of the screen, which indicates that the meter is ready for testing.
  • Page 22 Step 2 To obtain a suitable blood samples • Before obtaining a blood sample, wash your hands with warm, clean water and soap. Dry hands completely before testing. WARNING 1. All parts of the kit are considered bio hazardous and can potentially transmit infectious disease, even after you have conducted cleaning and disinfection measures.
  • Page 23 4. Put the lancing tip back and twist it tightly. 5. To adjust the depth setting: Use strength level 1-2 for soft skin, 3 for average skin, 4-5 for thick or calloused skin. 6. Pull the end of the lancet device back. 7.
  • Page 24 • Incorrect Incorrect Poorly filled GluNEO Lite Smart Test Meter only requires 0.5 ul sample volume. • Stick the edge of the test strip to the drop of blood. • The blood will automatically be ‘drawn’ into the test strip channel.
  • Page 25 Step 4 • After the beeping sound, test will begin automatically, and your results will appear in 5 seconds. It should begin counting down from 5 to 1 second on the LCD display window. LCD window will display the result of your blood glucose level, temperature and time.
  • Page 26: Test Strip Ejector Function

    Test Strip Ejector Function Disposing of your test strip 1. Once the test is complete, push the test strip forward to remove the test strip from the meter 2. Used test strips may be considered bio hazardous waste in your area. Be sure to contact your local representative for proper disposal.
  • Page 27: Reviewing Your Results

    Reviewing Your Results The GluNEO Lite Smart meter stores up to 365 test results in its built-in memory, along with the average sugar level for the number of days, which you had preset the glucose test meter to calculate. Reviewing Your blood sugar Press and release button.
  • Page 28: Reviewing Your Results With Activated User

    Reviewing your results with activated user ID Press the button once and the User ID option will appear the LCD screen. Use ▲ or ▼ buttons to scroll and choose the correct user ID, then press the button. If you choose the User ID to check the stored results, only the results of the selected user ID are displayed.
  • Page 29: Alternate Site Test

    Dorsal Hand Ventral Palm Fingertips GluNEO Lite Smart blood glucose monitoring system gives you the ability to your sugar from different areas of your body. ® The figure shown above displays the areas where you can test using your GluNEO Lite meter.You may test your blood sugar from your forearm, upper arm, palm (ventral palm...
  • Page 30 Ventral palm/Dorsal hand Choose a fleshy area on the palm, below your thumb or pinky finger. Select a spot without any visible veins and away from any deep lines, which may cause your blood sample to smear. Forearm Upper arm Choose a fleshy area of the forearm or upper arm away from bone, visible veins and thick hairs of the arm.
  • Page 31: Lancing Device Clear Tip Is Used For Alternate Site Testing

    To ensure accurate results when lancing your arm (forearm or upper arm), or palm (ventral palm or dorsal hand), clean the test site with soap and water. Make sure there is no cream or lotion on the test site. Thoroughly dry your hands and test site. WARNING To reduce the chances of getting infection: never share a lancet or lancing device with anyone.
  • Page 32 5. Continue to hold the lancing device and gradually increase pressure for several seconds. 6. While holding the lancing device on your test site, look through the clear tip until a round drop of blood appears. 7. Lift the lancing device straight up; be careful not to smear the blood on your testing site.
  • Page 33: Caring For Your Gluneo Lite Smart System

    2. Please store the GluNEO Lite Smart blood glucose monitoring system in a cool and dry area, out of reach of children. Do not freeze. For a more accurate test result, avoid testing under direct sunlight.
  • Page 34: Storage Of Your Gluneo Lite Smart System

    Storage of your GluNEO Lite Smart System To prevent the meter and test strips from getting dirt, dust or other contaminants, please wash and dry your hands thoroughly before use Meter Storage • Storage condition: 2~30 or 36~86 (Temperature) • Always store and transport the meter in its original storage case.
  • Page 35 CAUTION 1. Store the GluNEO Lite Smart test strip vials in a cool, dry place. Keep out of direct sunlight. Do not freeze. 2. Store test strips in their original vial only. Do not mix the test strips in new vials or in any other container.
  • Page 36: Bluetooth Function

    Bluetooth Function GluNEO Lite Smart Meter stores glucose test result data and can transmit it to wireless device such as a Smart phone and Tablet PC via Bluetooth wireless technology. STEP 1 Launch the application on your mobile phone. If Bluetooth function or Bluetooth icon is already activated on the LCD’s first screen of your mobile phone, please move on to the next step.
  • Page 37: Transferring Your Result Data

    Transferring Your Result Data You can transfer test results from the GluNEO Lite Smart meter to a computer. You will need to download the software from the following web site (http://www.osanghc.com) purchase the computer cable from your local representative. This is sold separately.
  • Page 38: Battery Installation

    Your GluNEO Lite Smart blood glucose test meter uses only one 3V lithium battery (CR2032), which is included. When replacing battery, only CR2032 or equivalent lithium battery should be used.
  • Page 40: Trouble-Shooting

    The following chart may help you identify certain problems, but may not solve all the problems that could occur. Contact your authorized representative or customer support center of OSANG Healthcare Co., Ltd., if the problem continues. OSANG Healthcare Co.,Ltd.
  • Page 41 Please insert a new glucose test strip and perform your test again. If “Sun”is blinking, avoid direct sunlight and re-test Contact your local representative or customer support at OSANG Healthcare Co., Ltd., if the problem persists.
  • Page 42 Change the batterry and if the problem persists, contact your local represen- tative or customer support at OSANG Healthcare Co., Ltd.
  • Page 44: Connection Between Mhealthgate And Gluneo Lite Smart

    ① Measure your blood glucose level by using GluNEO Lite Smart. ② Click mHealthGate app on your smartphone ③ Click “Connection device” button as below, and then select your connecting device. ④ Once the app is ready to receive your measuring value, following message is appeared.
  • Page 45 ⑤ If Bluetooth function is off status, ‘Bluetooth permission request’ is appeared as below. Push the <YES> to turn on Bluetooth to be connected with GluNEO Lite Smart device. ⑥ If the connection status between app and GluNEO Lite Smart is not set, the display of product number registration will be appeared.
  • Page 46 [Display of product number registration] [In case that registered product number is different from the device product number]...
  • Page 47 ⑦ Once the data is completely transferred, below display will be appeared. Select <Select & Save> button or <Save all> button, and store measuring value into your App. ※ If several people use one GluNEO Lite Smart device together, select <Select & Save> button to...
  • Page 48: Warranty

    Alternatively, you are requested to contact your place of purchase in your country for initial warranty request. OSANG Healthcare Co., Ltd. has trained specialists to help you 24 hours a day, 7 days a week, and 365days a year.
  • Page 49: Declaration Of Conformity

    Declaration of Conformity This Device Complies With Part 15 Of The FCC Rules. Operation is subject to the following two conditions; this device may not cause harmful interference, and this device must accept any interference received, including interference that may cause undesired operation. This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules.
  • Page 50: Specifications

    20~60% Bluetooth Bluetooth Smart (Bluetooth Low Energy)