WARNING
To reduce the risk of infection, illness, and/or injury from contamination and/or improper use, it is important to:
Clean and/or Disinfect your Handset Before First Use.
Rinse and Clean your Handset Immediately after Each Use (see Steps E1-E8).
Disinfect Your Handset Daily or as Recommended by Your Physician (see Steps E9-E10).
CAUTION
The Aerosol Head contains very small holes. To reduce the risk of damage to the Aerosol Head and Handset:
Do Not clean the Aerosol Head in a dishwasher.
Do Not clean the Aerosol Head with brushes or abrasives.
Do Not disinfect the Aerosol Head in the microwave oven.
Do Not apply liquid dish soap directly to the Aerosol Head.
Do Not use "white" liquid dish soaps or antibacterial liquid soaps as these may contain additives harmful to the Aerosol Head.
Do Not wash the Controller, Connection Cord or AC Power Supply.
Use clear liquid soap designed for washing dishes.
Use only distilled water to disinfect by boiling.
Follow manufacturer directions when using a chemical disinfectant, such as Control III®.
Handset Maintenance Summary (Details Can Be Found on the Following Pages)
Instruction When
Rinse
After each
use
Clean
After each
use
Disinfect
Daily or as
recommended
by your
physician.
Unclog
When needed
Parts cleaned
 Aerosol Head
 Medication Cap
 Blue Valve
 Reservoir and Chamber
 Mouthpiece
 Aerosol Head
 Medication Cap
 Blue Valve
 Reservoir and Chamber
 Mouthpiece
 Aerosol Head
 Medication Cap
 Blue Valve
 Reservoir and Chamber
 Mouthpiece
Aerosol Head
E. How To Clean and Maintain Your eRapid
Method
How Long
Rinse with warm
10 seconds
running tap water.
Soak in warm
5 minutes
soapy water.
Boil in distilled
5 minutes
water
or
Soak in Control
10 minutes
III® disinfectant.
Run the
5-10
Controller's
minutes
Easycare cycle
with distilled
water and the
Easycare
cleaning aid.
7
Comments
Rinse each side of the Aerosol Head
for At Least 10 Seconds.
Soak longer if Handset has dried
medicine or if visibly dirty.
Rinse thoroughly after disinfecting with
Control III®.
Disinfect after completing Easycare
cycle.
Do not exceed 30 Easycare cycles per
Aerosol Head.