ESG-400
Error
Error message
no.
E180
Temperature above
limit
Switch off the ESG-400
and wait until it has
cooled down.
„Contact OLY"
E181
Temperature below
limit
Switch off ESG-400 and
wait until operating
temperature is reached.
„Contact OLY"
E182
Temperature above
limit
Switch off the ESG-400
and wait until it has
cooled down.
„Contact OLY"
E183
Error
E184
„Contact OLY"
E185
E186
Error
„Contact OLY"
E187
Increased HF leakage
current
Check if an instrument,
the neutral electrode or
patient is unintentionally
grounded
„Contact OLY"
E188
Excessive HF leakage
current
Check if an instrument,
the neutral electrode or
patient is unintentionally
grounded
„Contact OLY"
E189
Error
[...]
„Contact OLY"
E196
E197
Error
E198
„Auto-restart"
„Contact OLY"
7.022.211 / ISSUE 5
Possible cause
The generator (Generator Board) is
too hot.
The generator (HVPS Board) is too
cold.
The generator (HVPS Board) is too
hot.
Internal hardware error.
The high frequency leakage current
has exceeded the limit of 150 mA
for monopolar application or 100 mA
for bipolar application.
The high frequency leakage current
has exceeded the limit of 300 mA
for monopolar application or 200 mA
for bipolar application.
63 / 110
Remedial action
Switch off the electrosurgical generator
and wait until it has cooled down or
reached the specified operating
temperature.
Switch off the electrosurgical generator
and wait until it has reached the specified
operating temperature.
Switch off the electrosurgical generator
and wait until it has cooled down or
reached the specified operating
temperature.
Refer to E132
Send back to Olympus Service
Check if an instrument, the neutral
electrode or the patient is unintentionally
grounded.
Check if an instrument, the neutral
electrode or the patient is unintentionally
grounded.
Refer to 132
Refer to 132