Page 2
In the event of a guarantee claim, customers receive a repair cost estimate from ELPRO-BUCHS AG to obtain the corresponding consent before starting work. The customer will bear the transport costs incurred for any repairs carried out by ELPRO-BUCHS AG. The DAP (value added tax) is borne by ELPRO-BUCHS AG.
Table of Contents Table of Contents System Description ................... 6 Configuration and Evaluation ................7 Notes........................ 7 ECOLOG-PRO LBR Communication Module ......8 Power Supply Unit Circuit Diagram..............9 ECOLOG-PRO 4PT Pt100 Measurement Module ....10 Sensor Connection Cable for M12 Plug............11 Circuit Diagram ....................
Page 4
Symbols and Description Codes Used Information IMPORTANT INFORMATION AND WARNINGS DELETE Windows functions, for example Delete. <xxxxxxxxx> Syntax for placeholders. Placeholders are written in < > brackets. Reference to related chapter or document. Presentation only refers to part of the user interface.
Page 5
- This product is subject to CE marking. - The manufacturer guarantees that this product complies with the relevant guidelines: EN 61000-6-2 : 2001 and EN 61000-6-4: 2001 - FCC ID: Z45-E11645398 - IC: 9954A-E11645398 - This product must be disposed of in accordance with WEEE (Waste Electrical and Electronic Equipment, 2002/96/EC).
SYSTEM DESCRIPTION System Description This is a quick-start reference manual for installing the ECOLOG-PRO system. The manual also applies to the modules listed below. ECOLOG-PRO LBR is the communication and power supply module. It has 1 Ethernet and 1 power supply unit connection. Its purpose is for remote modules to communicate over the Ethernet network with elproMONITOR which is installed on a server provided by the cus- tomer.
SYSTEM DESCRIPTION Configuration and Evaluation Configuration, calibration, and monitoring are carried out with the elproMONITOR software. SM3031D, elproMONITOR operating instructions Notes Power supply unit Follow the safety and application instructions of the power supply unit. Only operate the device using the delivered power supply unit. Use of an incorrect power supply unit may cause damage to the device.
E COLOG-P RO LB R C OM MU NICATION MOD ULE ECOLOG-PRO LBR Communication Module Module The communication module performs 2 functions: • Supplies power to the measurement modules in use. • Acts as a connecting link between the measurement modules and the configuration or evaluation software.
E COLOG-P RO LB R C OM MU NICATION MOD ULE Power Supply Unit Circuit Diagram Power supply unit Chapter 7 1250 mA Power Supply Unit, Accessories Functional ground Connect the functional ground when shielded sensor cables are used. Document: EP6001Eb 9 - EN...
1 minute to 60 minutes Each channel is capable of logging at a separate interval. Date and time Internal system clock The clock is synchronized permanently to elpro- MONITOR system time. Module power supply Via internal system bus Data security...
ECOLOG-PRO 4P T P T100 ME ASU REMEN T MOD ULE Calibration data Data is stored in the internal non-volatile module memory. Content of package Measurement module Connecting hook, see module rear Information leaflet Validation certificate The plugs for sensor connection are not included in the scope of supply. Connecting hook Modules are fixed using a connecting hook on the rear to ensure a reliable connection.
1 minute to 60 minutes Each channel is capable of logging at a separate interval. Date and time Internal system clock The clock is synchronized permanently to elpro- MONITOR system time. Module power supply Via internal system bus Data security...
E CO L OG - PRO 4 MA 4- 2 0 MA ME ASU REMEN T MOD ULE Calibration data Data is stored in the internal non-volatile module memory. Content of package Measurement module Connecting hook, see module rear Information leaflet Validation certificate The plugs for sensor connection are not included in the scope of supply.
Pulse length 200 ms Minimum time interval to detect status change. Date and time Internal system clock The clock is synchronized permanently to elpro- MONITOR system time. Module power supply Via internal system bus Data security Internal non-volatile memory...
MODULE WIT H CONTACT INPUTS ECOLOG-PRO 4 DI Connecting hook Modules are fixed using a connecting hook on the rear to ensure a reliable connection. Contact Statuses The module detects the following data in each measurement sampling period: A) Number of edge changes B) Total duty cycle as percentage of measurement sampling period C) Present contact status Example...
M OD U L E W IT H C ON T A C T O U T P U T S E C O L O G- P R O 4 D O Module with Contact Outputs ECOLOG-PRO 4DO Module Module with 4 contact outputs for alarm forward- ing.
M OD U L E W IT H C ON T A C T O U T P U T S E C O L O G- P R O 4 D O Connecting hook Modules are fixed using a connecting hook on the rear to ensure a reliable connection. Contact Statuses THE POWER SUPPLY UNIT MUST BE CONNECTED TO ENABLE ALARM SIGNALS.
M OD U L E W IT H C ON T A C T O U T P U T S E C O L O G- P R O 4 D O Alarm Indicator Power Supply and Ground Alarm indicator power supply Terminal 5 of all 4 contacts is connected internally.
1250 MA POWER SUPPLY UNIT 1250 mA Power Supply Unit Plug-in power supply unit In order to power ECOLOG-PRO modules, the power supply unit is connected to the ECOLOG- PRO LBR communication module. A plug-in power supply unit is used as the standard power supply unit.
ECOLOG-PRO MODULE CONFIGURA TOR ECOLOG-PRO Module Configurator The ECOLOG-PRO Module Configurator is used to assign the parameters required for operation in the network to the ECOLOG-PRO LBR. The assignment procedure is described by the 3 steps below: Preparing for configu- Connect the ECOLOG-PRO LBR module to the power supply unit and the network.
Page 21
ECOLOG-PRO MODULE CONFIGURA TOR 2. IP address... Define the ECOLOG-PRO LBR for configuration by entering the IP address. 3. Hostname... Define the ECOLOG-PRO LBR for configuration by entering the host name. Standard: Host name = serial number Step 3 After a successful search, the input fields are completed with the current settings of the selected module and enabled for processing.
Accessories Part number Designation Description 802096 PSU_ECOLOG-PRO_EU Power supply unit with EU plug 802097 PSU_ECOLOG-PRO_GB Power supply unit with U.K. plug 802098 PSU_ECOLOG-PRO_US Power supply unit with U.S. plug 802100 CTR_ECOLOG-PRO Channel Plug to connect a sensor 802101 CTR_ECOLOG-PRO LBR Plug to connect a power supply unit 800508 CTR_M12F_IDC...
Need help?
Do you have a question about the ECOLOG-PRO Series and is the answer not in the manual?
Questions and answers