    C. Getting Started These Instructions for Use contain information and safety D. Using Batteries or the A/C Power Supply precautions for the Lamira nebulizer that is made just for the medicine ARIKAYCE (amikacin liposome inhalation ® E. Maintaining Your Lamira suspension).
  • Page 3: Safety Precautions

    Do not use while bathing. contamination of the Handset, you must also disinfect the Handset Do not reach for the Lamira Nebulizer System if it has fallen into • and Aerosol Head after every cleaning. See Section G for water or other liquid.
  • Page 4: Getting Started

    Step C2: Check your Lamira Nebulizer System package to make sure you have the items shown below. Note that the package contains a Handset that should be set aside as a spare. If anything looks damaged, contact the Arikares Support Program at 1-833-ARIKARE (1-833-274-5273).
  • Page 5: Using Batteries Or The A/C Power Supply

    Using Batteries or the A/C Power Supply The Lamira is made to be used with “AA” batteries or with the A/C Power Supply. Using Batteries: Four (4) high quality “AA” batteries should provide 2 hours of total use. Step D1: Open the Battery Door on the Controller by placing your thumb on the tab of the Battery Door and firmly pulling the tab to open the Door (D-1).
  • Page 6: Maintaining Your Lamira

    • Aerosol Head first use water. environment. • Medication Cap and Seal and after • Blue Valve each use • Medication Reservoir and Aerosol Chamber • Mouthpiece Questions about Lamira? Contact the Arikares Support Program at 1-833-ARIKARE (1-833-274-5273) Reference ID: 4327567...
  • Page 7: Cleaning Your Lamira

    Step F3: Remove the Mouthpiece from the Aerosol Chamber by pulling straight off (F-3). The Blue Flap must still be attached to the mouthpiece as shown in the picture. Questions about Lamira? Contact the Arikares Support Program at 1-833-ARIKARE (1-833-274-5273) Reference ID: 4327567...
  • Page 8 Aerosol Head helps to prevent clogging and makes sure the Aerosol Head works properly. Never use a brush or any other object to clean the Aerosol Head. Questions about Lamira? Contact the Arikares Support Program at 1-833-ARIKARE (1-833-274-5273) Reference ID: 4327567...
  • Page 9 Controller. Caution Never let the Controller come in contact with water or cleaning agents. If liquid does get into the Controller, contact the Arikares Support Program. Questions about Lamira? Contact the Arikares Support Program at 1-833-ARIKARE (1-833-274-5273) Reference ID: 4327567...
  • Page 10: Disinfecting Your Lamira

    Reassemble just before taking your next treatment. This is to make sure the Blue Valve will not become damaged. Questions about Lamira? Contact the Arikares Support Program at 1-833-ARIKARE (1-833-274-5273) Reference ID: 4327567...
  • Page 11: Assembling Your Lamira Handset

    Blue Valve inside of the Aerosol Chamber. The Blue Valve should be placed on top of the Aerosol Chamber. AEROSOL CHAMBER VALVE FLAP VALVE FLAPS Questions about Lamira? Contact the Arikares Support Program at 1-833-ARIKARE (1-833-274-5273) Reference ID: 4327567...
  • Page 12 “snap” (H-6). If you do not hear a snap, open the Handset and check that the Blue Valve is seated properly (See Step H-4). Questions about Lamira? Contact the Arikares Support Program at 1-833-ARIKARE (1-833-274-5273) Reference ID: 4327567...
  • Page 13 Push the round end of the Connection Cord into the plug inlet port located under the digital display (H-9). Place the Controller with the attached Handset on a clean, flat, stable surface. Questions about Lamira? Contact the Arikares Support Program at 1-833-ARIKARE (1-833-274-5273) Reference ID: 4327567...
  • Page 14: Taking Arikayce

    Step I6: Make sure your Handset is placed on a clean, flat, stable surface. Pour 1 vial of ARIKAYCE into the Medication Reservoir (I-6). Do not use more than 1 vial for each treatment. MEDICATION RESERVOIR Questions about Lamira? Contact the Arikares Support Program at 1-833-ARIKARE (1-833-274-5273) Reference ID: 4327567...
  • Page 15 Mouthpiece until your treatment is complete. Your treatment should take about 14 minutes, but could take up to 20 minutes. Questions about Lamira? Contact the Arikares Support Program at 1-833-ARIKARE (1-833-274-5273) Reference ID: 4327567...
  • Page 16 Aerosol Head during the cleaning process. Follow the instructions in Steps F6 through F9 and replace the Change your Aerosol Head after 7 uses. Aerosol Head with a new one. Questions about Lamira? Contact the Arikares Support Program at 1-833-ARIKARE (1-833-274-5273) Reference ID: 4327567...
  • Page 17: Frequently Asked Questions

    (4) the Medication Reservoir and complete or after 20 minutes. If the Lamira turns off after Aerosol Chamber are closed properly and snapped 20 minutes, check the Medication Reservoir. If it is empty, together.
  • Page 18: Troubleshooting

    Lamira beeps 1 time and LCD shows this symbol Steady Controller has paused. To resume press On/Off button. Flashing Press button to start treatment. Brief Functioning properly, the Controller has started properly. Questions about Lamira? Contact the Arikares Support Program at 1-833-ARIKARE (1-833-274-5273) Reference ID: 4327567...
  • Page 19 The On button was accidently pressed for 10 seconds, which activated the easycare function. The easycare is an accessory that assists in cleaning the Aerosol Head but it is not used for your Lamira. Press the On button briefly to turn-off the Controller and stop the easycare function.
  • Page 20: Specifications

    For patent information: www.PARI.com/ip ©2018 PARI Pharma GmbH. All rights reserved. The PARI logo is a registered trademark of PARI GmbH. Altera®, eRapid®, and eFlow® logo are registered trademarks of PARI Pharma GmbH. Lamira™ is a trademark of PARI Pharma GmbH. ARIKAYCE is a registered trademark of Insmed Incorporated.. Touchspray® made under license from the Technology Partnership.