Service instructions
Laser sources
6.3 Error messages M = Machine safety
6.3.1 Error list
Bit
Message text
Significance
No.
Signal
BAK-protec
M 00
Pressure switch open, protect. glass broken
SLS 200 CL: Grounding connector line OS
not in inoperative position
Water LV
M 01
Water tank empty or leak in power supply
Leak OS
M 02
Leak in optical system
W-flow 1
M 03
Water flow too low in cavity 1
W-flow 2
M 04
Water flow too low in cavity 2
W-temp 1
M 05
Water temperature too high in cavity 1
W-temp 2
M 06
Water temperature too high in cavity 2
Airtemp A
M 07
Air temperature too high in cabinet A
Airtemp B
M 08
Air temperature too high in cabinet B
Airtemp C
M 09
Air temperature too high in cabinet C
Airtemp OS
M 10
Temp. OS much higher than cavity temp.
C-module 1
M 11
C-module 2
M 12
I N D U S T R I A L - L A S E R S
A COMPANY OF THE
Error discharge relay C-module level/cav 1
integrated in the RTPS 120/140 in the
FLS CL
Error discharge relay C-module level/cav 2
integrated in the RTPS 120/140 in the
FLS CL
6 Error messages
Measures
Insert new protection glass
Put away the grounding connector
line in the optical system
Call customer service
Seal leak, dry sensor
Check cavity for glass splinters in
cooling tube. Check water hoses
for kinks. Call customer service
Check cavity for glass splinters in
cooling tube. Check water hoses
for kinks. Call customer service
Check whether primary water is
flowing and not too warm. Adjust-
ment cooling valve.
Check whether primary water is
flowing and not too warm. Adjust-
ment cooling valve.
Call customer service
Call customer service
Call customer service
Danger of condensation in cavity
Set cooling water temperature
higher.
All laser systems:
Call customer service
SLS 200 CL / KLS 246 CL:
waiting time 5 minutes before the
laser system can be switched on
again
Call customer service
© Lasag AG, Thun, 26.02.08 CM
86 . 0621 - 06 - 09