Page 1
CA 1:1 EVO.15 1:1 CA 1:5 EVO.15 1:5 CA 10:1 ENG Instructions for use other languages available on www.bienair.com/ifu 0120 REF 2100294-0001/2019.08...
ENG Instructions for use 1 Symbols 1.1 Description of symbols used Symb Description Symb Description Manufacturer. Reference number. CE Marking with number of the noti- Serial number. fied body. 0120 Electrical or electronic materials to CAUTION! be recycled. Refer to the accompanying Recyclable materials.
This product is intended solely for pro- & Intended Use fessional use. For use in dentistry: 2.1 Identification • CA 1:1, CA 10:1 and EVO.15 1:1: for prophylaxis, restoration and endo- Medical devices made entirely in Switzer- dontic procedures. land by Bien-Air Dental SA.
CAUTION 3 Precautions Any excess maintenance product (lubri- cant, cleaning and disinfection prod- for use ucts) present on the contra-angle could find its way into the electric brush This medical device accessory must be motor and adversely affect its opera- used by a competent person, in particu- tion.
135°C 135°C FIG. 1 (EVO.15 1:1 L) 4 Description 4.1 Overview FIG. 1 (1) Centring for the micromotor (2) Light outputs (3) Bur (not supplied) (4) Transmission ratio marking strip (5) Push-button 4.2 Technical data Transmission ratio 10:1 Standard coupling As per ISO standard 3964 and Micro-Series version.
Type 1 / ISO 1797 min. 12 mm Code 4 / ISO 6360-1 max. 22 mm Type 3 / ISO 1797 min. 12 mm Code 4-5 / ISO 6360-1 max. 21 mm FIG. 3 FIG. 2 3. Check it is correctly locked by gently pushing and pulling the bur.
6.1.2 Suitable maintenance products 6 Cleaning and Automatic cleaning-disinfection: servicing • Low alkaline or enzymatic detergent recommended for cleaning in a washer-disinfector for dental or sur- 6.1 Maintenance gical instruments (pH 6 - 9.5). Manual cleaning/disinfection Clean, lubricate and sterilise the device separately between patients.
EVO.15 CA 1:1, 1:5, 10:1 FIG. 4 FIG. 5 6.2 Preliminary cleaning 6.2.2 Preliminary bath cleaning CAUTION Preparation The preliminary bath cleaning should be FIG. 4 followed by automatic cleaning-disinfec- tion. 1. Disconnect the coupling device, 1. Totally submerge the device in a remove the bur (FIG.
FIG. 6 6.3 Cleaning-disinfection Washer-disinfector Carry out automatic cleaning-disinfec- 6.3.1 Manual cleaning/disinfection tion using an approved washer-disin- 1. Dip the device in a tank containing fector which complies with appropriate detergent and, using a soft standard 15883-1 (e.g. Miele G 7781/G brush, clean and disinfect the device, 7881 or Steris Hamo LM-25).
FIG. 8 FIG. 7 6.4 Lubrication CAUTION Do not use a sterilisation procedure 6.4.1 Verifying cleanliness other than the one described below. Visually inspect the device to ensure it Procedure is clean. If necessary, clean again using Enclose the device and its accessories in a soft brush.
50 kPa and 106 kPa. The warranty period is: • 12 months for the CA 10:1 range 7.1.1 Packing • 24 months for the CA 1:1, CA 1:5 Pack the instrument in packaging and EVO.15 ranges approved for steam sterilisation.
1600384-001 CA 1:1 L (light) 1600424-001 CA 1:1 (without light) 1600385-001 CA 10:1 L (light) 1600425-001 CA 10:1 (without light) 1600691-001 CA 1:1 L Micro-Series (light) 1600386-001 CA 1:5 L (light) 1600325-001 CA 1:5 (without light) 1600690-001 CA 1:5 L Micro-Series (light)