Summary of Contents for Precision Medical Easy Go Vac PM66 Series
Page 1
U S E R M A N U A L PM66 Series PM66Li (Shown) SAVE THESE INSTRUCTIONS Read this manual before operating the device, save this manual for future reference. 300 Held Drive Tel: (+001) 610-262-6090 Northampton, PA 18067 USA Fax: (+001) 610-262-6080 ISO 13485 Certified www.precisionmedical.com...
Limited Warranty ad Limitation Of Liability ..............19 Receiving / Inspection Remove the Precision Medical, Inc. Easy Go Vac Aspirator from packaging and inspect for damage. If there is any damage, DO NOT USE and contact your Provider. Intended Use The Easy Go Vac Aspirator provides a portable, AC/DC powered medical vacuum source.
Safety Information - Warnings and Cautions Indicates an imminently hazardous situation which, if not DANGER avoided, will result in death or serious injury. Indicates a potentially hazardous situation which, if not WARNING avoided, could result in death or serious injury. Indicates a potentially hazardous situation which, if not CAUTION avoided, may result in minor or moderate injury.
Page 4
The PM66 Series Aspirators meet the requirements of IEC 60601-1-2 and are considered safe from producing electromagnetic interference with other devices when used normally in accordance with their intended use. Should the aspirator cause any interference, try repositioning it or moving it to another location. DANGER To reduce the risk of Electrocution: •...
Specifications For technical specifications contact Precision Medical, Inc. or visit www.precisionmedical.com Suction Canister: 800ml/1200ml Max Volume - Lid has a float shut off valve that protects the Aspirator from overfill. Vacuum is not to exceed 25 inHg (635 mmHg) Patient Tubing: 1/4 in.
Guidance and Manufacturer’s Declaration – Emissions All Equipment and Systems This device is intended for use in the electromagnetic environment specified below. The customer or user of this device should assure that it is used in such an environment. The PM66 Series Aspirators meet the requirements of IEC 60601-1-2 and are considered safe from producing electromagnetic interference with other devices when used normally in accordance with their intended use.
Page 7
For transmitters rated at a maximum output power not listed above, the recommended separation distance D in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Component Identification CAUTION If for any reason any label becomes illegible or lost, contact Precision Medical, Inc. Vacuum Vacuum Control Knob Patient Port On/Off Gauge Filter Switch 1200ml Suction Battery Canister Level (Built in overfill Indicator protection device) Charge Status...
Page 9
PM66Li (Lithium Ion) PM66 200 (Lithium Ion) Power Source When light is GREEN the indicated power source is being used. Charge Status (if equipped) LOW - The power switch is ON, AC not plugged in and light is AMBER, the Aspirator Internal Battery requires charging.
Operating Instructions 1. Connect to external power 1a. DC outlet and connector source (AC or DC) DC outlet DC connector 1b. Charge Device - Internal Battery 2. Connect suction tubing to canister Push together Device may be used while conncected to AC Ensure all connections are tight.
Operating Instructions (continued) 7. Power “OFF” when procedure Device in carry bag completed Push button CAUTION • Verify battery charge level prior to use. • Charging Light is on when Aspirator is plugged in and is in the “OFF” ( ) position only. •...
Maintenance Vacuum/Flow verification should be checked by the device provider as necessary - Recommended annually or between patients. Cleaning Instructions Device with Disposable 800cc Collection Canister Reference AARC Clinical Practice Guideline - Suctioning of the Patient in the Home, section HCS 13.0 INFECTION CONTROL Preparation 1.
Page 13
2. Wash canister, lid, o-ring, and overflow valve in a solution of warm water with a mild, liquid detergent (e.g. Dawn or Palmolive) and rinse with clean, warm tap water. 3. For single patient use: After washing, disinfect using one of the following method: Soak in 1 part vinegar (≥5% acetic acid concentration) to 3 parts water (131°F-149°F/55°C-65°C) solution for 60 minutes.
Canister Removal/Assembly (1200cc Reusable) 1. Grasp filter and pull apart 2. Tilt off base to remove 3. Remove lid, empty & clean 4. Place lid on canister 5. Align indicators on lid and canister 6. Tilt and place canister on device base 7.
Inline Filter Replacement CAUTION • When installing inline filter be sure the side that has small nubs and says “OUT” is facing toward inlet port on device. • Filter cannot be cleaned. If it becomes contaminated or clogged, it must be replaced. 800cc Disposable Canister Inlet Adaptor Kit Slightly twist and pull up...
Returns Returned products require a Returned Goods Authorization (RGA) number, contact Precision Medical, Inc. All returns must be properly packaged to prevent shipping damage. Precision Medical, Inc. will not be responsible for goods damaged in transit. Refer to Precision Medical, Inc. Return Policy available on the Internet, www.precisionmedical.com.
Troubleshooting If the Aspirator fails to function, consult this Troubleshooting section. If the problem cannot be corrected, consult your Provider. There are no user serviceable parts inside this device. DO NOT open case. Problem Probable Cause Remedy Aspirator will 1. On/Off Switch in “OFF” 1.
Only use accessories designed for use with this Aspirator. • Use tubing intended for suction only. • Use Collection Canisters designed for use with this Aspirator only. • Use Precision Medical: Inlet Filter, External AC & DC Power Cord & Carry Bag Only Description...
ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES. The representatives of Precision Medical, Inc. or any retailers are not authorized to make oral warranties about the merchandise described in this contract, and any such statements shall not be relied upon and are not part of the contract for sale.