January 18, 2025

What refrigerant type is used in Labcold RLDF0519 Underbench fridge

Summary of Contents for Labcold IntelliCold RLDF 19 Series

  • Page 1 IntelliCold® Pharmacy Refrigerator Operating Instructions - RLDF/G 19 www.labcold.com...
  • Page 2 Do not store explosive substances such as aerosol cans with fl ammable propellant in this appliance. Labcold RLDF/G refrigerators are designed for the storing of pharmaceuticals, medicines and laboratory products and are for professional use only by suitably WEEE: Disposal at end of life.
  • Page 3: Table Of Contents

    Guide......................13 Light Icon/Button....................14 Door Lock......................14 Digital Door Lock....................15 Micro SD Card......................18 Labcold are a UKAS calibration laboratory #8898 Caring for your Fridge..................20 8898 Troubleshooting....................21 We can calibrate your refrigerator AND your independent Useful Information...................22 thermometer IMPORTANT: These operating instructions contain information essential to the ...
  • Page 4: Before Use

    Before Use A Closer Look Thank you for purchasing a Labcold IntelliCold® Pharmacy Refrigerator. Your fridge has been purposely designed and precision built for the safe storage of vaccines and other temperature Controller Door Lock sensitive pharmaceuticals. Temperature Display Although great care has been taken in the manufacture of your refrigerator, damage can occur in transit.
  • Page 5: IntellicoldĀ® Controller

    IntelliCold® Controller Plug In and Switch On The IntelliCold® controller fitted to the front of your refrigerator has been designed by Labcold Before plugging in your Labcold refrigerator for the first time make sure the power cable is free from damage and check that your voltage corresponds to that on the serial plate.
  • Page 6: Muting The Alarm

    Muting the Alarm After setting the month you next need to set the day. The controller will now flash followed by 2 digits, for example d98. Adjust the digits using the arrow icons/buttons as above, so if it is to1H To mute the alarm, wait until the controller is showing the high alarm screen and touch June adjust the digits to 21.
  • Page 7: Viewing The Set Point

    When you access the menu the first menu option is to view the set point, see above. The set point is one of the parameters that you can not change but you can view it. Your Labcold IntelliCold® pharmacy refrigerator leaves the factory with a set point of 5oC as advised will flash, touch one of the arrows until the display shows F.
  • Page 8: Internal Light

    The battery alarm is enabled by default. This means if you put 4AA batteries (not supplied) in your refrigerator the alarm will not sound unless the batteries are faulty or removed. For Labcold extra security, the controller will alarm if the batteries are low or removed.
  • Page 9: Viewing Alarm Limits

    Green Book, Chapter 3 page 27 and you cannot change them. Should you need them to be this process for probe to2. You can tell which probe the high and low temperatures relate to 01256 705 580 or email service@labcold.com. altered please contact our service team on because the will display above the temperature displayed for the respective probe readings.
  • Page 10: Alarm Guide

    See page 9 for instructions on how to do that. The alarm will also sound when this screen is displayed. Labcold Your pharmacy and vaccine fridge is fitted as standard with a door lock with 2 keys unless the digital lock has been specified.
  • Page 11: Digital Door Lock

    Labcold digital combination lock is backed by a UK one year parts and labour warranty.
  • Page 12: Micro Sd Card

    Should you wish to use the lock as a digital lock again but have no record of the user code The option to download data from where the last download ended is only available if  01256 705 580 or service@labcold.com. please contact our service team on there has previously been at least one download.
  • Page 13: Caring For Your Fridge

    01256 705 580 service@ Cleaning labcold.com, and if you see it means the wrong Micro SD card for that refrigerator It is essential that you unplug the refrigerator from the mains to prevent electric ...
  • Page 14: Troubleshooting

    To register your warranty, please complete the enclosed warranty form or visit our web site doesn’t, please contact your retailer at www.labcold.com, click on the ‘Warranty Registration’ tab and follow the instructions. Full My fridge is making a gurgling noise...
  • Page 15 www.labcold.com...