Beurer FT 65 Instructions For Use Manual

Beurer FT 65 Instructions For Use Manual

Multi-function thermometer

Advertisement

DE Multifunktions-Thermometer
Gebrauchsanweisung .................................... 2
EN Multi-function thermometer
Instructions for use ...................................... 18
FR Thermomètre multifonction
Mode d'emploi ............................................. 33
ES Termómetro multifunción
Manual de instrucciones .............................. 48
IT
Termometro multifunzione
Istruzioni per l'uso ........................................ 63
TR Fonksiyonlu termometre
Kullanım kılavuzu ......................................... 78
RU Многофункциональный термометр
Инструкция по применению ...................... 93
PL Termometr wielofunkcyjny
Instrukcja obsługi ....................................... 109
FT 65

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the FT 65 and is the answer not in the manual?

Questions and answers

Summary of Contents for Beurer FT 65

  • Page 1 FT 65 DE Multifunktions-Thermometer Gebrauchsanweisung ........2 EN Multi-function thermometer Instructions for use ........18 FR Thermomètre multifonction Mode d’emploi ..........33 ES Termómetro multifunción Manual de instrucciones ......48 Termometro multifunzione Istruzioni per l’uso ........63 TR Fonksiyonlu termometre Kullanım kılavuzu .........
  • Page 2: Table Of Contents

    Thank you for choosing one of our products. Our name stands for high-quality, thoroughly tested products for applications in the areas of heat, gentle therapy, blood pressure/diagnosis, weight, massage and air. With kind regards, Your Beurer team Contents 1. Signs and symbols ............19 8.
  • Page 3: Signs And Symbols

    1. Signs and symbols The following symbols appear in these instructions for use and on the device: WARNING Warning notice indicating a risk of injury or damage to health. IMPORTANT Safety note indicating possible damage to the unit/accessory. Note Note on important information. Observe the instructions for use Application part, type BF Disposal in accordance with the Waste Electrical and Electronic Equipment EC Directive –...
  • Page 4: Notes

    Manufacturer Dispose of packaging in an environmentally friendly manner Protect from moisture Certification symbol for products that are exported to the Russian Federation and members of the CIS 2. Notes Safety notes Warning: • To measure temperature, insert the sensor tip of the thermometer carefully into the ear. •...
  • Page 5 • Protect the batteries from excessive heat. • Risk of explosion! Never throw batteries into a fire. • Do not charge or short-circuit batteries. • If the device is not to be used for a long period, take the batteries out of the battery compartment. •...
  • Page 6 • This device complies with the EU Directive 93/42/EEC concerning medical products, the Medizinproduktege- setz (German Medical Devices Act), the ASTM (American Society for Testing and Materials) E 1965 - 98 and the European Standard EN 12470-5: Clinical thermometers – Part 5: Requirements for infrared ear thermometers (with maximum device) ISO 80601-2-56 and the European Standard EN 60601-1-2 (in accordance with CISPR11, IEC61000-4-2, IEC61000-4-3, IEC61000-4-8), and is subject to particular precautions with regard to electromag- netic compatibility.
  • Page 7: Unit Description

    3. Unit description Overview 1 Forehead/cover cap 6 LCD 2 Forehead thermometer mode button Forehead 7 Sensor tip with lens (cover cap removed) Stirn 3 Ear thermometer mode/on button Ear / 8 Memory button 4 LED (red) (high temperature alarm) 9 Set button 5 LED (green) 10 Battery compartment lid...
  • Page 8: Initial Use

    The batteries are already inserted in all new devices. Before first use, remove the protruding plastic insulating strip from the battery compartment. The thermometer will then switch on automatically. Further information and FAQs available at: http://www.beurer.com 5. Setting the device Set the unit for the temperature display as well as the date and the time.
  • Page 9: Measuring

    Taking the temperature gives a measurement that provides information about a person’s current body temperature. If you are uncertain about interpreting the results or if the value is abnormal, please consult your doctor. This also applies in the case of slight temperature changes if there are other symptoms of illness such as agitation, severe sweating, flushed skin, fast pulse rate, tendency to collapse, etc.
  • Page 10: Thermometer

    Measuring body temperature in the ear • Some people produce different readings in their left and right ear. In order to record temperature changes, always measure a person’s temperature in the same ear. • The ear thermometer may be used by children only under adult supervision. Measurement is usually possible over the age of 6 months.
  • Page 11: Thermometer

    Make sure that the sensor tip and also the ear canal are clean. As the ear canal is slightly curved, you have to pull the ear slightly up and backwards before inserting the sensor tip. This is important so that the sensor tip can be pointed directly at the eardrum.
  • Page 12: Changing The Batteries

    Measuring surface temperature • Press the button for 1 second to switch the thermometer on. Following a successful self test, the device emits two short beeps. • Then press the buttons simultaneously for 3 seconds to switch to the Object temperature mode. symbol appears in the display.
  • Page 13: Storage And Maintenance

    Note: • When changing the batteries, use batteries of the same type, make and capacity. • Always replace all batteries at the same time. • Do not use rechargeable batteries. • Use batteries free from heavy metals. 1. Loosen the screw on the battery compartment lid and pull the lid off downwards.
  • Page 14: Disposal

    If using the device for commercial medical purposes, it must be regularly tested for accuracy by appropriate means. Precise instructions for checking accuracy may be requested from the service address. Name and model FT 65 Measurement range Ear/Forehead thermometer mode: 93.2 °F – 109.4 °F (34°C – 43°C) Object temperature mode: 32 °F –...
  • Page 15: Trouble-Shooting

    Interval between 2 measurements At least 5 seconds Clinical repeat precision Ear: Children, 1 – 5 years: ±0.14 °F (±0.08 °C) Adults: ±0.13 °F (±0.07 °C) Forehead: Children, 1 – 5 years: ±0.13 °F (±0.07 °C) Adults: ±0.14 °F (±0.08 °C) Measurement units °Celsius (°C) or °Fahrenheit (°F) Ambient operating conditions 59 °F –...
  • Page 16: Warranty / Service

    In case of returning the unit please add a copy of your receipt and a short report of the defect. The following warranty terms shall apply: 1. The warranty period for BEURER products is either 5 years or- if longer- the country specific warranty period from date of Purchase.