Summary of Contents for Precision Medical Live Active Five

  • Page 1 Portable Oxygen Concentrator...
  • Page 2 The Model [REF] and Serial Number [SN] label are located on the bottom of the device. Record the Model and Serial Number in the space provided below. This will be helpful if you need to contact Precision Medical about your POC in the future. For the latest information, visit us at www.precisionmedical.com.
  • Page 3: Table Of Contents

    Contents Device Overview ..........1 Powering On the POC .........16 About Your Live Active Five Portable Oxygen Adjusting the Pulse Setting ......16 Concentrator (POC) ..........1 Start-up Period ............. 17 Intended Use and Indications....... 1 Reading the Display Screens ......17 User Profile ..............
  • Page 4: Device Overview

    User Profile The Live Active Five POC is to be used by a trained patient and/or caregiver in or outside the home that requires prescribed supplemental oxygen. The user should have a visual acuity of 0 on the log Minimum Angle of Resolution (log MAR) scale or 6/6 (20/20), corrected if necessary.
  • Page 5: General

    Remove contents from packaging and inspect for damage. If there is any damage, DO NOT USE and contact your Provider. The POC package includes the following: Accessories Carry Bag (508623) Live Active Five POC (PM4155) Lithium Ion Battery (508561) *Extra Battery with Power Bundle...
  • Page 6: Symbols

    Symbols WARNING CAUTION CAUTION Indicates that personal safety Indicates a potentially Used without the safety alert of the patient may be involved. hazardous situation which, symbol indicates a potentially Disregarding a warning could if not avoided, may result in hazardous situation which, result in significant injury.
  • Page 7: Warnings

    Warnings WARNING Precision Medical Inc, and your equipment The device must be used in dry conditions. Do not submerge, operate under water, bathe or provider are accountable for ensuring the swim while in use. compatibility of the POC and all of the parts or accessories used.
  • Page 8: Features

    Features Control Panel / Component Descriptions Oxygen Concentrator 300 HELD DRIVE NORTHAMPTON, PA 18067 REF PM4155 MADE IN USA Matrix SN XXXXXXXX Here DC Input: 12 - 18VDC 5A (01)00855887006887(21)XXXXXXXX IP22 MEDICAL - GENERAL MEDICAL EQUIPMENT AS TO ELECTRICAL SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH ANSI/AAMI ES 60601-1 (2005) + AMD (2012), CAN/CAS C22.2 No.
  • Page 9: Power Options

    Oxygen Outlet Fitting - Connects the nasal cannula to the device. Display Screen - Displays pulse setting, power and battery status, and alert messages. Pulse Selection Buttons (1 - 5) Refer to Adjusting the Pulse Setting in Usage for more details. Paused Alert Button - Press to mute the audible signal.
  • Page 10: Accessories And Replacement Parts

    Live Active Five Battery Charger Bundle with AC Power Adapter and Battery – 508650 The Live Active Five Battery Charger Bundle comes with one Live Active Five Lithium Ion Battery so you can always have a charged battery ready to go. Additional batteries can be purchased separately.
  • Page 11 Live Active Five DC Car Adapter – 508558 The Live Active Five DC Car Adapter is designed for use with your Live Active Five POC. The DC cord allows you to power your concentrator and charge the battery simultaneously from a standard DC car outlet in your car, RV or boat.
  • Page 12: Setup

    Setup Follow these instructions to ensure safe operation of the device. Note: The device, its parts and accessories are intended for use by a single user and should be cleaned / disinfected before use on a new user. Powering the POC CAUTION Ensure battery is fully charged prior to first use.
  • Page 13: Reading The Battery Gauges

    Reading the Battery Gauges Installed Battery The battery gauge on the display shows the status of the installed battery. Battery time remaining with the device running without external power (may take up to 3 minutes to display). Battery charge level with the device running and external power connected.
  • Page 14: Battery Removal

    Battery Removal With an external power source connected, the battery can be removed without affecting the operation of the device. If not connected to an external power source, first power off the device. To remove the battery, pull latch until the battery disengages. Lift battery out of the device.
  • Page 15: External Dc Power

    External DC Power CAUTION Operating the device without the engine running may drain the vehicle’s battery. Ensure the 12-volt Accessory Jack is adequately fused for the device power requirements. The accessory jack should have a minimum 8 Amp (96 Watts) fuse. DC Power Plug (Plugs into DC outlet in vehicle) DC Power Adapter Power Output Plug (Plugs into device)
  • Page 16: Carry Bag Use And Setup

    Carry Bag Use and Setup The carry bag allows you to take your device with you as you go about your normal daily activities. It protects the POC and can be used with either the carry handle or shoulder strap when using the device. WARNING If the device is not placed in the carry bag properly, the air inlets and outlets will be blocked causing the device to overheat during operation and shut off.
  • Page 17: To Place The Poc Into Its Carry Bag

    Incorrect placement in Carry Bag Correct placement in Carry Bag Correct positioning of outlets and external power input connector To Place the POC into its Carry Bag 1. Insert the device into the top of the carry bag and pull the bag up around it. Check to ensure that the air inlets and outlets and external power input connector align with their openings in the carry bag.
  • Page 18: Usage

    Usage DO NOT use this device or optional accessories without first completely reading and understanding the instructions. DO NOT use this product in any way other than described in the specifications and intended use sections. Location and Operating Position The POC is intended to be used in dry locations. Use the device in a well ventilated location free of pollutants and fumes.
  • Page 19: Removing The Cannula From The Outlet

    ALWAYS confirm prescribed dose before administering to user and monitor on a frequent basis. The settings of other models or brands of oxygen therapy equipment may not correspond with the settings of the Live Active Five POC. The settings of the device may not correspond with continuous flow oxygen.
  • Page 20: Start-Up Period

    A visual and audible alarm will display and sound momentarily after the device has been powered on. “PRECISION MEDICAL” followed by device Serial Number and Hours of Operation will appear on the display while the device starts up. The display will then indicate the selected setting and the battery percentage remaining.
  • Page 21: Breathing With The Poc

    Breathing with the POC As you breathe through the nasal cannula, a pulse of oxygen will be delivered and display each time the unit senses an inhalation. If no breath is detected for 60 seconds, the alert will display and the device will enter into auto pulse delivery.
  • Page 22: Traveling With The Poc

    POC while on board the aircraft. Prior to the flight, ensure your Live Active Five is clean and in good working condition. Bring enough charged batteries with you to power your Live Active Five for at least 150% of the expected duration of your flight, including ground time before and after the flight, security screenings, connections and possible delays.
  • Page 23: Recommended Preventative Maintenance

    Except for tasks described in this manual, only trained personnel should perform preventive maintenance or performance adjustments on the device and its equipment. Users should contact your provider or Precision Medical for service. Service Life The expected life for the device is 5 years. The sieve beds and batteries have an expected life of 1 year.
  • Page 24: Replacing The Air Outlet Filter

    Replacing the Air Outlet Filter CAUTION Replace with Precision Medical Inc Outlet Filter 508583 only. The outlet filter is intended to protect the user from small particles in the oxygen gas flow. This filter is conveniently located behind the removable cannula outlet fitting. Precision Medical recommends only trained personnel replace the filter between users.
  • Page 25: After Use

    Fire Stop Check Valve to stop the flow of gas towards the user in the case that the cannula becomes ignited. The Fire Stop Check Valve should be located as close to the user as reasonably practicable. Fire Stop Check Valve - Precision Medical Part # 507706 2” Tubing Connector - Precision Medical Part # 507707...
  • Page 26: Troubleshooting

    Troubleshooting Technical Alerts WARNING Failure to resolve an alert condition may cause the device to shut down. Technical Alerts Description The device monitors various internal components and compares them to acceptable limits. An alert is generated when the acceptable limit has been exceeded. Alerts are classified as Low Priority Technical Alert Conditions.
  • Page 27 WARNING DO NOT disassemble or attempt to repair. There are no user serviceable parts inside. Contact your Provider or Precision Medical, Inc. for service. If the device fails to operate properly, refer to the following charts for possible causes and solutions. If necessary, contact your Provider or Precision Medical, Inc.
  • Page 28 POC Pulse Delivery Alerts Symptom Probable Cause Solution(s) Device does not deliver a Make sure the tubing is pulse of oxygen when the user Cannula tubing kinked, connected properly to the inhales. blocked or twisted. oxygen outlet port and that it is free of any obstructions.
  • Page 29: Shut Down Alerts

    Press setting to restart. If there are 5 unsuccessful restart attempts in less than 5 minutes, service of the device will be required. Connect to an alternative oxygen source and contact your home care provider or Precision Medical.
  • Page 30: Technical Data

    Technical Data Specifications General When measuring any published tolerance, be sure to consider the uncertainty of the measurement of the measuring equipment. Gas volume and flow rate specifications for gas delivered to the user are expressed at STPD (standard temperature and pressure, dry). Electrical Requirements: AC to DC Power Supply: Input: 100–240 VAC, 50–60 Hz, <2.0A Output: 18Vdc up to 5.56A...
  • Page 31: Regulatory Listings

    Expected Service Life of the Device: 5 Years *May vary based on age of device. Specifications are subject to change without prior notice. Regulatory Listings The Live Active Five POC has been designed, tested and certified to the following regulatory standards: ANSI/AAMI 60601-1; Ed: 3.1 IEC 60601–1–6 IEC 60601-1-2: 2014 IEC 60601–1–8...
  • Page 32: Electromagnetic Compliance (Emc)

    Electromagnetic Compliance (EMC) Guidance and manufacturer’s declaration - electromagnetic immunity The Live Active Five POC is intended for use in the electromagnetic environment specified below. The user of the device should make sure it is used in such an environment.
  • Page 33 Immunity Test IEC 60601 Compliance Electromagnetic Test Level Level Environment - Guidance Surge ± 1 kV line(s) to ± 1 kV line(s) to Mains power quality should be that of a line(s) line(s) typical commercial or hospital environment. ± 2 kV line(s) to ±...
  • Page 34 Guidance and Manufacturer’s Declaration – Electromagnetic Emissions The Live Active Five POC is intended for use in the electromagnetic environment specified below. The user of the device should assure that it is used in such an environment.
  • Page 35: Limited Warranty

    Precision Medical Limited Warranty Precision Medical, Inc. (PMI) warrants that the Live Active Five Portable Oxygen Concentrator (the Product) and Accessories will be free from defects in materials and workmanship for the following period: Product Period Live Active Five Portable Oxygen Concentrator...