5.2 Cleaning

5.2.1 CFRP (Carbon Fiber Reinforced Plastic) Plate of the Sensor Unit
Disinfect the CFRP plate of the sensor unit each time a different patient uses the instru-
ment. If the optional grid unit is being used, disinfect its surface.
WARNING
!
5.2.2 Cover
Clean the cover by the following procedure if it is dirty.
WARNING
!
(1) Shut down the system by referring to section 3.2.
(2) Turn OFF the power of remote switch, power box and control PC.
(3) Turn OFF the power of other connected instruments.
(4) Unplug the power cables of each instrument from the AC outlet.
(5) Wipe the cover with a piece of cloth soaked in neutral detergent diluted in water and
wrung dry.
(6) Wipe the cover with a piece of cloth soaked in water and wrung dry whenever neurtral
detergent has been used.
Wipe the CFRP (carbon fiber reinforced plastic) plate of the sen-
sor unit with ethanol or glutaraldehyde solution to disinfect them
each time a different patient uses the instrument, in order to pre-
vent infection.
If the optional grid unit is being used, disinfect its surface.
Please consult a specialist for the procedure for disinfection.
If you are using disinfectant other than those specified above, or
you are mixing another disinfectant with ethanol, please also
consult a specialist, because they may harm the CFRP plate and
grid unit.
When the instrument is going to be cleaned, be sure to turn OFF
the power of each instrument, and unplug the power cable from
the AC outlet.
Never use benzine, thinner or any other flammable cleaning
agents. Otherwise, fire or electric shock may result.