Opening the door in an emergency following a power outage

WARNING
The steam sterilizer must be completely pressure free.
Failure to observe this provision can result in scalding / injury.
n
No steam may be permitted to leave from between the sterile filter and the rear panel of the
steam sterilizer.
n
The slide seal grip must be easy to actuate.
n
It must be possible to push the door approx. 2 mm to the rear with light pressure.
n
It is imperative that you allow the steam sterilizer to cool. Metal parts such as the door and
chamber can be hot.
Should it not be possible to open the door e.g. following a power outage, taking into account the safety
information specified above, proceed as follows:
1.
Switch off the steam sterilizer at the mains and
remove the plug from the socket.
2.
To effect emergency release of the door, position long
side of the lever between the door and the side wall of
the steam sterilizer. The curve points forwards; the
lever is at the level of the locking slider.
If the lever is in the guide, pull it forwards with your
right hand. Push the slide locking grip upwards with
your other hand.
3.
Open the door.
Display with door unlocked
12 Malfunctions