Error message
Venting valve error! Restart the device.
If the error occurs again call service!
Electronic error! Restart the device. If
the error occurs again, call service!
Sensor error! Restart the device. If the
error occurs again call service!
Device temperature error! Do not use
device. Call service!
Valve error! Do not use device. Call ser-
vice!
Calibration error! The device must be re-
calibrated. Call service!
Contamination! Fluid has penetrated
the Control Unit internally! Ongoing sur-
gery can be finished! Do not use device
for next procedure! Contact technical
service!
Critical message
Check connection of Tri-Lumen Filtered
Tube Set! Insufflation will shut down in
15s!
Cannula partly occluded! Ensure black
lines on distal cannula are visible inside
cavity!
Filter contaminated! Insert new tube
set!
Fluid Threshold in Filter Exceeded.
AirSeal Insufflation Stopped. Change
Tube Set to Complete Current Proce-
dure. Inspection Required to Assess Pos-
sible Damage to iFS Unit. Call for
Service.
Fluid Threshold in Filter Exceeded.
Smoke Evacuation Function Stopped.
Change Tube Set to Complete Current
Procedure. Inspection Required to
Assess Possible Damage to iFS Unit. Call
for Service.
Change gas bottle! Insufflation stopped! Low insufflation gas supply (bottle or
Change gas bottle!
Change gas bottle! AirSeal Insufflation
shut down in 100 s! Pressure limited to
12 mmHg!
Insufficient gas! Check gas supply and
restart device!
Check gas supply!
Check gas supply! Insufflation stopped!
Check gas supply! AirSeal Insufflation
will stop in 100 s! Pressure limited to
12 mmHg!
Gas level low. Prepare to change gas
bottle.
Cause
Valve malfunction.
Malfunctions in the electronic system.
Sensor malfunction.
Temperature in device too high.
Do not use device.
The device is not calibrated properly.
The device is contaminated with fluid.
Cause
Tube set improperly connected.
Tube or cannula has a blockage.
High fluid level of the fluid trap in the
filter housing
High fluid level of the fluid trap in the
filter housing during AirSeal mode.
High fluid level of the fluid trap in the
filter housing during Smoke Evacua-
tion mode.
house supply) during insufflation.
Low insufflation gas supply (bottle or
house supply).
Low insufflation gas supply (bottle or
house supply) during AirSeal® insuffla-
tion.
Device has no gas supply.
Device has no gas supply.
Low insufflation gas supply (bottle or
house supply) during insufflation.
Low insufflation gas supply (bottle or
house supply) during AirSeal® insuffla-
tion.
Low insufflation gas supply (bottle or
house supply) during insufflation.
Informational, Warning and Error Messages
Remedy
Restart the device. If the error occurs
again, call service!
Restart the device. If the error occurs
again, call service!
Restart the device. If the error occurs
again, call service!
Serious defect. Do not use device. Call ser-
vice!
Call service!
The device must be re-calibrated. Call ser-
vice!
The device must be checked by an autho-
rized service technician or clearly marked
with a label referring to the contamina-
tion and then twice enclosed in a safety
foil, sealed, and returned to the manufac-
turer for repairs.
Remedy
Connect the tube set correctly.
Determine the cause for the occlusion.
Check the positioning of the cannula.
Please insert a new tube set.
Finish procedure and call service.
Finish procedure and call service.
Check gas supply.
Check gas supply.
Check gas supply.
Check gas supply.
Check gas supply.
Check gas supply.
Check gas supply.
Check gas supply.