Topcon CT-80 Troubleshooting Manual

Topcon CT-80 Troubleshooting Manual

Computerized (non-contact) tonometer
Hide thumbs Also See for CT-80:

Advertisement

T
rouble
CT
-80
(also applies to
CT
The Topcon
Computerized (Non-Contact) Tonometer brings
-80
quick and accurate tonometry readings to your fingertips.

1. Monitor screen does not work.

• Secure power
cable in AC plug.

2. Paper does not come out.

• Check paper supply. Ensure only Topcon paper is used (Part #42330 9003B).
• Check for paper jam and remove if jammed.

3. Machine body does not move.

• Check Clamping Knob under instrument. Loosen if found fastened.
• Check Fixing Knob near the joystick.
Turn to the left to loosen; turn to the
right to fasten.
CT-80
S
hooTing
CT
-60
• Plug power cable
into machine body.
)
• Check the fuse.
Page 1 of 2
®

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CT-80 and is the answer not in the manual?

Questions and answers

Summary of Contents for Topcon CT-80

  • Page 1 AC plug. into machine body. 2. Paper does not come out. • Check paper supply. Ensure only Topcon paper is used (Part #42330 9003B). • Check for paper jam and remove if jammed. 3. Machine body does not move.
  • Page 2 Warning: Regular alcohol must not be used. Moisten the Topcon q-tip with solution. Insert the q-tip into the nozzle. Lightly touch the glass surface while turning the q-tip. Replace the q-tip with a new one and wipe the surface in a similar fashion.