Summary of Contents for PerkinElmer Caliper IVIS Spectrum
Page 1
® IVIS Spectrum In Vivo and In Vitro 3D Optical Imaging Bioluminescence and Fluorescence Spectral Unmixing Transmission and Reflectance Imaging Hardware Manual March 2012...
IVIS Spectrum Hardware Manual ® Welcome About the IVIS Spectrum Imaging System ....1 IVIS Imaging Systems Technology Support ....3 1.1 About the IVIS Spectrum Imaging System The IVIS Spectrum Imaging System is a high-sensitivity, low noise, in vivo...
Page 8
1. Welcome camera XGI-8 Gas Anesthesia System High resolution monitor Imaging chamber Acquisition computer XWS-260 Workstation Compartment that contains the ThermoCube chiller unit and camera power supply Figure 1.1 IVIS ® Spectrum Imaging System The IVIS Spectrum is an integrated imaging system that includes a: •...
IVIS Spectrum Hardware Manual ® 1.2 IVIS Imaging Systems Technology Support For technical support, please contact Caliper at: Telephone 1.877.522.2447 (US) 1.508.435.9500 E-mail tech.support@caliperls.com 1.508.435.3439 Caliper Life Sciences Mail US Corporate Headquarters 68 Elm Street Hopkinton, MA 01748...
2. Important Safety Instructions 2.2 Instrument Labels Pay careful attention to the labels on the instrument. Label Description The Caliper IVIS Spectrum Imaging Instrument contains a Class 3R Laser module. ® The instrument itself is classified as a Class 1 device because the user is safely protected from the laser beam.
IVIS Spectrum Hardware Manual ® 2.4 Laser Safety The IVIS Spectrum is equipped with Class 3R laser (532 nm (green) laser source, 5 mW maximum continuous (CW), which is suitably embedded and limited in power to less than 1mW where accessible in the imaging compartment behind the front door.
2. Important Safety Instructions Water & Moisture VOLTAGE VOLTAGE! Do not use this product near water (for example, near a sink or wet room) due to risk of electric shock, electrical damage, and/or failure. The IVIS Spectrum requires a minimum of 100 square feet (9 square meters) Laboratory Space &...
IVIS Spectrum Hardware Manual ® A separate surge protector is provided for the computer and other accessory equipment. Facilities should be adequately wired according to local building codes. If the IVIS Imaging System experiences a loss of supply power, turn off the Power Outages power switch for all components and do not restart the system until reliable power has been restored.
2. Important Safety Instructions 2.10 Servicing Refer all servicing to IVIS Imaging Systems Technology Support. If the product is damaged and requires service, unplug the product from the outlet and contact technical support. Servicing by anyone other than those authorized by Caliper voids any warranty covering this product.
3. Warnings 3.2 Burn Hazard & Eye Safety Laser The IVIS Spectrum Imaging Instrument contains an embedded Class 3R Laser module: 532 nm (visible green) laser source, 5 mW maximum continuous (CW). The instrument itself is classified as a Class 1 device because the user is safely protected from the laser beam.
Page 19
IVIS Spectrum Hardware Manual ® Figure 3.2 IVIS Spectrum, imaging chamber door CAUTION CAUTION! DO NOT defeat any of the safety interlocks. Do not place anything under the imaging platform. The imaging platform cannot be moved unless the door is closed and locked. If the platform moves when the door is unlocked, shut down the system and contact IVIS Imaging Systems Technology Support.
Page 20
3. Warnings Filter wheel panel Figure 3.3 Filter wheel removable panel The laser beam is directed into the imaging chamber through an opening at the top of the chamber. Never place any object, tool, or fingers into this opening, otherwise damage to the galvanometer will result. WARNING WARNING! DO NOT look directly into this opening when the laser is operating.
IVIS Spectrum Hardware Manual ® 3.4 Chemical & Biological Safety Normal operation may involve the use of test samples that are pathogenic, toxic, or radioactive. It is your responsibility to ensure that all necessary safety precautions are taken before such materials are used. Dispose of all waste materials according to appropriate environmental health and safety guidelines.
4. Legal Notices OR ANY ASSOCIATED EQUIPMENT, COST OF CAPITAL, COST OF ANY SUBSTITUTE EQUIPMENT OR FACILITIES, DOWNTIME, THE CLAIMS OF ANY THIRD PARTIES, INCLUDING CUSTOMERS, AND INJURY TO PROPERTY, RESULTING FROM THE PURCHASE OR USE OF THE SYSTEM OR ARISING FROM BREACH OF THE WARRANTY, BREACH OF CONTRACT, NEGLIGENCE, STRICT TORT, OR ANY OTHER LEGAL OR EQUITABLE THEORY, EVEN IF CALIPER KNEW OF THE...
Page 25
IVIS Spectrum Hardware Manual ® Corporation in the United States and/or other countries. Adobe and Illustrator are either registered trademarks or trademarks of Adobe Systems Incorporated in the United States and/or other countries. Intel is a registered trademark of Intel Corporation.
Page 26
4. Legal Notices [This page intentionally left blank.]...
IVIS Spectrum Hardware Manual ® 5.6 Weights & Dimensions Item Specification Weight 275 kg (600 lbs) Depth 77 cm (30 inches) Width 65 cm (25.5 inches) Height 211 cm (83 inches) Imaging Chamber Internal 51 x 51 x 66 cm (D x W x H) Dimensions...
IVIS Spectrum Hardware Manual ® Figure 6.2 ThermoCube chiller unit 6.3 Imaging Chamber The imaging chamber is a highly specialized device consisting of the (Figure 6.3) imaging chamber housing, a heated, moveable platform, an auto focusing lens system with f/stop control, and synchronized filter wheels that control the spectral content of the luminescent or fluorescent images.
IVIS Spectrum Hardware Manual ® There are two 11-position filter wheels. Filter settings are selected by software Emission & Excitation control. For more details on how to access the filter wheels and change filters, Filter Wheels page Figure 6.5 Filter wheel For fluorescence studies, the IVIS Spectrum Imaging System can illuminate a ®...
6. IVIS ® Spectrum Components 6.6 Laser/Galvanometer System The laser/galvanometer system comprises a Class 3R laser (532 nm (visible green) laser source, 5 mW maximum continuous (CW). The laser beam is directed into the imaging chamber by two computer-controlled mirrors. WARNING WARNING! A laser beam can be harmful to the eyes.
IVIS Spectrum Hardware Manual ® Ventilation openings Ventilation openings Figure 6.7 IVIS Spectrum, back view 6.9 Acquisition Computer The system includes a Windows -based personal computer. The Living Image ® ® software as well as Microsoft Office software is installed on the computer. The Living Image software controls the IVIS Spectrum, and displays and analyzes the image data.
Page 38
6. IVIS ® Spectrum Components Table 6.1 Connector pin usage Pin Number Function 12 and 13 Use these pins to monitor the skin or rectal temperature of the animal. The use of a copper-constantan thermocouple is recommended. These pins may also be used to connect to the THM 100 ECG monitor (Indus Instruments).
IVIS Spectrum Hardware Manual ® 6.11 High Reflectance Hemisphere You can special order a high reflectance hemisphere that is used to check for light contamination in the imaging chamber (Figure 6.10) The hemisphere does not emit any photons when it is imaged in a light-tight and contaminant-free imaging chamber.
IVIS Spectrum Hardware Manual ® Operating the IVIS Spectrum ® Restarting the System ......Gas Plumbing ....... Door Operation .
Page 42
7. Operating the IVIS ® Spectrum Status light Bottom illumination input optics Gas valve Standby toggle switch Figure 7.1 IVIS Spectrum Table 7.1 IVIS Spectrum status light modes Status Light Is... Indicates... Blue The system is in standby mode. Blue and blinking The CCD camera cooling system is disabled due to a power interruption.
IVIS Spectrum Hardware Manual ® 5. In the Living Image software, click Initialize IVIS system in the IVIS System Control panel (Figure 7.2) — The status light is red. 6. Wait until the camera has cooled to operating temperature (requires about ten minutes).
Page 44
7. Operating the IVIS ® Spectrum CAUTION CAUTION! Caliper recommends using the XGI-8 Gas Anesthesia System when imaging small animals. This system supplies a controlled (Figure 7.3) amount of isoflurane to the imaging chamber and continuously reduces the build-up of isoflurane in the chamber. If you want to use a gas other than the recommended isoflurane/oxygen gas mixture or pure air, contact IVIS Imaging Systems Technology Support.
IVIS Spectrum Hardware Manual ® The mouse anesthesia manifold (part no. 117706) with transparent nose cones is provided with the IVIS Spectrum. The manifold has capacity for up to five ® mice and has a low profile (height) for operation with the system. NOTICE The IVIS Imaging System 100 Series and the XGI-8 Gas Anesthesia System are equipped with a different five-station manifold (PN 118274) that is taller...
Page 46
7. Operating the IVIS ® Spectrum To display the alignment grid, choose Alignment Grid On. Figure 7.4 IVIS System Control panel 2. Open the imaging chamber door to see the alignment grid projected by the laser/galvanometer on the imaging platform (Figure 7.5) —...
IVIS Spectrum Hardware Manual ® 7.5 Fluorescent Imaging The IVIS Spectrum has both bioluminescent and fluorescent imaging ® capabilities. To change from bioluminescent to fluorescent operation, select fluorescence mode in the IVIS System Control panel and select an appropriate excitation and emission filter (Figure 7.6) Choose luminescent or...
7. Operating the IVIS ® Spectrum In fluorescence mode, the excitation filter is automatically selected based on the Selecting an Excitation wavelength of the emission (imaging) filter selected from the IVIS System Filter Control panel. However, you can override the automatic selection in the IVIS System Control panel.
Page 49
IVIS Spectrum Hardware Manual ® 5. Turn off the main circuit breaker switch located on the rear of the IVIS Spectrum and disconnect the power cord from the wall socket. 6. Cover the gas inlet and outlet ports with masking tape to prevent entry of dust.
IVIS Spectrum Hardware Manual ® Troubleshooting Measured Temperature Does Not Equal the Demand Temperature . 45 Photographic Image Is Unacceptable ....Luminescent Image Is Unacceptable ....No Image Is Produced .
8. Troubleshooting 8.2 Photographic Image Is Unacceptable Default camera controls are programmed during the initialization of the Living Image software. Changes to these settings can greatly affect the photographic ® image. Check that the exposure time, f/stop and binning levels are set to the default conditions.
IVIS Spectrum Hardware Manual ® 8.3 Luminescent Image Is Unacceptable Binning, f/stop, and exposure time affect the appearance of a luminescent image. Please refer to the Living Image Software Manual for instructions on ® setting binning, exposure time, and f/stop values. In order to function properly and reduce camera noise, the CCD camera must be cooled to the demand temperature before acquiring an image.
8. Troubleshooting 8.4 No Image Is Produced If no image is produced, there may be an error in the Living Image software, a problem with the physical connections to the camera, or a hardware failure. 1. Close the Living Image software and restart the computer.
Page 55
IVIS Spectrum Hardware Manual ® 3. Turn off the rear panel circuit breaker switch. Front panel Back panel Toggle switch Figure 8.4 IVIS Spectrum, front and back panels 4. Check the connections between the computer and the rear panel of the IVIS Spectrum.
8. Troubleshooting 8.6 Accessing the Area Under the Imaging Platform During normal operation, the imaging platform always returns to the loading position before the door can be opened. This helps ensure user safety and convenience when loading subjects. Occasionally you may need access to the area beneath the platform for cleaning or to retrieve a subject.
IVIS Spectrum Hardware Manual ® Care & Maintenance Cleaning the IVIS Spectrum ..... . . Cleaning the Lens Protection Window ....Changing the Emission Filters Used in Fluorescence Imaging .
IVIS Spectrum Hardware Manual ® 4. Take an image to confirm that the image quality is restored. 9.3 Changing the Emission Filters Used in Fluorescence Imaging The following steps explain how to change one or more of the optional emission filters used for fluorescence imaging in the IVIS Spectrum.
Page 60
9. Care & Maintenance 7. To clean a filter, use a Dust-Off compressed gas duster to apply an air ® stream at an angle to the filter surface, otherwise dust particles could be driven into the filter coating. NOTICE Do not blow air onto a filter in the filter wheel as dust could be spread to other optical components.
IVIS Spectrum Hardware Manual ® 9.4 Moving the System It may be necessary to remove the plastic camera dome when you move the Removing the Dome IVIS Spectrum though a doorway or other low overhang. ® To remove the dome, wrap a flat blade screw driver with a thin cloth or ■...
Page 62
9. Care & Maintenance — The status light is red. The status light turns green when the system is initialized. Wait until the camera has cooled to the demand temperature before you begin an imaging session (about ten minutes).
IVIS Spectrum Hardware Manual ® Servicing & Maintaining the ThermoCube Chiller Unit Required Tools ......Turning Off Electric Power to the Main Console .
10. Servicing & Maintaining the ThermoCube Chiller Unit Figure 10.1 Lower rear panel Note that the power cord has been removed from the console. 10.3 Removing the Console Rear Panel 1. Use a 3/32" hex head wrench to remove the screws that secure the bottom rear (fan) panel (Figure 10.2) 2.
IVIS Spectrum Hardware Manual ® 10.4 Sliding the Tray Out The ThermoCube unit sits on a tray that must be pulled out to access the unit. 1. Use the ¼" hex head (Allen) wrench to loosen and remove the bolt that secures the tray under the ThermoCube unit to the console (Figure 10.3) Figure 10.3 Loosen the tray locking bolt...
10. Servicing & Maintaining the ThermoCube Chiller Unit 10.5 Filling the ThermoCube Unit 1. If additional coolant is required, wrap a piece of cloth or a paper towel around the neck of the tank to absorb accidental spills (Figure 10.5) 2.
IVIS Spectrum Hardware Manual ® 10.6 Reassembling the System 1. Replace the cap on the ThermoCube unit. Make sure that the cap is fully tightened to prevent rapid coolant evaporation. 2. Slide the tray back into the console while you hold the cables and hoses out of the way to prevent snagging.
Page 68
10. Servicing & Maintaining the ThermoCube Chiller Unit [This page intentionally blank.]...
IVIS Spectrum Hardware Manual ® Appendix A Spare Parts List Description Caliper Part Number Mouse Inhalation Nose Cone 117150 Low Profile, 5 Station Manifold 117706 XRH-1 High-Reflectance Hemisphere with container 118937 Black Paper, Swarthmore, Artagain. 9x12”, 24 sheet pad 117837 Low Fluorescence Imaging Mat, 11.6”...
Page 70
A. Spare Parts List [This page intentionally blank.]...
Page 71
IVIS Spectrum Hardware Manual ® Index emission filter selecting 41 accessory cable 31 emission filter wheel 29 back connector 32 emission filters stage connector 32 changing 53–54 anesthesia delivery system 38 environmental considerations 7 anesthesia manifold 39 heat 7 array plate 41 location of system 7 space and ventilation 8 water and moisture 8...
Page 72
Index gas plumbing 37 imaging platform patents 18 accessing space below 50 photographic image unacceptable 46 lock down bar 50 pinch hazard label 6 warming system 29 power cord 8 imaging system power line surges 8 proper location 7 power outages 9 information on manual instructions 6 power overloading 9 instrument labels...
Page 73
IVIS Spectrum Hardware Manual ® system cleaning 9 moving 9, 55–56 servicing 10 shut down 42–43 temperature measured unequal to demand 45 ThermoCube servicing 57–61 ThermoCube chiller unit 26 trademarks 18 transillumination 41 troubleshooting camera noise 47 light contamination 47 measured temperature does not equal demand temperature 45 no image produced 48...
Need help?
Do you have a question about the Caliper IVIS Spectrum and is the answer not in the manual?
Questions and answers