Summary of Contents for Compat Ella

  • Page 1 Enteral feeding pump...
  • Page 2 Language index Loading the Cassette Instructions for Use Page English...
  • Page 3 Pump and Accessory Overview...
  • Page 4 AC Adapter Pole Clamp Administration Set...
  • Page 5: Table Of Contents

    Pump ® Table of Contents Introduction Feeding in Continuous Mode --------------------------------------------------------------- ------------------------------------------- Intended Use -------------------------------------------------------------- Feeding in Interval Mode ----------------------------------------------- Support --------------------------------------------------------------------- Pausing a Nutritional Therapy ----------------------------------------- Safety Information ------------------------------------------------------- Warnings and Precautions Delivery Information ---------------------------------------------- ---------------------------------------------------- Contraindications 11.1 View/Clear Delivery Information in Continuous Mode --------------------------------------------------------- -------------- General complications associated with enteral tube feeding...
  • Page 6: Introduction

    ® the gastrointestinal system. The pump should be used only after proper training regarding care, use and maintenance of the Compat Ella pump and accessories.
  • Page 7: Contraindications

    ® • If enteral feeding is contraindicated; • For the administration of intravenous infusions. • With sets other than “Compat Ella ” administration sets ® • If intestinal obstruction • If ileus (Inability of the intestine to contract normally/ move waste out of the body) •...
  • Page 8: General Complications Associated With Enteral Tube Feeding

    In all circumstances such an incident will be detected by the Compat Ella pump and will trigger an occlusion alarm.
  • Page 9: System Overview

    Note: Different versions of the administration sets are available. To obtain further information please contact your sales representative or go to www.compat.com. 3.3 Button Function The pump always retains in memory the on-screen settings and values when scrolling from one screen to the next, as well as when the pump is turned off.
  • Page 10: Display

    During pump start up, confirm all the screen segments shown in the diagram hereafter are visible. If any are missing or if the screen is not readable, do not use the pump. Screen Layout The Compat Ella pump features a colour display with alphanumeric characters, symbols and a backlight. ®...
  • Page 11: Battery

    The battery is fully charged and is plugged into the AC power. Note: Compat Ella only pumps fluid a little at a time. Normal function is to pump fluid for a while, then stop pumping fluid and then start pumping fluid ®...
  • Page 12: Pump Mode Overview (Continuous Vs. Interval)

    The battery capacity is an estimate. If you are unsure that enough capacity remains for your intended use, connect to the AC power until recharged. 4. Pump Mode Overview (Continuous vs. Interval) The Compat Ella portable enteral feeding pump has two operational modes: Continuous and Interval. ®...
  • Page 13: Continuous Mode: Menu Overview

    4.1 Continuous Mode: Menu Overview 4.2 Interval Mode: Menu Overview SET RATE VOLUME. SET RATE BOLUS, SET INTERVAL Press to scroll through Press to scroll through SET RATE SET TOTAL VOLUME. Pressing SET RATE Pressing again cycles back to again cycles back to Button Device Display Function Button Device Display Function The delivery speed in mL / h at which the...
  • Page 14: Clear Nutritional Therapy Settings And Delivery Information

    RATE, BOLUS, INTERVAL TOTAL VOLUME Note: must all be set / reviewed if either one is changed or the pump has just been turned on. Only then can the nutritional therapy be started. 5. Clear Nutritional Therapy Settings and Delivery Information Button Direction / Result Hold down...
  • Page 15: Power On And Off

    Power Status Light Solid green = Pump is plugged into the AC power. 6.4 Inserting the Administration Set Warnings: • Use only Compat Ella administration sets. ® • Inspect the Compat Ella administration set. Do not use if damaged or altered.
  • Page 16: Filling The Administration Set (Priming)

    6.5 Filling the Administration Set (Priming) Priming fills the administration set with tube feed. Precautions: • Before priming, always verify that the administration set is not connected to the patient. • Verify that the administration set is correctly installed, is connected to the nutrition container and is ready to prime. Auto-prime Auto-prime should always be used if possible.
  • Page 17: Quick Start Guides

    7. Quick Start Guides 7.1 Continuous Mode Feeding 7.2 Interval Mode Feeding Step Button Function Step Button Function Turn the pump on Turn the pump on 2” 2” Review the VOLUME GIVEN THIS THERAPY Review the VOLUME GIVEN THIS THERAPY Clear the VOLUME GIVEN THIS THERAPY Clear the...
  • Page 18: Feeding In Continuous Mode

    8. Feeding in Continuous Mode Continuous mode is used to deliver tube feed at defined rates, with or without a volume limit. Step Button Function Instruction The most recent nutritional therapy settings are always saved in memory. When pow- Turn the pump on er is turned on, the display shows the last programmed flow rate and volume and the actual volume given.
  • Page 19: Feeding In Interval Mode

    Ensure nutritional therapy settings correspond to the nutritional therapy prescribed. Connect the administration set to the patient’s feeding tube. Press to run the nutritional therapy. The pump will automatically run the Start the nutritional therapy nutritional therapy and display the progress. Precautions: VOLUME GIVEN THIS THERAPY •...
  • Page 20 • Hold 2 seconds (2”) to activate auto-prime. The pump will fill the administration set Prime the administration with a predetermined volume. • Press again to stop priming at any time. 2” Note: Do not prime when connected to the patient’s feeding tube. Press to access the SET RATE...
  • Page 21: Pausing A Nutritional Therapy

    Precautions: VOLUME GIVEN THIS THERAPY • When programming a new therapy, make sure to clear the parameter. • After starting a therapy, verify all therapy parameters (including rate, dose, and volume given this therapy). Notes: • Holding changes values rapidly: •...
  • Page 22: Delivery Information

    11. Delivery Information 11.1 View/Clear Delivery Information in Continuous Mode At any time, press to review the delivery information. To clear : Ensure the pump is paused. Press and then to clear value to 0. Button Device Display Function Display the volume in mL given for the current nutritional therapy. Note: Must be manually cleared at the beginning of each new nutritional therapy.
  • Page 23: View/Clear Delivery Information In Interval Mode

    11.3 View/Clear Delivery Information in Interval Mode At any time, press to review the delivery information and programmed nutritional therapy details. Press to scroll through each menu item below. VOLUME GIVEN THIS THERAPY CUMULATIVE VOLUME: Ensure When the last menu is reached, pressing the button exits the info mode.
  • Page 24: User Settings

    12. User Settings 12.1 Alarm level, Backlight, Autodim, End of Nutritional Therapy Alarm BEEP LEVEL, BACKLIGHT, AUTODIM? At any time, press to review and / or adjust the pump settings. Press again to scroll between ALARM AT END OF THERAPY?. When the last setting is reached, pressing exits the option setting mode.
  • Page 25: Changing The Language

    12.2 Changing the Language Step Button Direction Press and hold for 2 seconds (2”) to activate language selection. 2” to scroll to the desired language. Press and hold again for 2 seconds (2") to select the language. 2” Immediately turn the pump off to save chosen language. Restart the pump. Note: If the pump is not immediately turned off, the language change will not be saved.
  • Page 26: Advanced Settings

    14. Advanced Settings Precaution: The following settings should only be adjusted by a qualified healthcare professional. When the pump is off, press and hold and only after the "HOLD FOR ON" message appears on the screen, press and hold to access and then adjust the advanced settings.
  • Page 27: Alarms & Troubleshooting

    SET DAY to set Day to set Hour SET HOURS SET MIN 00 59 to set Minutes Show the pump software version. PDMS Show the PDMS software version. Show the pump serial number. LOCK THERAPY? Return to the LOCK THERAPY screen Switch off to save settings.
  • Page 28: Alarm Overview And Troubleshooting

    15.2 Alarm Overview and Troubleshooting Alarm Message Alarm Cause Solution The programmed total volume End administration or program a new nutritional therapy. has been completely given. Clear VOLUME GIVEN THIS THERAPY to start a new nutritional therapy. The target volume programmed is The reprogrammed target volume must be higher than the VOLUME GIVEN THIS less than or equal to the...
  • Page 29 LOW BATTERY The battery is almost fully dis- Connect pump AC power via pump AC adapter (see page 4). charged and should be plugged in Charge battery. immediately; otherwise Compat LOW BATTERY Ella will shut down shortly. ® Acknowledge the...
  • Page 30: Cleaning & Disinfection

    There is magnetic interference Remove source of magnetic interference or move pump to a different location. from an outside source. REMOVE MAGNET Internal equipment fault (system Turn pump off and restart to clear the error. If the error returns or repeats, do not ERXX fault).
  • Page 31: Maintenance

    No calibration is required. The Expected Service Life (ESL) is 10 years. ® The required maintenance interval for the Compat Ella pump is every 5 years or sooner if the battery no longer holds a charge. In the case of a hardware or ®...
  • Page 32: Limited Warranty

    19. Limited Warranty All service should be performed by authorized personnel. Unauthorized repairs will void the warranty. Pump, pole clamp and AC power adapter are guaranteed by the manufacturer for a period of 24 months from date of delivery. • The warranty covers repair and replacement of components proven to be defective in material or workmanship.
  • Page 33 21. Specifications Mechanical Storage • Dimensions: L:140 mm x B:140 mm x H:62 mm. • Storage Temperature: -20°C to +60°C • Weight: 499 g • Storage Humidity: 10% to 95%, non-condensing • Atmospheric Pressure: 50 to 106 kPa Operational • Flow Rate: 1 mL/h to 600 mL/h Power...
  • Page 34: Recycling And Disposal

    5 yr 12 h Above 5 years Service Pump* * Service interval for the Compat Ella pump is every 5 years. Battery will be replaced. ® In the case of a malfunction, contact your sales representative or an approved service provider to arrange for the pump to be repaired or replaced.
  • Page 35 (>95 % dip in U AC power quality should be that of a typical com- tions and voltage variations for 0.5 cycle for 0.5 cycle mercial or hospital environment. The Compat on power supply input lines Ella pump allows continued operation during AC ®...
  • Page 36 RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Compat Ella pump is used exceeds the applicable RF compliance level above, the Compat Ella pump should be observed ®...
  • Page 37 Separation distance according to frequency of transmitter Rated maximum output power of transmitter (W) 150 kHz to 80 MHz d=1.2√P 80 MHz to 800 MHz d=1.2√P 800 MHz to 2.5 GHz d=2.3√P 0.01 0.12 0.12 0.23 0.38 0.38 0.73 For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estima- ted using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
  • Page 38 Emissions test Compliance Electromagnetic environment - guidance The Compat Ella pump uses RF energy on for its in- ® RF emissions ternal function. Therefore, its RF emissions are very Group 1 CISPR 11 low and are not likely to cause any interference in ne- arby electronic equipment.
  • Page 39 Tel: 020-780020 Suisse Infoline: 0848 000 303 12272020 02MAR2020 P001T080_00 12367032 Axium Mtech S.A. Allée de la Petite Prairie 4, 1260 Nyon, Switzerland ® Unless stated otherwise, all trademarks are owned by Axium MTech S.A. or used with permission. www.compat.com...