Replacing The Lamp; Fig. 13 Replacing The Lamp - Zeiss SL 120 Documentation

Slit lamp
Hide thumbs Also See for SL 120:
Table of Contents

Advertisement

Replacing the lamp

CAUTION - RISK OF BURNS
Before replacing the lamp, switch the instrument off and disconnect the
power cable. Allow the lamp to cool down or wear protective gloves.
1
Lamp holder
2
Lamp (6 V/20 W halogen)
3
Spring-loaded catch
4
Knurled screw
Fig. 13
Replacing the lamp
• Loosen the knurled screw (4, Fig. 13) and remove the cover cap from the
lamp housing.
• Withdraw the lamp holder (1, Fig. 13) of the lamp.
• Press spring-loaded catch (3, Fig. 13) to unlock the bulb (2, Fig. 13).
• Then pull out the defective bulb (2, Fig. 13). Push new lamp (6 V, 20 W)
into the opening. Observe alignment of the locating pins. Do not touch
the glass bulb.
Maintenance and care
000000-1490-499-GA-GB-090118
35