Canon CXDI-50G User Manual page 5

Digital radiography
Hide thumbs Also See for CXDI-50G:
Table of Contents

Advertisement

Immunity Test
Voltage dips, short
interruptions and volt-
age variations on
power supply input
lines
IEC 61000-4-11
Power frequency
(50/60Hz) magnetic
field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
IEC 60601
Compliance
Test Level
Level
<5% U
<5% U
T
(>95% dip in
(>95% dip in
U
) for 0.5
U
) for 0.5
T
T
cycle.
cycle.
40% U
40% U
T
(60% dip in
(60% dip in
U
) for 5
U
) for 5
T
T
cycles
cycles
70% U
70% U
T
(30% dip in
(30% dip in
U
) for 25
U
) for 25
T
T
cycles
cycles
<5% U
<5% U
T
(>95% dip in
(>95% dip in
U
) for 5 sec.
U
) for 5 sec.
T
T
3 A/m
3 A/m
Electromagnetic Environment – Guidance
Mains power quality should be that of a
T
typical commercial or hospital environ-
ment. If the user of the CXDI-50G requires
continued operation during power mains
interruptions, it is recommended that the
CXDI-50G be powered from an uninter-
T
ruptible power supply.
T
T
Power frequency magnetic fields should
be at levels characteristic of a typical loca-
tion in a typical commercial or hospital
environment.
(3)
Safety Information