Parameterizing; Main Menu - Aquametro Oil & Marine CONTOIL VZF II Mounting And Operating Instructions

Dn 15 – 50
Table of Contents

Advertisement

18
The display data and parameters are split into two menu groups:
Main Menu: displays measured data, accesses other menus, tests display segments and
displays error or alarm messages (if present).
Setup Menu: displays parameter settings for the display, output signals, additional
information about the flow meter and operating status.
When entering the access code, adjustment of parameters are possible.

5.3 Parameterizing

In order to adjust the parameters, scroll to the [SEtUP] item from the Main Menu and press the
Enter key
.
No code is required to view parameters.
To adjust any parameter in the Seutp menu, the device must be unlocked (
code. Press both keys (
Press Enter again and the first right digit will start flashing, which means, the unit is ready for the
entry of the first digit of the user code 1111.
Enter the first digit by using the Step key. Press the Enter key to accept the value (in our case 1)
and to move to the second right digit. Continue this way to enter all code digits.
At the end of the procedure a
parameters can be changed.
If no key is pressed within 1 minute, the device returns to the "home" display but the edit mode is
still active (15 minute timeout). Any entries that have not been completed by
pressing the Enter key are rejected.
Parameterizing the flow meter data
In order to guarantee accurate measurement, the electronic module of the flow meter
requires adjustment. During factory calibration, the data for nominal size and the exact measuring
chamber volume are entered for this purpose. These parameters cannot usually be changed
again.
If the display shows [UNCAL] the flow meter is not calibrated.
If the electronic counter has to be replaced, please mention the serial number of the defective
counter in your communication. Spare parts are set to the appropriate hardware size.
Reset of subtotalizers
To reset all subtotalizers, go to subtotalizer volume (*), press and hold both buttons for 4s until
the counter are reset.

5.4 Main Menu

The Main Menu displays most important measured data and accesses the setup menu.
The standard "Home" display of the Main Menu is the volume total [TOT], when Mass
compensation is turned on the "Home" display is mass total [TOT].
Quich return to "home": quickly press both key simultaneously.
+
) simultaneously for 4 seconds until [CodE0000] is displayed.
will be displayed. The flow meter is now in the edit mode and
) with the user

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CONTOIL VZF II and is the answer not in the manual?

Questions and answers

Joni
January 23, 2025

Alarm massege

User image 6791f6a9b017c
1 comments:
Mr. Anderson
February 10, 2025

The alarm messages for the Aquametro Oil & Marine CONTOIL VZF II indicate specific conditions detected by the electronic module during its self-test, which occurs approximately every 5 minutes. When an alarm condition is detected, the [ALARM] message appears on the display every 2 seconds. Some alarm messages include:

- [nO ALmS]: No active alarm. Action: None.
- [A-SEnSSI]: Alarm when sensor simulation is activated (for service technicians only). Action: None.
- [A-o1 SI]: Alarm when output 1 simulation is activated. Action: Turn off when no longer needed.
- [A-o2 SI]: Alarm when output 2 simulation is activated. Action: Turn off when no longer needed.
- [A-ANA SI]: Alarm when Analog current loop output simulation is activated. Action: Turn off when no longer needed.

The alarms indicate specific system conditions, such as active simulations, and require appropriate actions like disabling simulations when they are no longer needed.

This answer is automatically generated

This manual is also suitable for:

Contoil vzfa iiContoil vzoContoil vzoa

Table of Contents