33
CAUTION
, Totalizer value is lost.
[E-EEPTRP] EEPROM error in Trip Totalizer. Trip Totalizer value may be incorrect.
Action: Go to [DEFAULTS] correct Trip Totalizer memory block (under USER Code).
CAUTION,
Trip Totalizer is lost.
[E-EEPFAT] EEPROM error in File System.
Action: electronic module must be replaced. Please contact the supplier.
7.2 Alarm messages VZF II / VZFA II
The electronic module performs a self-test about every 5 minutes. If an alarm condition is
detected, [ALARM] message will appear on the display every 2 seconds.
Alarm messages are messages from the parameter settings
[nO ALmS] no alarm is active
Action: none.
[A-SEnSSI] alarm when sensor simulation is activated (Service Technician only).
Action: none. CAUTION Totalizer and Trip Totalizer accumulate the simulated value.
[A-o1 SI] alarm when output 1 simulation is activated.
Action: turn off when not required anymore.
[A-o2 SI] alarm when output 2 simulation is activated.
Action: turn off when not required anymore.
[A-ANA SI] alarm when Analog current loop outputsimulation is activated.
Action: turn off when not required anymore.
[A-o1 LI] alarm when output 1 is over its limit.
Action: adjust the settings of output1.
[A-o2 LI] alarm when output 2 is over its limit.
Action: adust the settings of output2.
[A-AnA LI] alarm when Analog current loop outputis over its limit.
Action: adust the settings of analog current loop output.
Need help?
Do you have a question about the CONTOIL VZF II and is the answer not in the manual?
Questions and answers
Alarm massege
The alarm messages for the Aquametro Oil & Marine CONTOIL VZF II indicate specific conditions detected by the electronic module during its self-test, which occurs approximately every 5 minutes. When an alarm condition is detected, the [ALARM] message appears on the display every 2 seconds. Some alarm messages include:
- [nO ALmS]: No active alarm. Action: None.
- [A-SEnSSI]: Alarm when sensor simulation is activated (for service technicians only). Action: None.
- [A-o1 SI]: Alarm when output 1 simulation is activated. Action: Turn off when no longer needed.
- [A-o2 SI]: Alarm when output 2 simulation is activated. Action: Turn off when no longer needed.
- [A-ANA SI]: Alarm when Analog current loop output simulation is activated. Action: Turn off when no longer needed.
The alarms indicate specific system conditions, such as active simulations, and require appropriate actions like disabling simulations when they are no longer needed.
This answer is automatically generated